Product introduction and general description of Variostat thermostats
Specp. 2
Technical specifications according to DIN 12876 standard
p. 1table, 32 rows
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-30...150 °C
Temperature stability at -10°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
1 kW
Cooling power at 100°C
0,3 kW
Cooling power at 20°C
0,3 kW
Cooling power at 0°C
0,2 kW
Cooling power at -10°C
0,18 kW
Cooling power at -20°C
0,12 kW
Cooling power at -30°C
0,03 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Pressure pump
yes
max. delivery
25 l/min
max. delivery pressure
0,7 bar
max. delivery (suction)
18,5 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
sound pressure level +/- 4 dB(A)
54 dB(A)
Overall dimensions WxDxH **
183x493x443 mm
Net weight
25 kg
Power supply requirement
220-240V 1~/2~ 50/60Hz
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.
1.0/26
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit's cooling capacity falls sharply at low temperatures, so the process load must be checked against the published curve at the actual working temperature.
Ensures the selected model can remove the required heat load at the coldest set point.
Source
“at -30°C 0,03 kW”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source assessment.
Flammable refrigerant classification drives installation location, ventilation and safety review.
Source
“R-290 (A3, H220)”
Installation
The unit is air-cooled, so sufficient clearance and cool intake air are required; capacity drops as ambient temperature rises.
Prevents undersized cooling performance and overheating in confined or warm rooms.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The permitted ambient range is 5 to 40 °C, which must be respected for reliable operation.
Defines the acceptable plant-room conditions for installation.
Source
“max. ambient temperature 40 °C”
Installation
The pump connections are M16x1 male, so mating fittings and hoses must match this thread.
Avoids mismatch of process plumbing and adapters during installation.
Source
“Pump connection M16x1 male”
Performance
The pressure pump delivers up to 25 l/min at 0.7 bar, which sets the maximum external circuit flow and pressure drop budget.
Needed to size external tubing, heat exchangers and pressure losses.
Source
“max. delivery 25 l/min”
Performance
The maximum permissible kinematic viscosity is 50 mm²/s, limiting the use of very viscous fluids at low temperatures.
Prevents pump overload and poor circulation when selecting the bath fluid.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The power supply tolerance is ±5% voltage with ±2% frequency, but the combination of -5% voltage and +2% frequency is not allowed.
Guides verification of site power quality before commissioning.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single/two-phase devices are supplied with a country-specific plug that must be specified at ordering.
Ensures the correct mains connection is delivered for the installation site.
Source
“with power cable and country-specific plug (please specify when ordering)”
Conformité
The equipment is EMC Class A Group 1, which may require additional measures in residential or sensitive environments.
Class A equipment is intended for industrial environments and may cause interference elsewhere.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Protects the bath and process from overheating if the control loop fails.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The controller supports remote control and data transmission via USB with Spy software and via the network port.
Relevant for integration into lab data acquisition and remote operation.
Source
“The USB port allows connection of the system to a PC or notebook.”