Quality Assessment of NGS

Sommaire

  1. Procedure p. 2
    Describes experimental methods used for NGS quality assessment
  2. p. 3
    Presents results and discussion of quality assessment experiments
  3. Spec p. 4
    Compares sizing and quantification approaches with data tables
  4. Overview p. 5
    Summarizes key findings and conclusions of the assessment
  5. p. 2 table, 10 rows
  6. p. 4 table, 4 rows

Table 1. Agilent automated electrophoresis instruments and assay specifications.

Assay SpecificationsBioanalyzer High Sensitivity DNAFragment Analyzer High Sensitivity NGSFragment Analyzer High Sensitivity Small FragmentTapeStation High Sensitivity D1000
Sizing Range50 - 7,000 bp100 - 6,000 bp50 - 1,500 bp35 - 1,000 bp
Typical Resolution50 - 600 bp: 10%100 - 1,000 bp: ≤5% 250 -900 bp: ≤5% 235 - 300 bp: 15%
600 - 7,000 bp: 20%1,000 - 6,000 bp: ≤10% 2900 - 1,500 bp: ≤10% 2300 - 1,000 bp: 10%
Sizing Precision5 %CV 12 %CV 12 %CV 1,35 %CV 1
Sizing Accuracy (%error)±10% 1±5% 1±5% 1,3±10% 1,4
Quantitative Precision50 - 2,000 bp: 15 %CV 115 %CV 115 %CV 115 %CV 1
2,000 -7,000 bp: 10 %CV 1
Quantitative Accuracy (%error)±20% 1±25% 1±25% 1±20% 1
Quantitative RangeFragment 5 - 500 pg/µLFragment 5 - 500 pg/µL 1,3Fragment 5 - 500 pg/µL 1,310 - 1,000 pg/µL
Smear 100 - 10,000 pg/µLSmear 50 - 5,000 pg/µL 3Smear 100 - 5,000 pg/µL 3

1 Determined using ladder as sample; 2 Determined using short array; 3 Determined using DNA samples in 1X TE buffer; 4 Sizing accuracy for analysis with electronic ladder: ±20%

Table 1

Instrument and KitAverage Size (bp) Library 1Average Size (bp) Library 2%CV Library 1%CV Library 2
Bioanalyzer High Sensitivity DNA4314700.350.43
Fragment Analyzer High Sensitivity NGS4194480.960.22
Fragment Analyzer High Sensitivity Small Fragment4294570.540.25
TapeStation High Sensitivity D10004234461.540.22

Figure 2. Sizing comparison. The average size of two NGS libraries reported by the Agilent Bioanalyzer, Agilent Fragment Analyzer, and Agilent TapeStation systems. Error bars represent standard deviation. n = 3 replicates per instrument.

Engineering Facts

11 facts · every claim linked to a page
Selection

Choose the assay kit so its sizing range covers the expected NGS smear, since the kits differ in sizing range and resolution.

Kit sizing range must match the sample smear to obtain valid sizing and quantification.

Source
“The DNA analysis kits chosen for analysis of the NGS libraries in this paper have similar specifications (Table 1), with a sizing range well-suited for the NGS smear, allowing for comparison between the instruments.”
Selection

The Bioanalyzer High Sensitivity DNA kit covers 50-7,000 bp, wider than the other three kits listed.

Defines whether a kit can size fragments outside the 200-1,000 bp smear window.

Source
“Sizing Range 50 - 7,000 bp”
Selection

The TapeStation High Sensitivity D1000 kit only sizes up to 1,000 bp, so larger fragments fall outside its range.

Prevents selecting this kit for libraries with fragments above 1,000 bp.

Source
“35 - 1,000 bp”
Performance

Quantitative accuracy differs by platform: Bioanalyzer and TapeStation are ±20% error while Fragment Analyzer kits are ±25%.

Sets the expected concentration error budget when comparing platforms.

Source
“±20% 1 ±25% 1 ±25% 1 ±20% 1”
Operation

The quantitative range depends on whether fragments or smears are measured, with smear ranges differing per kit.

Sample concentration must fall inside the applicable fragment or smear range for reliable quantification.

Source
“Fragment 5 - 500 pg/µL Fragment 5 - 500 pg/µL 1,3 Fragment 5 - 500 pg/µL 1,3 10 - 1,000 pg/µL Smear 100 - 10,000 pg/µL Smear 50 - 5,000 pg/µL 3 Smear 100 - 5,000 pg/µL 3”
Operation

Use the smear/region function to quantify only the portion of sample inside the set size window and to get its average size.

Correct region setup is required to get meaningful library concentration and size values.

Source
“The smear analysis functions allow the user to determine the quantification of only the portion of sample that lies within the smear range, as well as the average size of the sample within the range.”
Performance

Sizing precision across all instruments was better than 1.6 %CV in triplicate measurements.

Confirms sizing repeatability is adequate for library QC decisions.

Source
“The sizing across each instrument showed excellent precision, with each system displaying less than 1.6 %CV (Figure 2B).”
Performance

Bioanalyzer and Fragment Analyzer read slightly lower concentration than TapeStation, so cross-platform values are not identical.

Explains systematic offsets when transferring concentration values between platforms.

Source
“the Bioanalyzer and Fragment Analyzer consistently displayed a slightly lower value than the TapeStation (Figure 3).”
Performance

Measured quantification precision stayed under 15% CV and error versus Qubit stayed under 18%.

Gives realistic tolerance for concentration results versus a fluorometric reference.

Source
“The %CV for each library was less than 15% within each system, and the %error was below 18%.”
Selection

Because platform performance is equivalent, selection can be driven by throughput or speed needs.

Justifies choosing a platform on operational criteria rather than analytical differences.

Source
“researchers can have confidence in selecting any of the systems based on factors important to their laboratory’s needs, such as sample throughput or speed.”
Conformité

The instruments are for research use only and not for diagnostic procedures.

Restricts deployment to research settings, not clinical diagnostics.

Source
“For Research Use Only. Not for use in diagnostic procedures.”