Optimizing Microplastic Characterization by LDIR: Automated versus Manual Workflows

Technical Reference Summary

Covered Models & Document Scope

Resumen técnico del sistema de imágenes químicas Agilent 8700 LDIR con el software de control Clarity; abarca el flujo de trabajo de análisis de partículas (automatizado) con un intervalo de tamaños predeterminado de 20 a 500 µm y el flujo de trabajo de análisis manual; evaluado con microesferas de látex de poliestireno trazables según NIST de 2, 5 y 10 µm sobre portaobjetos low-E (número de pieza de Agilent M7300-68010) y filtros de poliéster recubiertos de Al (número de pieza de Agilent M7300-68011)

Technical Data

Parameter Value Condition
Default particle size range (Particle Analysis workflow) 20 to 500 µm p.1 ↗ adjustable based on user methodology
Minimum detectable particle size, automated mode 5 µm p.7 ↗ 5 µm PS beads, low-E slides and filters
IR wavenumber used for imaging 1,442 cm–1 p.3 ↗
Low-E IR reflective slide dimensions 25 × 75 mm p.2 ↗ Agilent part number M7300-68010
Al-coated polyester filter diameter 25 mm p.2 ↗ Agilent part number M7300-68011
Al-coated polyester filter pore size 0.8 µm p.2 ↗ Agilent part number M7300-68011
Al-coated polyester filter coating 100/0 nm p.2 ↗ Agilent part number M7300-68011
PS latex bead sizes evaluated 2, 5, and 10 µm p.1 ↗ NIST-traceable

Operating Limits

  • Default particle size range (Particle Analysis workflow): 20 to 500 µm p.1 ↗
  • Minimum detectable particle size, automated mode, low-E slides: 2 µm p.7 ↗
  • Minimum detectable particle size, automated mode, low-E slides and filters: 5 µm p.7 ↗
  • Minimum detectable particle size, automated mode, low-E slides and filters: 10 µm p.7 ↗

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
Low-E infrared (IR) reflective slide p.2 ↗ M7300-68010 25 × 75 mm Substrate used in the study
Al-coated polyester filter p.2 ↗ M7300-68011 25 mm, 100/0 nm coating, 0.8 µm pore size Substrate used in the study