Simplify Your Method Development Using the Agilent 1260 Infinity II Prime LC System and LC/MSD iQ
Technical Reference Summary
Covered Models & Document Scope
Resumen técnico que demuestra un flujo de trabajo de desarrollo de método para el kit de idoneidad del sistema de 7 analitos Agilent LC/MS utilizando el Agilent 1260 Infinity II Prime LC System acoplado al LC/MSD iQ; abarca la configuración del instrumento, las columnas, las fases móviles y los resultados del cribado (sin variantes de producto ni códigos de pedido).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Flow rate | 1 mL/min p.3 ↗ | |
| Mobile phase A | Water + 10 mM ammonium bicarbonate, pH 6.6 p.3 ↗ | |
| Column temperature screening range | 40 to 60 °C p.4 ↗ | 5 °C increments |
| Initial column temperature | 40 °C p.6 ↗ | |
| Buffer concentration screening range | 10 to 2 mM p.6 ↗ | automatically diluted with water |
| Column 1 | InfinityLab Poroshell 120 Phenyl-Hexyl, 2.1 × 50 mm, 1.9 µm p.3 ↗ | |
| Column 2 | InfinityLab Poroshell 120 PFP, 2.1 × 50 mm, 1.9 µm p.3 ↗ | |
| Software | Agilent OpenLab CDS (Version 2.4) p.3 ↗ | |
| Detector | Agilent 1260 Infinity II Diode Array Detector HS (G7117C) p.3 ↗ | |
| Pump | Agilent 1260 Infinity II Flexible Pump p.3 ↗ | |
| Autosampler | Agilent 1260 Infinity II Multisampler p.3 ↗ | |
| Column thermostat | Agilent 1260 Infinity II Multicolumn Thermostat (G4237A), standard heat exchanger p.3 ↗ | |
| Solvent selector valve capacity | 12-position/13-port p.2 ↗ | one pump channel accesses up to 12 additional solvent channels |
| Water purification | LC-Pak polisher and 0.22 μm membrane cartridge (Millipak, Merck-Millipore) p.3 ↗ | point of use |
| Compound 8-BG logP | −2.27 p.2 ↗ | |
| Compound DOP logP | 8.97 p.2 ↗ | |
| Ami monitored mass | 278.2 m/z p.4 ↗ | positive |
| Ami average retention time | 1.655 min p.5 ↗ | |
| Ami retention time RSD | 0.120 % p.5 ↗ | |
| Minimum resolution (EC-C18, Water + 10 mM AmBic pH 6.6, ACN) | 8.4 p.5 ↗ | |
| Minimum resolution sum (EC-C18, Water + 10 mM AmBic pH 6.6, ACN) | 96.1 p.5 ↗ | |
| Peak symmetry (EC-C18, Water + 10 mM AmBic pH 6.6, ACN) | 0.58 p.5 ↗ |
Operating Limits
- Column temperature screening range: 40 to 60 °C p.4 ↗
- Buffer concentration screening range: 10 to 2 mM p.6 ↗