Introduces the IMG-300 inverted magnetron gauge and its principle
Warningp. 11
Electromagnetic compatibility warnings and compliance notes
Overviewpp. 13-18
Describes general purpose inverted magnetron gauge details
Specpp. 15-19
Technical specifications and requirements tables for the gauge
Orderingpp. 19-20
Ordering codes and configuration options for gauge variants
pp. 20-24
Vacuum Products Division contact and company information
p. 15table, 9 rows
p. 20table, 40 rows
Table 1-1 IMG-300 Technical Specifications
Specification
Definition
Operating voltage (HV range)
3 kV DC
Sensitivity
2.5 A/Torr + 100%; -5% @ 5x 10−6 Torr
Measurement Range
10-3 to 10-11 Torr
Maximum Operating Temperature
250 °C, CFF metal gasket; with magnet
Electrical Connector
SHV style, 5kV rated
Feedthrough/UHV
Ceramic brazed
Ignition Time
< 5 seconds at pressures > 1E-6 Torr, typically 20 min @ 1E-10 Torr
Response Time
< 10 milliseconds for increasing pressure
Material Exposed to the Vacuum
304L Stainless steel, Nickel, 2% Thoriated Tungsten, and ceramic
Request for Return Form (Health and Safety Certification)
Section
Field
Value
1) CUSTOMER INFORMATION
Company Name
1) CUSTOMER INFORMATION
Contact Name
1) CUSTOMER INFORMATION
Tel
1) CUSTOMER INFORMATION
Email
1) CUSTOMER INFORMATION
Fax
1) CUSTOMER INFORMATION
Customer Ship To
1) CUSTOMER INFORMATION
Customer Bill To
1) CUSTOMER INFORMATION
Europe only: VAT reg. Number
1) CUSTOMER INFORMATION
USA/Canada only: Taxable
1) CUSTOMER INFORMATION
USA/Canada only: Non-taxable
2) PRODUCT IDENTIFICATION
Product Description
2) PRODUCT IDENTIFICATION
Agilent P/N
2) PRODUCT IDENTIFICATION
Agilent S/N
2) PRODUCT IDENTIFICATION
Original Purchasing Reference
3) TYPE OF RETURN
3A. Non-Billable
3) TYPE OF RETURN
3A. Billable
3) TYPE OF RETURN
3A. New PO # (hard copy must be submitted with this form)
3) TYPE OF RETURN
3B. Exchange
3) TYPE OF RETURN
3B. Repair
3) TYPE OF RETURN
3B. Upgrade
3) TYPE OF RETURN
3B. Consignment/Demo
3) TYPE OF RETURN
3B. Calibration
3) TYPE OF RETURN
3B. Evaluation
3) TYPE OF RETURN
3B. Return for Credit
4) HEALTH and SAFETY CERTIFICATION
HAS NOT pumped or been exposed to any toxic or hazardous materials
4) HEALTH and SAFETY CERTIFICATION
HAS pumped or been exposed to the following toxic or hazardous materials
4) HEALTH and SAFETY CERTIFICATION
Toxic
4) HEALTH and SAFETY CERTIFICATION
Corrosive
4) HEALTH and SAFETY CERTIFICATION
Reactive
4) HEALTH and SAFETY CERTIFICATION
Flammable
4) HEALTH and SAFETY CERTIFICATION
Explosive
4) HEALTH and Safety Certification
Biological
4) HEALTH and Safety Certification
Radioactive
4) HEALTH and Safety Certification
List all toxic/hazardous materials. Include product name, chemical name, and chemical symbol or formula
4) HEALTH and Safety Certification
Print Name
4) HEALTH and Safety Certification
Authorized Signature
4) HEALTH and Safety Certification
Date
5) FAILURE INFORMATION
Failure Mode (REQUIRED FIELD. See next page for suggestions of failure terms)
5) FAILURE INFORMATION
Detailed Description of Malfunction: (Please provide the error message)
5) FAILURE INFORMATION
Application (system and model)
Форма содержит чекбоксы и поля для заполнения; пустые ячейки оставлены пустыми строками.
Engineering Facts
14 facts · every claim linked to a page
Safety
The controller and vacuum system must be separately connected to a common ground, and a ground wire must not be placed between the vacuum chamber and controller chassis.
Improper grounding can create large circulating currents and lethal high-voltage hazards, so grounding must be verified before operation.
Source
“Do not place a ground wire between the vacuum chamber and the controller chassis; large continuous currents could flow through it.”
Safety
The system ground must be tested to confirm it is complete and can support at least 10 A.
A verified low-impedance ground is required to safely handle the 3 kV high voltage present in the gauge.
Source
“Test the system ground to be sure that it is complete and capable of supporting at least 10 A.”
Safety
The gauge is not intrinsically safe and must never be turned on when combustible gases or mixtures are present.
The high-voltage discharge and filament/grid can act as ignition sources, so the gauge must be isolated from combustible atmospheres.
Source
“As with all ionization gauges, this device is not intrinsically safe.”
Installation
The IMG-300 must not be mounted closer than a few inches to any magnetic material because the magnet generates an external field.
Magnetic material near the gauge distorts the field and can change calibration and starting behavior.
Source
“Do not mount the IMG-300 closer than a few inches to any magnetic material.”
Installation
Rough handling or prolonged contact with the metal surface can weaken the magnet and shift calibration.
Mechanical abuse of the magnet assembly degrades measurement accuracy, so careful mounting and handling are required.
Source
“Avoid rough handling or prolonged contact with the metal surface which can cause a decrease in magnet strength and result in a calibration change.”
Operation
The gauge can be baked out at up to 250 °C with the cable unplugged.
Bakeout capability determines UHV compatibility, but the cable must be disconnected to avoid damage during bakeout.
Source
“The IMG-300 with CFF handles bakeout temperatures of up to 250 °C, with the cable unplugged.”
Performance
The gauge covers the pressure range from 10-3 Torr down to 10-11 Torr.
The measurement range must match the application's operating pressure to select the correct gauge.
Source
“measures vacuum pressure from 10-3 Torr (10-3 mbar) down to 10-11 Torr (10-11 mbar).”
Performance
Ignition time is under 5 seconds above 1E-6 Torr but typically about 20 minutes at 1E-10 Torr.
Slow starting at high vacuum affects process timing and may require patience or starting aids in UHV applications.
Source
“Ignition Time < 5 seconds at pressures > 1E-6 Torr, typically 20 min @ 1E-10 Torr”
Performance
The response time is under 10 milliseconds for increasing pressure.
Fast response is important for leak detection and rapid pressure-change monitoring.
Source
“Response Time < 10 milliseconds for increasing pressure”
Selection
The gauge uses a standard SHV connector rated to 5 kV and a coaxial cable for connection to a Multi-Gauge.
The connector type and 5 kV rating must match the controller cable and high-voltage supply.
Source
“A standard SHV connector and coaxial cable provides for a safe, easy connection to a Multi-Gauge.”
Selection
All materials exposed to vacuum are high-vacuum compatible and no elastomers are used in construction.
Elastomer-free all-metal construction reduces outgassing and is suitable for UHV applications.
Source
“All materials exposed to the vacuum system are high vacuum compatible. No elastomers are used in IMG-300 construction.”
Maintenance
Silicone oil and silicone grease must not be used when servicing the vacuum equipment.
Silicone contamination migrates and degrades vacuum performance, so only approved lubricants may be used.
Source
“Do not use silicone oil or silicone grease.”
Maintenance
Alconox is incompatible with aluminum and must not be used to clean aluminum parts.
Using the wrong cleaning agent can corrode aluminum components and introduce contamination.
Source
“Do not clean any aluminum parts with Alconox. AIconox is not compatible with aluminum and will cause damage.”
Conformité
The product is Class A equipment and may cause radio interference in domestic environments.
Class A rating means the gauge is intended for commercial/industrial use, not residential, and EMC measures may be needed.
Source