Protector XL Laboratory Hoods Manual

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Protector XL Laboratory Fume Hoods in 4', 5', 6', 8', 10', 12', and 16' widths, with standard vertical-rising sash and optional A-Style Combination Sash, with or without bypass block, and with Eco Energy Saving or Flush air foils. Also covers optional accessories: Guardian airflow monitors, sash stops, ceiling enclosures, rear panels, additional service fixtures, distillation grids, upper bypass, electrical duplex outlets, and 8' dual exhaust adapter. Appendix D covers the Protector I-S Hood with Intelli-Sense Blower 3-speed operation.

Section-by-Section Technical Breakdown

pp. 9-10
Location and Support Requirements

The hood must be located away from traffic, doors, windows, fans, and ventilation registers that could disrupt airflow, with room windows closed. Installation must be permanent and stationary on a base cabinet and chemically resistant work surface; carts, dollies, and mobile benches are prohibited.

  • Locate away from traffic patterns, doors, windows, fans, ventilation registers, and other air-handling devices.
  • All windows in the room should be closed.
  • Do not install on a cart, dolly, or mobile bench; all installations must be permanent and stationary.
  • Supporting structure usually consists of a base cabinet and chemically resistant work surface.
pp. 10-11
Exhaust Requirements

The exhaust duct connection is designed for 12-inch nominal duct (12.75-inch OD) to minimize static pressure loss at 60-100 fpm face velocity and provide 1000-2500 fpm transport velocity. Tables list airflow volumetric rates and static pressures for each hood width, sash position, and bypass configuration.

  • Exhaust connection designed for 12" nominal duct (12.75" OD).
  • Face velocity range 60 to 100 fpm.
  • Transport velocity away from hood 1000 to 2500 fpm.
  • Airflow tables cover 4', 5', 6', 8', 10', 12', and 16' hoods at sash full open (28") and 62.5% open (18").
  • Example: 4' XL at 100 fpm full open requires 725 CFM at 0.27" w.c.; 8' XL at 100 fpm full open requires 1640 CFM at 0.31" w.c.
pp. 12, 18
Electrical Requirements

Hoods are normally wired for 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A service. All internal wiring for LED light, light switch, blower switch, and duplex outlets terminates at a single-point junction box for connection by a qualified electrician. Each circuit should be a dedicated branch circuit; if combined, maximum load is the sum of individual outlets plus 2 A unit rating.

  • Standard electrical service: 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A.
  • Verify voltage on I.D. plate behind front panel.
  • Single-point internal junction box for lights, blower, and duplex outlets.
  • Dedicated branch circuit recommended; combined maximum load = sum of outlets + 2 A.
  • 115 V wiring: black = phase (hot), white = neutral; 230 V: brown = Phase 1, blue = Phase 2.
  • Ground to MAINS protective earthing lug using ring terminal for 10-24 screw.
pp. 12, 20-21
Service Line Requirements

Service lines should be 1/4-inch OD copper (brass for natural gas) with an accessible shut-off valve. Recommended operating pressure is 40 PSI with a maximum allowable pressure of 200 PSI. Standard service fixtures support air, hot water, vacuum, cold water, and natural gas; oxygen requires special fixtures from Labconco.

  • Supply tubing: 1/4" OD copper; brass for natural gas.
  • Recommended operating pressure: 40 PSI; maximum allowable: 200 PSI.
  • Consider a pressure regulator to reduce line pressure to 40 PSI.
  • Standard services: air, hot water, vacuum, cold water, natural gas.
  • Do not use oxygen with standard service fixtures; contact Labconco for oxygen fixtures.
  • Natural gas must only be used in the fixture pre-plumbed with brass tubing.
pp. 13-15
Unpacking and Shipping Skid Removal

Hoods weigh 400 to 1600 lbs (182-726 kg) and must remain on the shipping skid until close to the final location. Move with a floor jack or furniture dolly under the skid; do not tilt onto a hand truck. Baffles ship loose behind the hood: upper 16.5" tall, middle 23.3" tall, lower perforated 17.3" tall.

  • Hood weight range: 400 to 1600 lbs (182-726 kg).
  • Leave hood attached to shipping skid until as close to final location as possible.
  • Move using a suitable floor jack or furniture dolly under the skid; do not tilt onto a hand truck.
  • Baffles ship loose: upper 16.5" tall, middle 23.3" tall, lower perforated 17.3" tall.
  • Do not lift by the front air foil.
pp. 15
Sash Weight Release

Sash weights are secured to the back of the hood with screws for shipment. Remove screws and verify sash cables or chains are on pulleys or sprockets before operating the sash. On multi-sash models, weights are lag-screwed to the skid and must be attached to respective sash cables using provided hooks.

  • Remove shipping screws securing sash weight to back of hood.
  • Verify sash cables or chains are on pulleys or sprockets before sash operation.
  • On models with more than one sash, remove weights from skid and attach to respective sash cables using hooks.
  • Sash weight is individually matched to the specific hood and must not be exchanged with another unit.
pp. 15-16
Installing the Hood on a Supporting Structure and Work Surface

The work surface must support the combined weight of hood and equipment, be at least as wide as the hood, and be aligned flush with the back of the hood for proper bypass airflow. A cross support must be installed under the rear of the work surface after leveling base cabinets and before placing the hood.

  • Work surface must safely support combined weight of hood and related equipment.
  • Work surface should be at least as wide as the hood.
  • Work surface aligned flush with back of fume hood for correct spacing under air foil.
  • Install cross support after base cabinets and work surface are leveled and before installing hood.
  • Work surface should be smooth, nonporous, chemically resistant (e.g., epoxy resin) with dished recess for spills.
  • Hood depth options: 31.7", 37.7", 43.7"; internal depths 24", 30", 36"; work surface depths 30", 36", 42".
pp. 17
Connecting to the Hood Exhaust System

Exhaust ductwork weight must be supported independently of the hood superstructure. The exhaust connection is 12-inch nominal (12.75-inch OD) and should be installed by a qualified HVAC contractor. Duct material must match hood procedures and chemicals for compatibility.

  • Support exhaust ductwork weight independently; do not load hood structure.
  • Exhaust connection: 12" nominal pipe (12.75" OD).
  • Installation by qualified HVAC contractor.
  • Select duct material compatible with procedures and chemicals used.
pp. 17-18
Connecting the Electrical Supply Source

Building electrical supply must include overload protection with a switch or circuit breaker in close proximity, marked as the disconnecting device. The hood must be grounded to MAINS protective earthing. Wire color codes: 115 V black/white, 230 V brown/blue. Refer to Appendix C wiring diagram.

  • Provide overload protection and a marked disconnecting means within easy reach of operator.
  • Consult NEC-2002 for proper installation.
  • Ground to MAINS protective earthing lug; only MAINS ground conductors on this lug.
  • Use ring terminal sized for 10-24 machine screw for ground connection.
  • 115 V: black = phase (hot), white = neutral; 230 V: brown = Phase 1, blue = Phase 2.
  • All connections terminate at field wiring terminal box for hook-up by qualified electrician.
pp. 20
Connecting the Service Lines

Hoods with service fixtures are plumbed from valve to hose connector or gooseneck. Supply tubing by qualified installer can enter from above, through the back, or through the work surface. Inspect all fittings for leakage and tighten slightly if needed.

  • Supply tubing provided by qualified installer.
  • Tubing can enter from above, through the back, or through the work surface.
  • Inspect all fittings for leakage; tighten slightly if needed.
  • Service access plate on front corner posts allows access to valve bodies.
  • Do not use oxygen with standard service fixtures.
pp. 21
Sealing and Certifying the Hood

After setting, ducting, wiring, and plumbing, seal the hood to the work surface with silicone sealant to prevent spilled materials from collecting under walls. Certify by measuring face velocity across the sash opening; Labconco recommends average face velocity of 60 to 100 fpm. Hoods are factory tested per ASHRAE 110-1995 with as-manufactured rating <0.05 ppm at 4 L/min.

  • Seal hood to work surface with silicone sealant after installation.
  • Cupsinks sealed with silicone sealant to work surface.
  • Recommended average face velocity: 60 to 100 fpm.
  • Factory tested per ASHRAE 110-1995; as-manufactured rating AM<0.05 ppm at 4 L/min.
  • Face velocity profiles and smoke testing should be done periodically.
pp. 22-23
Performance Features

The Protector XL is a bypass hood with baffle and air foil designed for OSHA-approved velocities as low as 60 fpm. Features include 37.5-inch high visibility sash with 28-inch operating travel, 48-inch interior height, Opti-Zone baffle, LED lighting rated 50,000 hours, ADA-compliant switches, and factory preparation for up to 8 service fixtures.

  • Sash provides 37.5" visibility with 28" operating travel.
  • Interior height 48".
  • Bypass airflow design maintains containment as sash moves.
  • Opti-Zone baffle optimizes airflow and face velocity profile.
  • LED T8 bulbs rated 50,000 hours, use approximately half the power of fluorescent.
  • Factory prepared for up to 8 service fixtures.
  • Duplex receptacles mounted on right and left corner posts, factory-wired to single-point junction box.
  • Guardian airflow monitor factory-prepared location on right corner post.
pp. 28-30
Routine Maintenance

Chapter 6 covers service safety precautions, a routine maintenance schedule, and routine service operations. Specific maintenance intervals and procedures are detailed in the manual.

  • Service safety precautions must be followed.
  • Routine maintenance schedule provided.
  • Routine service operations described.
pp. 31-36
Modifying Your Protector XL

Chapter 7 covers field installation of work surfaces, ceiling enclosures, rear panels, additional service fixtures, Guardian airflow monitors, distillation grids, sash stop kits, upper bypass, electrical duplex outlets, and 8' dual exhaust adapter.

  • Installing work surfaces.
  • Installing ceiling enclosures above the fume hood.
  • Installing rear panels behind the fume hood.
  • Installing additional service fixtures.
  • Installing Guardian airflow monitors.
  • Distillation grids field installation.
  • Sash stop kit field installation.
  • Upper bypass field installation.
  • Installing electrical duplex outlet.
  • 8' dual exhaust adapter.
pp. 37-39
Troubleshooting

Chapter 8 provides a table of problems, probable causes, and corrective actions for electrical, airflow, sash, service valve, and duplex outlet issues.

  • Covers remote blower and light failures.
  • Covers sash operation problems.
  • Covers service valve and duplex outlet issues.
  • Covers contaminant and face velocity problems.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication S'applique à
Exhaust duct connection designed for 12-inch nominal pipe (12.75-inch OD). p.10 ↗ Ductwork must be sized and installed to this connection; use qualified HVAC contractor. All Protector XL models
Recommended average face velocity at sash opening is 60 to 100 fpm. p.21 ↗ Certification and periodic testing must verify face velocity within this range. All Protector XL models
Transport velocity away from hood should be 1000 to 2500 fpm. p.10 ↗ Exhaust system must maintain this velocity range for proper transport. All Protector XL models
Hood weight range: 400 to 1600 lbs (182-726 kg). p.13 ↗ Supporting structure and lifting equipment must be rated for these weights. All Protector XL models
Standard electrical service: 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A. p.18 ↗ Verify voltage on I.D. plate; provide dedicated branch circuit. All Protector XL models
If circuits are wired together, maximum load allowed is sum of individual outlets plus 2 A unit rating. p.18 ↗ Do not exceed this combined load when sharing branch circuits. All Protector XL models
Service line tubing: 1/4-inch OD copper (brass for natural gas); recommended operating pressure 40 PSI, maximum 200 PSI. p.12 ↗ Use pressure regulator to reduce line pressure to 40 PSI; ensure material compatibility. Hoods with service fixtures
Baffle dimensions: upper 16.5" tall, middle 23.3" tall, lower perforated 17.3" tall. p.14 ↗ Verify correct baffle installation per figures; baffles ship loose. All Protector XL models
Sash provides 37.5-inch visibility with 28-inch operating travel; interior height 48 inches. p.22 ↗ Sash stops can limit height to reduce energy usage. All Protector XL models
LED T8 bulbs rated 50,000 hours and use approximately half the power of fluorescent bulbs. p.23 ↗ Consider LED for energy efficiency; serviceable from outside hood cavity. All Protector XL models
Factory prepared for up to 8 service fixtures. p.23 ↗ Plan service fixture additions within this capacity. All Protector XL models
Hoods tested per ASHRAE 110-1995 with as-manufactured rating AM<0.05 ppm at 4 L/min. p.21 ↗ Field certification should be performed by certified contractor. All Protector XL models
Work surface depth options: 30", 36", 42" corresponding to hood depths 31.7", 37.7", 43.7". p.16 ↗ Select work surface depth to match hood model and ensure flush rear alignment. All Protector XL models
Example airflow: 4' XL at 100 fpm full open requires 725 CFM at 0.27" w.c.; 8' XL at 100 fpm full open requires 1640 CFM at 0.31" w.c. p.11 ↗ Use airflow tables to size blower and duct system. 4' and 8' XL models
Ground connection requires ring terminal sized for 10-24 machine screw; only MAINS ground conductors on protective earthing lug. p.18 ↗ Ensure proper grounding for safe operation. All Protector XL models
Natural gas must only be used in service fixture pre-plumbed with brass tubing due to sulfur content deteriorating copper. p.21 ↗ Verify fixture tubing material before connecting natural gas. Hoods with natural gas service fixture

Faults & Troubleshooting

Code / Symptôme Possible Cause Remedy Note
Remote blower and lights won’t operate p.37 ↗ Wires not connected at junction boxes or switches. Check connection of switches. Check connection to control box on top of unit.
Remote blower and lights won’t operate p.37 ↗ Circuit breakers tripped in building electrical supply. Reset circuit breakers.
Remote blower won’t operate, but lights work p.37 ↗ Blower wiring is disconnected. Inspect blower wiring and switch.
Remote blower won’t operate, but lights work p.37 ↗ Belt broken Replace belt.
Remote blower won’t operate, but lights work p.37 ↗ Blower motor is defective. Replace blower motor.
Fume hood blower operates but lights will not operate p.37 ↗ Lamp not installed correctly. Inspect lamp installation.
Fume hood blower operates but lights will not operate p.37 ↗ Lamp is defective. Replace lamp.
Fume hood blower operates but lights will not operate p.37 ↗ Lamp circuit breaker in building is tripped. Reset the lamp circuit breaker.
Fume hood blower operates but lights will not operate p.38 ↗ Lamp wiring is disconnected. Inspect lamp wiring.
Fume hood blower operates but lights will not operate p.38 ↗ Defective lamp ballasts. Replace lamp ballasts.
Contaminants outside of fume hood p.38 ↗ Improper user techniques for the fume hood. See “Certifying the Hood” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix E) Ref. Appendix E
Contaminants outside of fume hood p.38 ↗ Restriction of the baffle air slots or – blockage of the exhaust outlet. Remove baffles to ensure that all air slots, and the exhaust outlet are unobstructed.
Contaminants outside of fume hood p.38 ↗ External factors are disrupting the fume hood airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix E) Ref. Appendix E
Fume hood has improper face velocity. p.38 ↗ Not specified. Have fume hood re-certified and check remote blower exhaust system. Hood should have average face velocity of 60-100 fpm.
Vertical sash no longer operates smoothly p.38 ↗ Cable is frayed or plastic protection is damaged. Inspect cable and replace cable if worn or damaged immediately; otherwise injury could result.
Vertical sash no longer operates smoothly p.38 ↗ Pulley bearing is damaged. Replace pulley, bearing or add grease.
Vertical sash no longer operates smoothly p.38 ↗ Cable or chain has slipped off the pulleys or sprockets. Re-install, cable or chain must be replaced immediately if damaged.
Vertical sash no longer operates smoothly p.38 ↗ Weight has broken pulleys or sprockets. Replace pulleys or sprockets.
Combination A-Style sash no longer operates smoothly p.38 ↗ Horizontal glass panels have come off the tracks. Re-install horizontal glass on tracks.
Combination A-Style sash no longer operates smoothly p.39 ↗ Vertical sash frame is distorted. Place horizontal glass symmetrically and pull sash down to air foil. Straighten damaged frame.
Combination A-Style sash no longer operates smoothly p.39 ↗ Cable is frayed or has slipped off the pulleys. Re-install, cable must be replaced immediately if damaged.
Electrical duplex outlets no longer have power p.39 ↗ Wires not connected or faulty duplex. Check wire connection or replace duplex.
Electrical duplex outlets no longer have power p.39 ↗ Circuit breakers tripped in building electrical supply. Reset circuit breakers.
Service valves no longer operate p.39 ↗ Faulty building supply. Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI.
Service valves no longer operate p.39 ↗ Valve no longer operates. Replace valve and check for leaks.
Service valves no longer operate p.39 ↗ Supply line or outlet line has leaks. Inspect line for leaks and fix any leaking plumbing connections.