Experimental materials and methods for quality analysis
Qualitypp. 3-4
Results, discussion, and SDS-PAGE sample preparation details
Specpp. 5-6
Sizing accuracy and precision data for protein standards and CAII
Specpp. 7-8
Concentration measurements and CAII performance results
Overviewp. 9
Summary of quality analysis findings
p. 10
Bibliographic references cited in the document
p. 2table, 16 rows
p. 5table, 11 rows
p. 6table, 11 rows
p. 7table, 10 rows
p. 8table, 10 rows
Table 1. Agilent Protein Broad Range P240 kit specifications (LM: lower marker; UM: upper marker).
Analytical Specifications
ProteoAnalyzer Protein Broad Range P240 kit
Sizing Range
LM only
10 to 240 kDa
LM and UM
10 to 200 kDa
Typical Sizing Accuracy (% Sizing Error)
LM only
< 15% for BSA, CAII
LM and UM
< 10% for BSA, CAII
Typical Resolution
< 10% molecular weight resolution between 15 to 150 kDa (based on ladder)
R≥1 NIST mAb NGHC/HC (using reduced conditions)
Sizing Precision
LM only
< 8% CV for BSA, CAII, GREMLIN-1, and NIST mAb (using reduced conditions)
< 10% CV for intact NIST mAb (using non-reduced conditions)
LM and UM
< 5% CV for BSA, CAII, GREMLIN-1 and NIST mAb (using reduced conditions)
Quantitative Range
2 ng/µL to 2,000 ng/µL for BSA in PBS
Sensitivity (Signal/Noise > 3)
1 ng/µl for BSA, CAII in PBS
Quantification Reproducibility
<15 %CV for 20 – 2,000 ng/µL BSA
<25 %CV for 2 – 20 ng/µL BSA
LM: lower marker; UM: upper marker.
BSA sizing assessment
Sample Concentration
Average Size (kDa) — ProteoAnalyzer
Average Size (kDa) — SDS-PAGE
Accuracy (% error) — ProteoAnalyzer
Accuracy (% error) — SDS-PAGE
Precision (%CV) — ProteoAnalyzer
Precision (%CV) — SDS-PAGE
2,000 ng/µL
66.7
50.0
1.01%
–24.24%
0.57%
11.14%
1,000 ng/µL
68.5
51.7
3.84%
–21.72%
0.59%
11.83%
500 ng/µL
69.0
54.0
4.49%
–18.18%
0.08%
13.98%
250 ng/µL
69.4
56.7
5.10%
–14.14%
0.46%
12.52%
125 ng/µL
69.6
58.7
5.40%
–11.11%
0.87%
9.39%
62.5 ng/µL
69.4
61.3
5.10%
–7.07%
0.46%
8.98%
31.3 ng/µL
70.4
62.7
6.62%
–5.05%
1.37%
7.20%
15.6 ng/µL
69.8
65.0
5.71%
–1.52%
0.54%
11.09%
7.8 ng/µL
69.6
66.7
5.40%
1.01%
0.79%
14.80%
3.9 ng/µL
69.6
65.7
5.40%
–0.51%
0.87%
12.31%
Overall
69.2
59.2
4.81%
–10.25%
1.53%
13.71%
Figure 3. (B) Average size, % error, and %CV for each dilution of BSA using the ProteoAnalyzer and SDS-PAGE. (n = 3)
Average size, % error, and %CV for each dilution of CAII using the ProteoAnalyzer and SDS-PAGE. (n = 3)
Sample Concentration
Average Size (kDa) — ProteoAnalyzer
Average Size (kDa) — SDS-PAGE
Accuracy (% error) — ProteoAnalyzer
Accuracy (% error) — SDS-PAGE
Precision (%CV) — ProteoAnalyzer
Precision (%CV) — SDS-PAGE
2,000 ng/µL
27.9
23
–3.68%
–20.69%
0.41%
0.00%
1,000 ng/µL
28.3
23
–2.41%
–20.69%
3.06%
0.00%
500 ng/µL
27.9
23
–3.91%
–20.69%
0.41%
0.00%
250 ng/µL
27.8
23
–4.14%
–20.69%
0.72%
0.00%
125 ng/µL
28.3
23.3
–2.53%
–19.54%
2.27%
2.47%
62.5 ng/µL
27.9
23.3
–3.91%
–19.54%
0.41%
2.47%
31.3 ng/µL
27.7
23.7
–4.37%
–18.39%
0.42%
2.44%
15.6 ng/µL
28.3
24
–2.53%
–17.24%
2.27%
0.00%
7.8 ng/µL
27.8
23.7
–4.14%
–18.39%
0.00%
2.44%
3.9 ng/µL
27.9
23.7
–3.91%
–18.39%
0.41%
2.44%
Overall
28.0
23.4
–3.55%
–19.43%
1.43%
2.10%
Figure 4B. (n = 3)
Concentration Precision (%CV) of BSA
Nominal Concentration
ProteoAnalyzer
SDS-PAGE
2,000 ng/µL
1.28%
33.00%
1,000 ng/µL
1.45%
37.65%
500 ng/µL
0.74%
18.52%
250 ng/µL
1.33%
16.71%
125 ng/µL
1.99%
12.36%
62.5 ng/µL
2.23%
31.08%
31.3 ng/µL
1.35%
12.71%
15.6 ng/µL
3.09%
9.48%
7.8 ng/µL
4.46%
3.08%
3.9 ng/µL
7.84%
16.40%
n = 3
CAII Concentration Precision (%CV)
Nominal Concentration Range
ProteoAnalyzer
SDS-PAGE
2,000 ng/µL
8.02%
11.52%
1,000 ng/µL
1.34%
5.07%
500 ng/µL
2.37%
3.90%
250 ng/µL
6.03%
6.69%
125 ng/µL
2.35%
10.56%
62.5 ng/µL
4.67%
9.49%
31.3 ng/µL
9.27%
4.39%
15.6 ng/µL
6.19%
1.13%
7.8 ng/µL
11.98%
1.58%
3.9 ng/µL
8.95%
30.67%
(n = 3)
Engineering Facts
12 facts · every claim linked to a page
Selection
The instrument uses a 12-channel capillary array and can run multiple samples in parallel with multiple programmed runs, which matters for throughput planning.
Determines batch capacity and scheduling when selecting the system for a QC workflow.
Source
“uses a 12-channel capillary array to analyze multiple samples in parallel and allows for multiple runs to be programmed at once”
Operation
Only 1 µL of sample is required, which reduces sample consumption and matters when sample material is limited.
Low sample consumption is a key selection criterion when sample is scarce or expensive.
Source
“allow for the detection of impurities while only requiring 1 µL of sample”
Performance
Sizing range is 10–240 kDa with lower marker only and 10–200 kDa with both markers, so the marker configuration must match the target protein size.
Defines whether the assay can cover the protein of interest and which marker protocol to use.
Source
“has a sizing range from 10 to 240 kDa when using only the lower marker (LM), and 10 to 200 kDa when using the LM and upper marker (UM)”
Operation
The lower marker aligns samples and ladder for sizing, and the optional upper marker improves alignment, so marker addition is an operational choice affecting accuracy.
Guides method setup and whether to enable the optional upper marker for better alignment.
Source
“A LM is used for alignment of the samples and ladder, allowing for accurate and precise sizing, along with an optional UM that can be added for improved alignment”
Performance
SDS-PAGE gels can be overloaded and produce smearing, making sizing and quantification unreliable at high concentrations.
Warns that loading amount must be controlled on slab gels to avoid invalid QC results.
Source
“the gels can be easily overloaded and thus misrepresent the size and quantification of the sample”
Performance
SDS-PAGE lacks alignment markers, so migration artifacts such as smiling can cause inaccurate sizing.
Explains a systematic error source in slab-gel sizing that the capillary system avoids.
Source
“Unlike the ProteoAnalyzer, SDS-PAGE does not use alignment markers with the samples to help with correcting migration issues”
Performance
The ProteoAnalyzer showed consistent sizing error across concentrations, while SDS-PAGE error depended on concentration.
Indicates the capillary method is more robust across a dilution range, reducing re-run risk.
Source
“the percent error achieved by the ProteoAnalyzer remained consistent despite the concentration of the sample”
Performance
The system provides a three-log quantitative dynamic range from 2 to 2,000 ng/µL, which supports low-level impurity detection.
Defines the usable concentration window and detection capability for quantification.
Source
“The system provides a three-log quantitative dynamic range, with a sample input concentration range of 2 to 2,000 ng/μL”
Performance
Coomassie staining on SDS-PAGE has a dynamic range of two logs or less, limiting low-level impurity detection compared with the capillary system.
Explains why slab-gel quantification is less suitable for trace impurity assessment.
Source
“This stain has a dynamic range of two logs or less”
Performance
The ProteoAnalyzer showed strong linearity (R² 0.9995 for BSA, 0.9998 for CAII) versus lower SDS-PAGE linearity.
Supports selection when quantitative linearity across a concentration range is required.
Source
“The ProteoAnalyzer demonstrated a calculated R2 of 0.9995 for BSA”
Performance
Quantification precision from the ProteoAnalyzer was within kit specifications and significantly better than SDS-PAGE.
Confirms the method meets stated precision limits for regulated or repetitive QC.
Source
“The %CV values determined from the ProteoAnalyzer were within the kit specifications for both proteins”
Conformité
The product is for research use only and not for diagnostic procedures, which constrains its use in regulated diagnostic workflows.
Must be checked against the intended application before selection for clinical or diagnostic use.
Source
“For Research Use Only. Not for use in diagnostic procedures.”