Agilent InfinityLab Pro iQ Series User Guide

Technical Reference Summary

Covered Models & Document Scope

Covers the Agilent InfinityLab Pro iQ Series LC/SQ Mass Detector systems running OpenLab CDS 2.7 or 2.8 with 3.2 LC/MS Driver. Includes ESI, AJS (Pro iQ Plus only), APCI and MMI ion sources, the Calibrant Delivery System (CDS), and the Edwards nXR40i Dry Pump.

Section-by-Section Technical Breakdown

pp. 16
Instrument Status Indicator Lights

Table 1 maps each instrument state to the Status LED and Power Switch LED colors/blink patterns. States include BOOTING (FPGA Loaded), BOOTING (LINUX Loaded), ACQUISITION, FAULT, PUMPING DOWN, VENTING, STANDBY, Vented, IDLE, POWERING DOWN, BACKGROUND SCAN and Off.

  • BOOTING (FPGA Loaded): Status LED blinking yellow, Power Switch LED yellow
  • ACQUISITION: Status LED blinking green, Power Switch LED green
  • BOOTING (LINUX Loaded): Status LED yellow, Power Switch LED green
  • FAULT: Status LED red, Power Switch LED green
  • PUMPING DOWN: Status LED blinking yellow, Power Switch LED green
  • VENTING: Status LED blinking red, Power Switch LED yellow
  • STANDBY: Status LED yellow, Power Switch LED green
  • Vented: Status LED red, Power Switch LED yellow
  • IDLE: Status LED yellow, Power Switch LED green
  • POWERING DOWN: Status LED off, Power Switch LED blinking yellow
  • BACKGROUND SCAN: Status LED green, Power Switch LED green
  • Off: Status LED off, Power Switch LED off
pp. 17
Instrument Status Indicators in Acquisition

The OpenLab CDS Acquisition Instrument Status window indicates instrument state by color: Gray (Offline/Disconnected), Yellow (Unknown), Light Orange (Not Ready), Green (Idle), Teal (Standby), Magenta (Pre-run/Injecting), Blue (Running, Post Run), Orange (Tuning), Red (Instrument Error). In Standby the source interface stays on, drying gas remains heated and nebulizer flows are maintained, but source and ion optics voltages are off and spectra are not sent to the computer.

  • Gray: Offline or Disconnected
  • Yellow: Unknown run state
  • Light Orange: Not Ready (e.g. temperature or pressure not reached)
  • Green: Idle, ready to process samples
  • Teal: Standby/sleep state; use when changing the ion source
  • Magenta: Pre-run/Injecting
  • Blue: Running, Post Run (collecting data)
  • Orange: Tuning
  • Red: Instrument Error, cannot process samples
  • In Standby: source interface on, drying gas heated, nebulizer flows maintained; source and ion optics voltages off; no spectra sent; APCI/MMI vaporizer heaters turned down; AJS sheath gas temperature, drying gas flow and drying gas temperature reduced; spray chamber high voltages off
pp. 22-33
Tuning

Tuning adjusts MS parameters to maximize sensitivity, maintain acceptable resolution and ensure accurate mass assignments. Three tune types exist: autotune, checktune and diagnostic tune; all except diagnostic tune can be scheduled. A checktune verifies peak width and mass axis before acquisition; an autotune is only needed after preventive maintenance or if checktune fails.

  • Checktune verifies tuning mix ion masses, peak width and mass axis before data acquisition
  • Autotune runs automatically after preventive maintenance or if checktune is out of tolerance
  • Frequent tuning not required; generally not more often than monthly, at most weekly
  • Tuning mix selection: ESI and AJS use ESI-L Low Concentration Tuning Mix (G1969-85000); MMI uses MMI-L Low Concentration Tuning Mix (G1969-85020); APCI uses APCI-L Low Concentration Tuning Mix (G1969-85010)
  • Checktune can be run with ESI, MMI, APCI and AJS sources (AJS only on Pro iQ Plus)
  • If Checktune is Out of Tolerance, run an Autotune; if Autotune is Out of Tolerance, contact Agilent
  • Diagnostic Tune produces log file TCDBackup.zip retrievable from OpenLab Control Panel under Administration > Diagnostics
  • Tune control locks the SQ instrument; single sample runs and sequences cannot be started while tune has control
pp. 30-31
Scheduling a Tune

Checktune or Autotune can be scheduled to run automatically Weekly or Monthly. Weekly scheduling allows a recurrence value from 1 to 99 weeks, selected days of the week, a start date and a time. Monthly scheduling supports 'Day X of every Y month(s)' or 'The X day of every Y month(s)' with recurrence from 1 to 99.

  • Weekly recurrence value 1 to 99 weeks
  • Monthly recurrence value 1 to 99 months
  • Monthly options: Day X of every Y month(s) or The X day of every Y month(s) (First, Second, Third, Fourth, Last)
  • If the current month has fewer days than the entered day, the occurrence falls on the last day of the month
pp. 34-52
Basic Operation

Covers starting up and shutting down the instrument, preparing LC modules, switching the LC stream between waste and MS, performing checktune or autotune, viewing the System Activity Log, and setting up acquisition methods using Auto Acquire or Advanced Acquire. Also covers running single samples and sequences, including method override for parameter optimization and fraction collection trigger masses.

  • When not acquiring data, switch the LC stream away from the MS ion source to waste
  • Auto Acquire mode supports ESI source only
  • Auto Acquire allows up to 4 Scan rows and up to 500 SIM rows (500 total rows with up to 4 Scan rows for SIM/Scan)
  • Chromatogram types: TIC, Channel TIC, EIC, BPC
  • Timetable can set Diverter to To MS or To waste, and Calibrant On or Off
  • When Calibrant is off, diverter is set to MS; when Calibrant is on, diverter valve is set to waste
  • Acquisition method extension .amx; processing method extension .pmx
  • Sequence method override can optimize one parameter at a time (e.g. Gas Temperature 250, 260, 270, 280, 290)
  • Fraction target mass overrides use SQ: T1/T2/T3 Fraction target mass columns; formula or monoisotopic mass required
  • System Activity Log accessed via Activity Log in the Windows section of the ribbon
pp. 54-64
Maintenance

Maintenance is divided into routine tasks (daily/weekly) and periodic tasks. Routine tasks include daily spray chamber cleaning, weekly calibrant bottle check/fill, weekly spray chamber cleaning and weekly nebulizer cleaning. Periodic tasks include checking rough pump oil level and replacing MS45+ pump oil and filter. General maintenance covers nebulizer needle, capillary, electron multiplier horn, valve rotor seal, corona needle and Calibrant Delivery System flushing. Early Maintenance Feedback (EMF) tracks injections and days since maintenance.

  • Daily spray chamber cleaning: lower drying gas temperature to minimum; for APCI/multimode lower vaporizer heater to minimum; wait 20 minutes for cooling; use 50:50 LC/MS-grade isopropanol and water
  • Weekly calibrant bottle check: refill if tuning mix level is within a few millimeters of the intake tube end; rinse bottle once with water then twice with ACN
  • Weekly spray chamber cleaning: sonicate parts in 50:50 isopropanol/water for 5 minutes; optional 4000-grit abrasive paper for persistent contaminants
  • Weekly nebulizer steam cleaning conditions: LC pump flow 0.5 ml/min, nebulizer pressure 60 psi, drying gas 13 L/min, drying gas temperature 350 °C
  • Steam cleaning solvent mix: 30 to 70% methanol in water; overnight steam cleaning improves signal-to-noise
  • Tuning mixes are not interchangeable; changing source or inlet assembly requires changing tuning mix to avoid miscalibration
  • EMF icon turns yellow when an enabled limit is exceeded; tooltip displays cause in device pane

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication S'applique à
Operating temperature range is 15 °C to 35 °C; storage temperature range is -40 °C to 70 °C. p.2 ↗ Install the instrument in a laboratory with controlled temperature within this range to ensure reliable operation and safe storage. whole series
This guide is valid for LC/SQ Mass Detector systems running on OpenLab CDS 2.7 or 2.8 with 3.2 LC/MS Driver. p.2 ↗ Verify software version compatibility before installing or operating the system. whole series
Auto Acquire mode supports ESI source only. p.42 ↗ For AJS, APCI or MMI sources, use Advanced Acquire to set up acquisition methods. whole series
Auto Acquire allows up to 4 Scan rows and up to 500 SIM rows (500 total rows with up to 4 Scan rows for SIM/Scan). p.43 ↗ Plan acquisition methods within these row limits to avoid method creation errors. whole series
Tuning mix part numbers: ESI-L Low Concentration Tuning Mix G1969-85000 (ESI and AJS), MMI-L Low Concentration Tuning Mix G1969-85020 (MMI), APCI-L Low Concentration Tuning Mix G1969-85010 (APCI). p.24 ↗ Order the correct tuning mix for the installed source; using the wrong mix causes miscalibration and mass assignment errors. whole series
AJS source is supported only by the Pro iQ Plus instrument. p.26 ↗ Checktune with AJS is only possible on Pro iQ Plus; other models cannot use AJS. Pro iQ Plus
If Checktune is Out of Tolerance, run an Autotune; if Autotune is Out of Tolerance, contact Agilent. p.26 ↗ Follow this escalation path for tuning failures before contacting service. whole series
Frequent tuning is not required; tuning is generally not needed more often than monthly, or at most weekly. p.27 ↗ Avoid unnecessary autotunes; use checktune to confirm out-of-adjustment before retuning. whole series
Weekly steam cleaning conditions: LC pump flow 0.5 ml/min, nebulizer pressure 60 psi, drying gas 13 L/min, drying gas temperature 350 °C. p.63 ↗ Use these parameters for weekly nebulizer cleaning to reduce contamination and improve signal-to-noise. whole series
Steam cleaning solvent mix is 30 to 70% methanol in water; overnight steam cleaning improves signal-to-noise. p.63 ↗ Prepare the correct solvent mix and allow overnight cleaning for best contamination removal. whole series
Daily spray chamber cleaning requires lowering drying gas temperature to minimum and, for APCI/multimode, lowering vaporizer heater to minimum; wait 20 minutes for cooling. p.57 ↗ Prevent burns and source damage by ensuring the spray chamber is cool before cleaning. whole series
Weekly spray chamber cleaning: sonicate spray shield, capillary cap, end plates and contact ring in 50:50 LC/MS-grade isopropanol/water for 5 minutes; optional 4000-grit abrasive paper for persistent contaminants. p.60 ↗ Follow this procedure when daily cleaning is insufficient or discoloration persists. whole series
Calibrant bottle should be refilled when tuning mix level is within a few millimeters of the intake tube end; rinse bottle once with water then twice with ACN before refilling. p.58 ↗ Prevent air intake and calibration failures by maintaining adequate calibrant level. whole series
Tuning mixes are not interchangeable; failure to change tuning mix when source or inlet assembly is changed can result in miscalibration and mass assignment errors. p.59 ↗ Always switch to the correct tuning mix when changing ion sources. whole series
In Standby mode, source interface is on, drying gas remains heated, nebulizer flows are maintained, but source and ion optics voltages are off and no spectra are sent to the computer. p.17 ↗ Use Standby when changing the ion source or when the instrument is not in use to save consumables while keeping the source ready. whole series
When Calibrant is off, the diverter is set to MS; when Calibrant is on, the diverter valve is set to waste. p.46 ↗ Understand diverter logic when setting up Timetable events for in-acquisition calibration. whole series
Sequence method override can optimize one parameter at a time; example Gas Temperature values 250, 260, 270, 280, 290 °C. p.49 ↗ Use method override columns to efficiently screen parameter values without editing the method repeatedly. whole series
Fraction target mass overrides require a formula or monoisotopic mass and are recorded in the activity log. p.51 ↗ Ensure correct formula or mass entry when overriding fraction collection trigger signals. whole series
Diagnostic Tune log file is named TCDBackup.zip and is retrieved from OpenLab Control Panel under Administration > Diagnostics. p.33 ↗ Provide this log to Agilent service personnel for troubleshooting. whole series
EMF dashboard tracks number of injections and days since maintenance; EMF icon turns yellow when an enabled limit is exceeded. p.56 ↗ Set EMF limits to plan maintenance conveniently and increase instrument uptime. whole series

Faults & Troubleshooting

Code / Symptôme Possible Cause Remedy Note
Status LED = Red, Power Switch LED = Green p.16 ↗ Instrument State = FAULT Check the System Activity Log for errors to assist with troubleshooting. Table 1 Status indicator lights on instrument
Status LED = Blinking Red, Power Switch LED = Yellow p.16 ↗ Instrument State = VENTING Wait for venting to complete; check system status. Table 1 Status indicator lights on instrument
Status LED = Red, Power Switch LED = Yellow p.16 ↗ Instrument State = Vented Instrument is vented; restart or pump down as required. Table 1 Status indicator lights on instrument
Red - Instrument Error p.17 ↗ The instrument has an error and cannot process samples. View the System Activity Log for events and errors to troubleshoot. Instrument Status Indicators in Acquisition
Light Orange - Not Ready p.17 ↗ The instrument is connected but is not ready to run (due to not reaching the correct temperature or pressure required by the method, for example). Wait for the instrument to reach the required temperature or pressure conditions. Instrument Status Indicators in Acquisition
Yellow - Unknown p.17 ↗ The device is in an unknown run state. Check the System Activity Log and instrument status. Instrument Status Indicators in Acquisition
Checktune is Out of Tolerance p.26 ↗ MS tuning is out of adjustment. Run an Autotune. Calibrating the LC/SQ (Checktune)
Autotune is Out of Tolerance p.28 ↗ MS cannot be automatically adjusted to meet performance goals. Contact Agilent. Calibrating the LC/SQ (Autotune)
Error shown when selected polarity does not match the autotuned polarities of the instrument p.43 ↗ Selected polarity in acquisition method does not match the autotuned polarities. Select a polarity that matches the autotuned polarities of the instrument. Setting up an acquisition method using Auto Acquire

Technical Data

Parameter Value Condition
Operating temperature 15 to 35 °C p.2 ↗
Storage temperature -40 to 70 °C p.2 ↗