Let Your Autosampler Do Your Pipetting
Technical Reference Summary
Covered Models & Document Scope
Resumen técnico que describe el uso del programa de inyector del Agilent 1260 Infinity II Vialsampler (G7129C) y del 1260 Infinity II Multisampler (G7167A) para la preparación de soluciones de calibración mediante dilución seriada; abarca la configuración predeterminada de cabezal analítico de 100 µL/bucle de 100 µL y el cabezal analítico de 900 µL (opción #163)/bucle de 900 µL (opción #156) para el Multisampler, con la opción de termostatización de muestras #101; el sistema incluye la 1260 Infinity II Flexible Pump (G7104C), el Multicolumn Thermostat (G7116A) y el Diode Array Detector HS (G7117C).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analytical head volume | 100 µL p.2 ↗ | Vialsampler G7129C, default setup |
| Loop volume | 100 µL p.2 ↗ | Vialsampler G7129C, default setup |
| Sample thermostat option | #101 p.2 ↗ | Vialsampler G7129C and Multisampler G7167A |
| Diluent draw volume | 80 µL p.3 ↗ | injector program, location D1F-E1 |
| Wait time | 0.6 minutes p.3 ↗ | injector program |
| Column | Agilent InfinityLab Poroshell 120 EC-C18 p.5 ↗ | |
| Calibration concentration levels | 20, 4, 0.8, 0.16 p.2 ↗ | serial dilution |
| Caffeine stock solution concentration | 100 µg/mL p.7 ↗ | |
| Calibration function slope (Multisampler 100 µL, manual) | 27.85 p.7 ↗ | y=27.85*x+0.53 |
| Calibration function slope (Multisampler 100 µL, autosampler) | 27.64 p.7 ↗ | y=27.64*x-0.18 |
| Calibration function slope (Vialsampler 100 µL, manual) | 27.76 p.7 ↗ | y=27.76*x+0.43 |
| Calibration function slope (Vialsampler 100 µL, autosampler) | 27.65 p.7 ↗ | y=27.65*x-0.73 |
| Calibration function slope (Multisampler 900 µL, manual) | 27.69 p.7 ↗ | y=27.69*x+0.54 |
| Calibration function slope (Multisampler 900 µL, autosampler) | 27.53 p.7 ↗ | y=27.53*x+0.02 |
| Calibration function R² (Multisampler 100 µL, manual) | 0.99999 p.7 ↗ | |
| Calibration function R² (Multisampler 100 µL, autosampler) | 1.00000 p.7 ↗ | |
| Calibration function R² (Vialsampler 100 µL, manual) | 0.99999 p.7 ↗ | |
| Calibration function R² (Vialsampler 100 µL, autosampler) | 0.99999 p.7 ↗ | |
| Calibration function R² (Multisampler 900 µL, manual) | 0.99999 p.7 ↗ | |
| Calibration function R² (Multisampler 900 µL, autosampler) | 1.00000 p.7 ↗ | |
| Signal-to-noise acceptance criterion | 3:1 or 2:1 p.7 ↗ | LOD estimation |
Operating Limits
- LOD signal-to-noise ratio acceptance: 3:1 or 2:1 p.7 ↗