Rev g xpert bulk powder enclosures and systems stations user manual

Technical Reference Summary

Covered Models & Document Scope

Cubre las Cabinas para Polvo a Granel XPert de Labconco (anchos de 5', 6', 8'; 115 V y 230 V), los Sistemas Filtrados para Polvo a Granel XPert (5', 6') y las Estaciones Filtradas para Polvo a Granel XPert (5', 6'). Códigos de pedido: Cabinas 3963500/3963520 (5' L), 3963501/3963521 (5' C), 3963502/3963522 (5' R), 3963600/3963620 (6' L), 3963601/3963621 (6' C), 3963602/3963622 (6' R), 3963800/3963820 (8' L), 3963801/3963821 (8' C), 3963802/3963822 (8' R); Sistemas Filtrados 3961500/3961520 (5' L), 3961501/3961521 (5' C), 3961502/3961522 (5' R), 3961600/3961620 (6' L), 3961601/3961621 (6' C), 3961602/3961622 (6' R); Estaciones Filtradas 3962500/3962520 (5' L), 3962501/3962521 (5' C), 3962502/3962522 (5' R), 3962600/3962620 (6' L), 3962601/3962621 (6' C), 3962602/3962622 (6' R).

Section-by-Section Technical Breakdown

pp. 10-12, 16-19
Requisitos de escape y soplador

Define las dimensiones del collarín de escape, los adaptadores de transición y el volumen de escape requerido, la pérdida de presión estática, la cantidad de FilterMate y el nivel de ruido a velocidades de cara de 60, 80 y 100 fpm para cada ancho de cabina. También cubre los datos de escape del Sistema/Estación filtrados y el requisito de una compuerta ajustable cuando se conecta al escape del edificio.

  • Exhaust connection accepts a 2" x 10" (5.1 cm x 25.4 cm) nominal collar.
  • 5' and 6' enclosures require one exhaust connection to house exhaust; 8' requires two.
  • 6" Transition Adapter 3912401 recommended for house exhaust on all 5', 6', and 8' enclosures.
  • 5" Transition Adapter 3912400 required for FilterMate Portable Exhausters.
  • 8' enclosure house exhaust may use 3981600 Exhaust Transition (6" x 6" x 8").
  • Adjustable damper (or valve) must be installed when connected directly to house exhaust.
  • Filtered Station requires 10" diameter ductwork for 5' and 6' models and a remote blower.
pp. 12, 20
Requisitos eléctricos y conexión de suministro

Especifica las capacidades estándar de los receptáculos dúplex, la caja eléctrica cuádruple debajo de la superficie de trabajo y el amperaje máximo para el circuito del soplador remoto en la Estación XPert. También cubre los requisitos de cableado y la prohibición de cables de alimentación desmontables con capacidad inadecuada.

  • Standard duplex receptacles: 115V, 12A or 230V, 6A.
  • Four-plex electrical box: four outlets rated 115V, 12A or 230V, 6A.
  • XPert Station remote blower circuit: maximum 9A at 115V or 4.5A at 230V.
  • All building wiring requires a licensed electrician and must conform to local codes.
  • Do not use any detachable power cord not adequately rated for the unit.
pp. 13-15, 20
Instalación y montaje

Cubre el desembalaje, traslado e instalación de la cabina superior y la superficie de trabajo sobre el soporte elevador, incluidos los números de pieza y las cantidades de los herrajes de fijación. También cubre el sellado de la cabina superior a la superficie de trabajo con sellador de silicona.

  • Upper enclosure with attached work surface weighs 100 to 550 lbs (45 to 248 kg); lift stands weigh 380 to 500 lbs (171 to 225 kg).
  • Do not lift by the front or side air foils.
  • Fastening hardware: 1880740 Hex Bolt ¼-20 x 2.5 (qty 8); 1880708 Hex Head Screw ¼-20 x 0.5 (qty 3).
  • Seal upper enclosure to work surface with a clear bead of silicone sealant.
pp. 20
Ajuste y validación de la velocidad de cara

Especifica el rango de velocidad de cara para operaciones de pesaje y describe las pruebas de validación de fábrica e independientes. Requiere perfiles de velocidad de cara y pruebas de humo según el sistema de calidad de la organización.

  • Face velocity should be adjusted from 60 to 100 fpm for all weighing operations.
  • Factory tested per ASHRAE 110-1995; as-manufactured rating less than 0.05 ppm at 4 L/min.
  • Independent validation performed using sodium naproxen powder.
  • Face velocity profiles and smoke testing should be performed frequently.
pp. 21-23
Características de alto rendimiento

Describe las características de diseño que mejoran la contención, incluidos los perfiles aerodinámicos de contención de barrera, el deflector perforado trasero, el suministro superior de aire de dilución, el carro de transferencia, la elevación automatizada, los anillos de retención, la manga de seguridad, las bandas de retención, la superficie de trabajo de acero inoxidable, la filtración HEPA, el conducto de residuos y la caja eléctrica cuádruple.

  • Automated lift: 12" travel and 350 pounds capacity; photoelectric eye stops container below work surface.
  • Retaining rings available in various diameters up to 19.6" dia.
  • Waste chute requires hazardous waste bags minimum 8-1/2" wide x 11" long.
  • Interior dimensions: 30" depth and 32" height.
  • HEPA filters include true Bag-In/Bag-Out filter change system.
pp. 25-28
Uso de su cabina

Proporciona pasos de planificación, puesta en marcha, carga, técnicas de trabajo, procedimiento de trabajo, descarga, limpieza y apagado. Incluye requisitos de tiempo para purgar contaminantes aerotransportados y el manejo de contenedores de origen con mangas de seguridad y anillos de retención.

  • Allow enclosure to operate for 1 minute prior to starting procedures.
  • After loading and attaching bags, wait 15 seconds to purge airborne contaminants.
  • Upon completion of work, allow enclosure to operate for two minutes undisturbed.
  • Elevate large objects at least ¼" above the work surface to permit airflow underneath.
  • Keep all materials behind the lower air foil and perform contaminated operations as far to the rear as possible.
pp. 25-26
Mantenimiento y programación de la Alarma Digital Guardian 1000

Cubre el programa de mantenimiento rutinario y la programación de la Alarma Digital Guardian 1000.

  • Routine maintenance schedule is provided.
  • Guardian 1000 Digital Alarm programming instructions are included.
pp. 28-34
Accesorios para su cabina

Enumera los accesorios disponibles, incluidos mangas de seguridad, bandas de retención, anillos de retención, kits de luz fluorescente, adaptadores de transición de escape, sopladores remotos, compuertas de extractores, Extractor Portátil FilterMate y filtros, filtros HEPA y de control de olores traza, conexión de campana, carro de transporte, kits de estantes, kit de filtración HEPA doble, kits de mangueras, kit de ruedas y extensión de plataforma elevadora.

  • Remote blowers and FilterMate Portable Exhausters are available for exhaust.
  • HEPA and trace odor control filters are available.
  • Double HEPA Filtration Kit is available.
  • Canopy Connection aids in connecting to remote blower or house exhaust.
pp. 39-41
Solución de problemas

Proporciona una tabla de problemas, causas y acciones correctivas para contaminación, funcionamiento del soplador, velocidad de cara, iluminación, monitor de flujo de aire y fallos del elevador.

  • Covers contamination outside enclosure, blower won't operate, low face velocity, blower and lights won't operate, lights do not work, airflow monitor malfunction, and lift won't operate or malfunctions.
  • Includes service operating record log.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication S'applique à
La velocidad de cara debe ajustarse de 60 a 100 fpm para todas las operaciones de pesaje. p.20 ↗ Set exhaust blower or impeller speed to achieve average face velocity within this range for proper containment. All XPert Bulk Powder Enclosures, Systems, and Stations
La cabina de 5' a 60/80/100 fpm requiere volúmenes de escape de 290/385/480 CFM con pérdidas de presión estática de .20"/.35"/.54" H2O y 2/2/3 FilterMates. p.10 ↗ Select blower or number of FilterMate Portable Exhausters to match required exhaust volume and static pressure at the desired face velocity. 5' XPert Bulk Powder Enclosure
La cabina de 6' a 60/80/100 fpm requiere volúmenes de escape de 350/465/580 CFM con pérdidas de presión estática de .27"/.48"/.75" H2O y 2/3/3 FilterMates. p.10 ↗ Select blower or number of FilterMate Portable Exhausters to match required exhaust volume and static pressure at the desired face velocity. 6' XPert Bulk Powder Enclosure
La cabina de 8' a 60/80/100 fpm requiere volúmenes de escape de 465/620/775 CFM con pérdidas de presión estática de .15"/.27"/.42" H2O y 3/3/3 FilterMates. p.10 ↗ Select blower or number of FilterMate Portable Exhausters to match required exhaust volume and static pressure at the desired face velocity. 8' XPert Bulk Powder Enclosure
El Sistema/Estación Filtrados de 5' a 60/80/100 fpm requiere volúmenes de escape de 290/385/480 CFM con pérdida de presión estática inicial con filtro HEPA de .38"-.46"/.55"-.67"/.72"-.85" H2O. p.11 ↗ Size remote blower for Station or verify System impeller performance against these pressure losses. 5' XPert Bulk Powder Filtered Balance System or Station
El Sistema/Estación Filtrados de 6' a 60/80/100 fpm requiere volúmenes de escape de 350/465/580 CFM con pérdida de presión estática inicial con filtro HEPA de .46"-.56"/.67"-.81"/.87"-1.03" H2O. p.11 ↗ Size remote blower for Station or verify System impeller performance against these pressure losses. 6' XPert Bulk Powder Filtered Balance System or Station
Amperaje máximo del circuito del soplador remoto de la Estación XPert: 9 amperios a 115 V o 4.5 amperios a 230 V. p.12 ↗ Ensure remote blower electrical draw does not exceed these limits when connected to the enclosure blower switch. XPert Filtered Balance Station
Tomas de la caja eléctrica cuádruple nominales a 115 V, 12 A o 230 V, 6 A. p.12 ↗ Total connected load to the convenience outlets must not exceed these ratings. All models
El elevador automatizado tiene un recorrido de 12" y una capacidad de 350 libras. p.22 ↗ Do not exceed 350 lb source container weight; lift height adjustment limited to 12 inches. All models with automated lift
Anillos de retención disponibles en varios diámetros hasta 19.6" de diámetro. p.22 ↗ Select retaining ring slightly larger than source container diameter; maximum container diameter approximately 19.6 inches. All models
El conducto de residuos requiere bolsas de residuos peligrosos de un mínimo de 8-1/2" de ancho x 11" de largo. p.23 ↗ Procure bags meeting minimum dimensions for proper waste chute use. All models
Dimensiones interiores: 30" de profundidad y 32" de altura. p.22 ↗ Verify equipment fits within these interior clearances. All XPert Bulk Powder Enclosures
Mantener el nivel de humedad entre 45 y 60% para minimizar problemas electrostáticos. p.9 ↗ HVAC or humidification may be required to sustain this range for accurate weighing and reduced powder attraction. All models
Dejar 6-12" alrededor de la cabina para un flujo de aire adecuado. p.12 ↗ Maintain clearance from walls and other equipment during installation. All models
Clasificación de contención de fábrica tal como se fabrica inferior a 0.05 ppm a 4 L/min según ASHRAE 110-1995. p.20 ↗ Baseline containment performance; consult Labconco for individual ratings. All XPert Bulk Powder Enclosures
Permitir que la cabina funcione durante 1 minuto antes de iniciar los procedimientos; esperar 15 segundos después de la carga; operar durante dos minutos sin perturbaciones al finalizar. p.25 ↗ Follow these purge times to ensure airborne contaminants are cleared before and after work. All models
Elevar objetos grandes al menos ¼" por encima de la superficie de trabajo para permitir el flujo de aire por debajo. p.27 ↗ Use stands or supports to allow air to sweep under equipment and into rear baffle slots. All models
La cabina superior con superficie de trabajo adjunta pesa de 100 a 550 lbs (45 a 248 kg); los soportes elevadores pesan de 380 a 500 lbs (171 a 225 kg). p.13 ↗ Use hydraulic lift, electric lift, or forklift for moving; do not lift by front or side air foils. All models
Herrajes de fijación de la superficie de trabajo al soporte elevador: 1880740 Perno hexagonal ¼-20 x 2.5 (cant. 8); 1880708 Tornillo de cabeza hexagonal ¼-20 x 0.5 (cant. 3). p.15 ↗ Use specified bolts and screws for secure assembly; long bolts pass through 2" sq. tubing. All models
Adaptador de transición de 6" 3912401 recomendado para escape del edificio en todas las cabinas de 5', 6' y 8'; Adaptador de transición de 5" 3912400 requerido para Extractores Portátiles FilterMate. p.17 ↗ Order correct adapter for exhaust configuration. All XPert Bulk Powder Enclosures

Faults & Troubleshooting

Code / Symptôme Possible Cause Remedy Note
Contamination outside of enclosure p.39 ↗ Improper user techniques for the enclosure. See “Using your Enclosure” Chapter 5 and “Safety Precautions” Chapter 4 sections in the manual. Chapter 8
Contamination outside of enclosure p.39 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 8
Contamination outside of enclosure p.39 ↗ External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. Chapter 8
Contamination outside of enclosure p.39 ↗ Enclosure has improper face velocity. Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures. Chapter 8
Blower won’t operate p.39 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 8
Blower won’t operate p.39 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 8
Blower won’t operate p.39 ↗ Blower wiring is disconnected. Inspect blower wiring. Chapter 8
Blower won’t operate p.39 ↗ Blower switch is defective. Replace switch. Chapter 8
Blower won’t operate p.39 ↗ Motorized impeller or blower is defective. Replace motorized impeller or blower. Chapter 8
Low face velocity or poor containment of contaminants p.39 ↗ Enclosure sash not closed. Close sash to the lowest position. Chapter 8
Low face velocity or poor containment of contaminants p.39 ↗ HEPA filter clogged. Replace HEPA filter or increase speed. Chapter 8
Low face velocity or poor containment of contaminants p.39 ↗ Carbon filter loaded with chemicals. Replace carbon filter. Chapter 8
Blower and lights won’t operate p.39 ↗ Unit not plugged into outlet. Plug enclosure into appropriate electrical service. Chapter 8
Blower and lights won’t operate p.39 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Lights do not work p.40 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Lights do not work p.40 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Lights do not work p.40 ↗ Defective lamp. Replace lamp. Chapter 8
Lights do not work p.40 ↗ Light switch is defective. Replace light switch. Chapter 8
Lights do not work p.40 ↗ Defective electronic ballast. Replace ballast. Chapter 8
Airflow monitor malfunction: No power. No lights. No display. p.40 ↗ Power supply is not plugged into proper voltage; plug in power supply. Verify that airflow monitor interface cables are connected. Check fuses on FilterMate or accessory light. Plug in power supply; verify interface cables; check fuses. Chapter 8
Airflow monitor malfunction: No audible alarm. p.40 ↗ Alarm has been temporarily silenced using “test/reset” or “enter” buttons. Press “test/reset” or “enter” to re-enable audible alarm. Chapter 8
Airflow monitor malfunction: Wrong alarm set point. p.40 ↗ Airflow monitor was not properly adjusted. Repeat calibration steps outlined in the User’s Manual. Chapter 8
Airflow monitor malfunction: Constant audible alarm. p.40 ↗ Check airflow and calibration of airflow monitor. Check airflow and calibration of airflow monitor. Chapter 8
Airflow monitor malfunction: Continuous alarm. p.40 ↗ Check the face velocity of the enclosure as the airflow of the system may have changed. If face velocity is correct, calibrate the airflow monitor as outlined in the User’s Manual. Check face velocity; calibrate airflow monitor. Chapter 8
Airflow monitor malfunction: Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.40 ↗ Air way to airflow monitor sensor is blocked by insects, dust or debris. Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. Chapter 8
Airflow monitor malfunction: Audible disable will not stay ON. p.40 ↗ An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. Chapter 8
Lift won’t operate or malfunctions: No power. p.40 ↗ Lift motor is not connected to a proper outlet. Connect the plug into the four-plex electrical box located under work surface. Chapter 8
Lift won’t operate or malfunctions: Intermittent operation. p.40 ↗ Thermal override activated. Allow lift system’s thermal override to cool for 10 minutes then adjust height. Chapter 8
Lift won’t operate or malfunctions: Photoelectric eye malfunctions. p.40 ↗ Photoelectric eye malfunction. Either adjust photoelectric eye sensitivity or replace photoelectrical eye if it malfunctions. See replacement parts. Chapter 8
Lift won’t operate or malfunctions: Hydraulic lines pinched. p.40 ↗ Hydraulic lines pinched. Replace entire lift system. See replacement parts. Chapter 8