Users manual capture bt fuming chamber
Technical Reference Summary
Covered Models & Document Scope
This manual covers the Labconco CApture BT Fuming Chamber, catalog numbers 3170000 (115 V, 60 Hz, 12 A, 1 phase) and 3170001, -02, -03, -04, -05 (230 V, 50/60 Hz, 6 A, 1 phase). Accessories include the CApture BT Stand, hanging rods, perforated/wire/half-wire shelves, casing holder, long gun holder, large clips, mini bag clips, security tags, and exhaust kit.
Section-by-Section Technical Breakdown
The chamber must be installed on a level, solid surface capable of supporting 500 lbs (227 kg). Minimum clearances are specified for top, right side, rear, and front to allow service access and unobstructed door swing.
- Surface must support 500 lbs (227 kg).
- Top clearance: at least 6 inches (150 mm).
- Right side clearance: 12 inches (300 mm) for service.
- Rear clearance: 2 inches (50 mm).
- Front clearance: 36 inches (914 mm) for door swing.
- Leg levelers are provided if not mounted on accessory stand.
A dedicated electrical outlet with appropriate circuit breaker must be located within 10 feet (3 m) of the installation site. The outlet must provide a protective earthing ground. Catalog number 3170000 requires 115 V, 60 Hz, 12 A; 3170001 through -05 require 230 V, 50/60 Hz, 6 A.
- 3170000: 115 V, 60 Hz, 12 A, 1 phase; typical operating current 4 A.
- 3170001, -02, -03, -04, -05: 230 V, 50/60 Hz, 6 A, 1 phase; typical operating current 2 A.
- Outlet must be within 10 feet (3 m) of the product.
- Dedicated outlet must provide protective earthing ground.
The CApture BT Fuming Chamber does not require exhaust ducting. Chamber air is filtered through a pre-filter and main carbon filter to remove cyanoacrylate vapors, and clean air is recirculated into the room.
- No external exhaust ducting required.
- Air passes through pre-filter and main carbon filter.
- Filtered air is recirculated back into the room.
The chamber is shipped on a pallet secured by four bolts. Lift brackets and 2x4 boards can be used for lifting. The main carbon filter is shipped sealed in protective plastic and must be unsealed before use. Leveling legs are installed if not using the accessory stand.
- Remove four pallet bolts using a 1/2" wrench.
- Install four leveling legs into the holes left by shipping bolts.
- Chamber is heavier on the right side; use at least two additional persons for tipping.
- Remove protective plastic from main carbon filter before use; failure results in CA fumes not exhausting.
- Main carbon filter gasket must face up when reinstalled.
- Remove lift brackets and protective spacers after positioning.
After electrical connection, turning on the power switch should display the CApture BT logo, then the Main Menu, with a single beep and interior light on. Pre-run steps include filling the water bottle with RO filtered or distilled water and verifying the valve is in NORMAL FILL (down) position.
- Power switch on: display shows logo then Main Menu, single beep, interior light on.
- Fill water bottle with approximately 500 mL of RO filtered or distilled water.
- Verify valve behind water bottle is flipped down (NORMAL FILL).
- Water pump runs approximately 60-90 seconds, removing one-third to one-half of water from bottle.
The chamber includes a main carbon pre-filter, main carbon filter, independent recirculation and exhaust blowers, LED light, hot plate, and humidifier. The hot plate is user-controllable from 100 °F (38 °C) to 425 °F (218 °C). The humidifier can raise relative humidity to a maximum of 80%.
- Main carbon pre-filter removes CA vapor before main carbon filter.
- Main carbon filter uses granular activated carbon.
- Recirculation and exhaust blowers operate independently.
- LED light stays on during programs and while main door is open; goes off after 15 minutes of inactivity.
- Hot plate range: 100 °F (38 °C) to 425 °F (218 °C).
- Humidifier maximum relative humidity: 80%, displayed in 1% increments.
- Humidity can only be increased from ambient; chamber cannot decrease humidity.
The Settings menu allows configuration of clock (12/24-hour), temperature units (°F/°C), standby purge (5 minutes every 2 hours, or night mode 7pm-5am), post purge time (0, 5, or 10 minutes), keypad beep mute, diagnostic mode, and filter reorder information.
- Standby Purge: purges chamber for 5 minutes every 2 hours on even hours when power switch is on; will not start if a program is running.
- Standby Purge NIGHT setting enables function only between 7pm and 5am.
- Post Purge Time: 0, 5, or 10 minutes after main door is opened following a fuming cycle.
- Closing main door turns blowers off.
Program parameters include Grams CA (0.0-99.9), Humidity (20-80%), RH Dwell (0:00-99:30 in 30-second increments), CA Heat (100-425 °F / 38-218 °C), Fume time (0:00-99:30 in 30-second increments), and Purge time (4:00-10:00 in 30-second increments). Programs 1-3 are factory defaults that revert after modification; Programs 4-20 save changes.
- Grams CA: 0.0 to 99.9, user reference only.
- Humidity: 20% to 80%, set in 1% increments.
- RH Dwell: 0:00 to 99:30, set in 30-second increments; 0:00 means fuming begins immediately after humidity reached.
- CA Heat: 100 °F (38 °C) to 425 °F (218 °C); CA volatizes quickly around 170 °F (77 °C).
- Fume time: 0:00 to 99:30, set in 30-second increments; time begins when hot plate turns on.
- Purge time: 4:00 to 10:00, set in 30-second increments; chamber purges completely in 4 minutes under ideal conditions.
- Programs 1-3 factory defaults: Program 1 (New Prints, Faster Run Time): 2.5 g CA, 80% RH, 0:00 RH Dwell, 350 °F (177 °C) CA Heat, 10:00 Fume, 5:00 Purge.
- Program 2 (Older Prints, More Dwell Time): 2.5 g CA, 80% RH, 2:00 RH Dwell, 250 °F (121 °C) CA Heat, 18:00 Fume, 5:00 Purge.
- Program 3 (Dye Staining, Lumicyano): 2.7 g CA, 75% RH, 0:00 RH Dwell, 250 °F (121 °C) CA Heat, 17:00 Fume, 5:00 Purge.
- Low viscosity cyanoacrylate recommended; Arrowhead Forensics A-2601 CA used to develop Programs 1 and 2.
The fuming cycle consists of Humidity, optional RH Dwell, Fuming, and Purge stages. During Fuming, both main and CA chamber doors lock. The hot plate maintains set point within 10 °F (5 °C). Pressing STOP or PURGE during Fuming initiates a 5-minute purge.
- Humidity increases approximately 2%-3% per minute; maximum 80% RH reached in less than 35 minutes under normal conditions.
- If desired humidity not reached within pre-set time, program stops and displays error screen.
- During RH Dwell, do not open main door or CA chamber door.
- During Fuming, both doors lock and remain locked until program complete; if power outage occurs, doors remain locked.
- Hot plate maintains set point within 10 °F (5 °C).
- Pressing STOP or PURGE during Fuming initiates 5-minute purge.
- Pressing STOP before Fuming stage begins stops unit without purging.
- Time can be added to Fuming stage only before stage ends; once Purge begins, no more time can be added.
- If hot plate is above 125 °F (52 °C), a Fume Time adjustment recommendation is displayed.
The Humidify cycle can rehydrate prints without running a fuming cycle. Humidity level can only be increased from ambient. After the Humidify cycle, an audible beep and flashing chamber light occur for 30 minutes or until main door is opened.
- Humidity increases approximately 2%-3% per minute; maximum 80% RH reached in less than 35 minutes.
- If desired humidity not reached within pre-set time, program stops and displays error screen.
- After Humidify cycle, audible beep and flashing chamber light for 30 minutes or until main door opened.
The manual provides a recommended maintenance schedule including weekly draining of humidifier tank, monthly inspection/replacement of pre-filter, cleaning of internal surfaces, vacuuming of recirculation blower intake, annual replacement of main carbon filter, and quarterly recalibration of humidity sensor.
- Drain humidifier tank weekly.
- Inspect/replace pre-filter monthly.
- Clean internal surfaces monthly.
- Vacuum/clean recirculation blower intake monthly.
- Replace main carbon filter annually or when prompted.
- Recalibrate humidity sensor quarterly.
Filter life cycles can be set between 200 and 500 cycles in 50-cycle increments via the SERVICE menu. The appropriate number depends on average cycles per day and average grams of CA per cycle, with additional factors reducing filter life.
- Filter life cycles adjustable 200-500 in 50-cycle increments.
- Service password: [UP] [DOWN] [LEFT] [RIGHT] [OK].
- Filter life chart: for <1.5 g CA/cycle: 500 cycles (1-2 cycles/day), 400 (3-4/day), 350 (5+/day); for 1.5-2.5 g: 400, 350, 300; for 2.5+ g: 350, 300, 250.
- Factors reducing filter life: not regularly cleaning CA deposits, not replacing pre-filter regularly, enabling Standby Purge.
Diagnostic mode allows individual testing of recirculation blower, purge blower, humidify components, hot plate, and door locks. Each component activates for 30 seconds when doors are closed.
- Recirculation Blower test: alternates between slower and faster speeds for 15 seconds each.
- Purge Blower test: run twice in succession (no more than 5 seconds between tests) to test exhaust vent valve, which should open by end of second test.
- Humidify test: recirculation blower, humidifier fan, and humidifier valve activate (moderate click).
- Hotplate test: hot plate heats up; do not touch.
- Door Locks test: both main and CA chamber doors lock and stay locked while active.
If the user-selected humidity is not reached within the default 35-minute timeout, an error occurs and the program ends. The timeout can be adjusted in 5-minute increments via the SERVICE menu.
- Default Humidify Timeout: 35 minutes.
- Adjust via SERVICE > RH TIMEOUT; press [RIGHT] to increase in 5-minute increments, [OK] to accept.
- Before changing timeout, ensure no other problems exist.
Serviceable parts are located behind the right side panel. Always unplug power cord before servicing. Wear a static dissipative wrist band when handling electrical components.
- Remove two screws at top of right side panel; pivot top away, then lift to clear bottom pins.
- Side panel is heavy; lay flat while removed.
- Wear static dissipative wrist band grounded to chamber frame when replacing electrical components.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | S'applique à |
|---|---|---|
| Surface must support 500 lbs (227 kg). p.9 ↗ | Ensure floor or bench can safely bear the chamber's weight plus contents. | All models |
| Minimum clearances: 6 inches (150 mm) top, 12 inches (300 mm) right side, 2 inches (50 mm) rear, 36 inches (914 mm) front. p.9 ↗ | Plan installation space to allow service access and unobstructed door swing. | All models |
| Electrical: 3170000 requires 115 V, 60 Hz, 12 A, 1 phase; 3170001 through -05 require 230 V, 50/60 Hz, 6 A, 1 phase. p.9 ↗ | Verify dedicated circuit and outlet match the model's electrical requirements. | Model-specific |
| Outlet must be within 10 feet (3 m) of the product and provide protective earthing ground. p.9 ↗ | Locate outlet accordingly and ensure proper grounding for safe operation. | All models |
| Maximum load on internal floor: 50 lbs (23 kg). p.15 ↗ | Do not exceed this limit to avoid damage to perforated floor and supports, and reduce overturning risk. | All models |
| Maximum weight limits: Hanging Bar 15 lbs (6.8 kg), Drawer 10 lbs (4.5 kg), Full Shelf 25 lbs (11.3 kg), Half Shelf 15 lbs (6.8 kg). p.15 ↗ | Distribute evidence weight according to accessory limits. | All models |
| Hot plate temperature range: 100 °F (38 °C) to 425 °F (218 °C). p.27 ↗ | Set CA Heat parameter within this range; CA volatizes quickly around 170 °F (77 °C). | All models |
| Humidifier maximum relative humidity: 80%, displayed in 1% increments. p.27 ↗ | Humidity can only be increased from ambient; chamber cannot decrease humidity. | All models |
| Humidity increases approximately 2%-3% per minute; maximum 80% RH reached in less than 35 minutes under normal conditions. p.42 ↗ | If humidity not reached within timeout, program stops and error displays; adjust timeout if low ambient humidity. | All models |
| Hot plate maintains set point within 10 °F (5 °C) during Fuming stage. p.43 ↗ | Temperature control is tight; ensure proper calibration if deviations occur. | All models |
| Pressing STOP or PURGE during Fuming initiates a 5-minute purge. p.43 ↗ | Use to remove CA fumes if evidence over-processes. | All models |
| If hot plate is above 125 °F (52 °C), a Fume Time adjustment recommendation is displayed. p.39 ↗ | When running successive cycles, reduce fume time based on elevated hot plate temperature. | All models |
| Filter life cycles adjustable between 200 and 500 cycles in 50-cycle increments. p.18 ↗ | Set according to usage chart to ensure timely filter replacement. | All models |
| Default Humidify Timeout: 35 minutes; adjustable in 5-minute increments. p.80 ↗ | Increase timeout in low-humidity environments if needed. | All models |
| Water pump runs approximately 60-90 seconds, removing one-third to one-half of water from bottle. p.24 ↗ | Normal fill operation; verify valve is in NORMAL FILL (down) position. | All models |
| LED light goes off after 15 minutes of inactivity. p.27 ↗ | Normal behavior; opening door or pressing button wakes chamber. | All models |
| Standby Purge: purges chamber for 5 minutes every 2 hours on even hours when power switch is on; will not start if a program is running. p.33 ↗ | Reduces CA fume build-up; can be set to NIGHT mode (7pm-5am). | All models |
| Post Purge Time settings: 0, 5, or 10 minutes after main door is opened following a fuming cycle. p.33 ↗ | Keeps blowers on to pull lingering CA vapors away from user; closing main door turns blowers off. | All models |
Faults & Troubleshooting
| Code / Symptôme | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber Display will not turn on p.86 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |
| Chamber will not achieve humidity p.87 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |
| Chamber does not exhaust CA fumes p.88 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |
| Door alarm active when door(s) closed p.89 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |
| Hot Plate will not achieve temperature set point p.90 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |
| White CA residue builds on inside of Chamber p.91 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |
| Heavy Background CA on evidence p.92 ↗ | Not specified in manual excerpt. | Not specified in manual excerpt. | Section 10: Troubleshooting |