System introduction, fluidics, components, workstation, fluidics cart
Procedurepp. 51-56
Space requirements, instrument installation, workstation and cart connections
Procedurepp. 57-70
Power-up, configuration, QC test, and shutdown procedures
Procedurepp. 71-104
Experiment creation, acquisition parameters, compensation, data analysis
Procedurepp. 105-134
Preventative and unscheduled maintenance, cleaning, priming, filter care
Procedurepp. 135-138
Troubleshooting tables for common instrument issues
pp. 139-140
Document revision history
Specpp. 141-146
Laser and detection channel configuration appendix
Specpp. 147-150
Instrument, power adapter, workstation, and monitor specifications
p. 141table, 8 rows
p. 147table, 19 rows
p. 143table, 10 rows
p. 150table, 6 rows
p. 113table, 6 rows
p. 126table, 9 rows
p. 148table, 19 rows
p. 149table, 17 rows
Table 15 Laser Configuration
Model Number
405 nm
488 nm
561 nm
640 nm
1000/1040/1050
√
1000Y
√
2000R/2060R/2070R
√
√
2070V/2090V/2110V
√
√
2100YB
√
√
√
3000/3005
√
√
√
3000VYB
√
√
√
3000RYB
√
√
√
√
Table 21 NovoCyte Technical Specifications1
Parameter
Specification
Optics
Laser
Solid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality.
Laser Power
20 or 60 mW for 488 nm laser, 50 mW for 405 nm and 561 nm lasers, 40 mW for 640 nm laser
Laser Beam Profile
10 x 80 µm elliptical beam.
Optical Alignment Procedure
Fixed, no operator alignment required
Fluorescence Detection
Photomultiplier Tube (PMT)
FSC/SSC Sensitivity
FSC: 0.5 µm; SSC: 0.2 µm
Fluorescence Resolution
<3% CV for CEN
Optical Filters
User Exchangeable
Cell Size
0.2 - 50 µm
Fluidics
Flow Cell
170 x 290 µm rectangular quartz flow cell
Sample Acquisition Rate
35,000 events/second
Sample Delivery
Positive-displacement syringe pump enabling direct absolute count without the need for reference counting beads.
Volumetric Absolute Count Precision
< 5%
Sample Flow Rate
5-120 µL/min, continuously adjustable
Sheath Flow Rate
6.5 mL/min
Sample Aspiration Volume
5 µL – 5 mL
Compatibility to Autosampler
No fluidic tubing disassembly or re-connection required.
1 Сноска в тексте страницы не приведена.
Table 17 Detection Channel Configuration1
2090V-Config 2
2090V-Config 3
2110V-Default
2110V-Config 1
2110V-Config 2
2100YB-Default
3000-Default
3000-Config 1
3000VYB-Default
3000VYB-Config 1
3000VYB-Config 2
V
V
V
V
V
V
V
V
V
V
V
V
V
VB
VB
VB
VB
VB
B
VB
VB
VB
VB
VB
B
B
B
BV
BV
BV
VB
VB
VB
VB
VB
VB
BV
BV
BY
VB
VB
VBY
VBY
VBY
BY
BY
BY
VB
VB
VB
VB
VB
BY
VB
VB
VBY
VBY
BY
BY
BYR
VB
BY
VBR
VBR
VBY
VBY
BYR
BYR
BYR
BVR
BVR
BVR
VB
VB
VB
VBY
BYR
BYR
BYR
VB
VB
VB
BY
VB
VB
VBR
VBR
VY
VY
VY
YR
YR
YR
BVR
BVR
BVR
VB
Y
BVR
BVR
BVR
1V: 405 nm Laser; B: 488 nm Laser; Y: 561 nm Laser; R: 640 nm Laser. Different models of NovoCyte system use different lasers for FSC and SSC detection.
Table 24 NovoCyte Fluidics Station Specifications
Parameter
Specification
Dimensions (with reagents containers) (W × D × H)
11.4 × 9.8 × 16.5 in (29.0 × 25.0 × 42.0 cm)
Weight
13.2 lb (6.0 kg)
NovoFlow sheath fluid container
3 L
NovoClean solution container
500 mL
NovoRinse solution container
500 mL
Waste container
3 L
Table 11 List of Fluidic System Consumables to be Replaced
NovoCyte V1 Fluidics System
NovoCyte V2 Fluidics System
NovoFlow sheath fluid in-line filter (2 pcs)
NovoFlow sheath fluid in-line filter (1 pcs)
Waste filter (1 pcs)
Waste filter (1pcs)
NovoFlow sheath fluid inlet filter (1 pcs)
NovoFlow sheath fluid inlet filter (1 pcs)
NovoRinse solution inlet filter (1 pcs)
NovoRinse solution inlet filter (1 pcs)
NovoClean solution inlet filter (1 pcs)
NovoClean solution inlet filter (1 pcs)
Sheath Pump Tubing (1 pcs)
Table 12 Unscheduled Maintenance and When to Perform
Unscheduled Maintenance
When to Perform
Priming
• The CV of the data is larger than normal. • QC test failed. • No events showed up after sample collection started for certain period of time. • Instrument was not used for over one week.
Clean the External Surface of the Instrument
Keep the external surface of the instrument clean.
Decontamination
When suspecting contamination in the system.
Prime Fluidics System before Long-Term Storage
No plan to run the instrument for more than two weeks.
Prime Fluidics System after Long-Term Storage
Instrument has not been used for more than two weeks.
Purge Fluidics System before Shipment
The instrument needs to be shipped.
Calibrate the NovoCyte Fluidics Station or NovoCyte Fluidics Cart
Abnormal warning about the liquid level in the reagent containers is observed.
Clean the Optical Filters
The optical filters get dirty.
Clean Fluidics System
• The CV of the data is larger than normal. • QC test failed. • Excessive debris in display.
Table 21 NovoCyte Technical Specifications1
Characteristic
Specification
Fluid Level Sensing
Active sensing using weight sensors with automated warnings when any fluid level is out of specified range.
Fluid Container Capacity
3 L sheath, 3 L waste, 500 mL cleaning, 500 mL rinse. Optional large container for sheath (15 L) and waste (15 L)
Carryover
< 0.1%
Sample Injection Probe (SIP) Rinse
Automated flying collar wash of inner and outer SIP surface after each sampling.
Fluidics System Monitoring
In-line pressure sensor monitors the pressure in real time. Automated system recovery when pressure is out of range due to clogging.
Fluidics Maintenance
Automated startup and shutdown with fluidic system cleaning. Automated user executable cleaning, debubble, rinse, unclog, priming. Automated execution of user definable maintenance sequence.
Data Acquisition Software
NovoExpress
Parameters
Height and Area for FSC, SSC and all Fluorescent Channels, Width off FSC and Time.
Dynamic Range
24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector Gain Control
0-1000 User adjustable, optimized default gain setting for each individual channel.
Compensation
Full inter-beam matrix, during or post acquisition
File Export
FCS 3.0, FCS 3.1; CSV; Batch PDF report
Data Report
Automatic report. Customizable report. Batch PDF report.
Usage Monitor
Comprehensive Transaction Log and System Log.
User Management
Administrative creation of individual user accounts and user groups with privilege control. Configurable roles for individual users. User operation time tracking.
Quality Control Instrument Daily QC
Automatic daily quality control with Agilent Flow Cytometer QC Particles. Levey-Jennings Plot.
Production QC
Each instrument is verified for assembly integrity and performance to specifications.
Sampling Manual Sample Loading
12 X 75 mm tube; 1.5 mL or 2.0 mL tube with Agilent tube holder
Automatic Sample Loading with NovoSampler (Optional)
Agilent Flow Cytometer 24 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom) microtiter plates
1 Сноска в тексте страницы не раскрыта.
Table 21 NovoCyte Technical Specifications1
Specification
Value
Automatic Sample Loading with NovoSampler Pro (Optional)
Agilent Flow Cytometer 24 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom) microtiter plates
Operating Conditions
Instrument Dimension (W × D × H)
23.6 × 17.7 × 15.4 in (60.0 × 45.0 ×39.0 cm)
29.5 × 17.7 × 15.4 in (75.0 × 45.0 ×39.0 cm) with NovoSampler
31.9 × 17.7 × 15.4 in (81.0 × 45.0 ×39.0 cm) with NovoSampler Pro
1Specification and performance claims would be based on certain specific experiment conditions.
Engineering Facts
13 facts · every claim linked to a page
Selection
The instrument is available with either V1 or V2 fluidics, which differ in pump type, filter count, and pressure sensor count, so the specific variant must be confirmed before specifying consumables or service procedures.
Determines which fluidics consumables, filters, and pressure sensors apply to the purchased unit.
Source
“NovoCyte flow cytometer is equipped with either V1 or V2 fluidics system.”
Installation
The instrument weighs more than 23 kg, so two-person lifting or mechanical assistance is required during installation and any relocation.
Affects site preparation, manpower planning, and bench/floor load during installation.
Source
“This product is large and the weight is more than 51 lb (23 kg).”
Installation
The instrument must be connected only to a three-pin socket with reliable protective ground, otherwise shock protection is compromised.
Electrical installation must provide a grounded three-pin outlet at the instrument location.
Source
“the power cord must be inserted in the three-pin socket with a protective ground connection”
Safety
The instrument is a Class I laser product with lasers fully enclosed, so no special laser work area is needed except during service by qualified Agilent personnel.
Laser safety classification determines lab signage, access control, and service restrictions.
Source
“Agilent flow cytometer is a Class I laser product according to IEC/EN 60825-1:2014.”
Operation
The fluidics station LED turns red and prohibits operation when NovoFlow is below 0.7 L, NovoRinse/NovoClean below 100 mL, or waste exceeds 3 L.
Defines minimum reagent volumes and maximum waste capacity that must be planned for uninterrupted runs.
Source
“Reagent volume is too low to run experiments (NovoFlow < 0.7 L; NovoRinse < 100 mL; NovoClean < 100 mL), or the waste container is full (> 3 L).”
Operation
The fluidics station LED turns orange when NovoFlow is between 0.7 and 0.9 L, NovoRinse/NovoClean between 100 and 150 mL, or waste between 2.8 and 3 L, warning that refill or empty is needed soon.
Sets the refill/empty thresholds that operators must monitor to avoid aborted experiments.
Source
“0.7 L < NovoFlow < 0.9 L; 100 mL < NovoRinse < 150 mL; 100 mL < NovoClean < 150 mL”
Operation
The optical configuration should be verified before each experiment and can be changed by replacing optical filters, after which a QC test with automatic photodetector gain adjustment is required.
Filter changes affect which fluorochromes can be measured and require a QC run to restore valid performance.
Source
“It is recommended for users to view and verify the current optical configuration before each experiment.”
Performance
For automatic fluorescence compensation, the default stop condition is 10,000 events or 20 µL at slow flow rate with FSC-H threshold above 100,000.
Defines the acquisition settings that must be used to obtain valid automatic compensation matrices.
Source
“Stop condition: total 10,000 events or 20μL sample volume.”
Performance
Compensation samples should be prepared at 1,000,000 to 2,000,000 per mL to ensure correct automatic fluorescence compensation.
Sample concentration directly affects the validity of the compensation matrix and downstream data quality.
Source
“it is recommended to prepare the compensation sample concentration to be 1,000,000 ~ 2,000,000 /mL.”
Maintenance
A QC test must be performed every time fluidic system consumables are replaced to verify system performance.
Makes QC test a mandatory step in the consumable replacement procedure, affecting maintenance workflow.
Source
“A QC Test must be conducted each time when fluidic system consumables are replaced in order to check the performance of the system.”
Maintenance
The SIP cleaning apparatus must be wiped with NovoClean (1X) and then with Milli-Q or 0.2 µm filtered DI water at least five times while the instrument is powered off.
Specifies the cleaning agents and repetition required to avoid carryover and SIP contamination.
Source
“Prepare a damp cloth soaked with Milli-Q or 0.2 µm filtered DI water. Wipe the SIP cleaning apparatus back and forth at least five times.”
Conformité
The instrument is Class A ISM equipment and may cause radio interference in domestic environments, so it should be installed in a commercial/industrial setting.
EMC class restricts the installation environment and may require coordination with other sensitive equipment.
Source
“This product complies with the requirement of CISPR 11, Group 1, Class A as radiation professional equipment.”
Safety
If a leak is detected, power must be cut by holding the front power button for about 6–7 seconds and Agilent support contacted immediately.
Defines the emergency shutdown action for fluid leaks that could damage electronics or create biohazard.
Source
“immediately turn off the power by pressing and holding the power button on the front of the instrument for about 6-7 seconds”