Performance Evaluation of a Reactor Sampling Device for Online LC

Technical Reference Summary

Covered Models & Document Scope

Evaluación de un dispositivo de muestreo de reactor para LC en línea utilizando un Agilent 1260 Infinity II Prime Online LC System (G3167AA) con Snapdragon Chemistry SRS y Mettler Toledo EasySampler 1210; abarca instrumentación, parámetros del método y ajustes de software/hardware del SRS.

Technical Data

Parameter Value Condition
Analytical flow rate 1.0 mL/min p.2 ↗
Solvent composition Isocratic 15% B p.2 ↗
Column temperature 45 °C p.2 ↗
Feed speed 80% of analytical flow rate p.2 ↗
Injection volume 5 µL p.2 ↗
Needle wash 3 s, Water: ACN 1:1 + 0.1% FA (S1) p.2 ↗
Diode array detector wavelength 273 ± 4 nm p.2 ↗ reference: off, 20 Hz data rate
Draw speed 100 µL/min p.2 ↗ Setting 2
Wait time 3.6 s p.2 ↗ Setting 2
Dilution 1:150, 1:200, 1:400, 1:600, 1:700 p.2 ↗
Mixing time 10 s p.2 ↗
Cleaning volume 1.6 mL p.2 ↗ 2 Wash repeats
Sampling interval 10 min p.2 ↗
Inlet pressure 30 psi p.2 ↗ on inlet of instrument
Regulator pressure 15 psi p.2 ↗ on regulator on front of instrument
Flow meter 0.1 liter/min p.2 ↗
Column dimensions 3.0 × 50 mm p.2 ↗ ZORBAX RRHD Eclipse Plus C18, 1.8 µm (p/n 959757-302)
Detector flow cell path length 10 mm p.2 ↗ InfinityLab Max-Light Cartridge Cell (G4212-60008)

Operating Limits

  • Inlet pressure of instrument: 30 psi p.2 ↗
  • Regulator pressure on front of instrument: 15 psi p.2 ↗
  • Flow meter setting: 0.1 liter/min p.2 ↗

Type Code Decoder

Position Token Meaning
5067-6680 p.2 ↗ 5067-6680 Reactor valve pod part number
G4212-60008 p.2 ↗ G4212-60008 InfinityLab Max-Light Cartridge Cell (10 mm)
959757-302 p.2 ↗ 959757-302 ZORBAX RRHD Eclipse Plus C18 column part number