* (According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange and using the recommended pre-vacuum pump)
Tab. 3 Accessories and Spare Parts
DESCRIPTION
PART NUMBER
Inlet screen, DN 40
9699309
Inlet screen, DN 63
9699300
Heater band 220 V
9699801
Heater band 120 V
9699802
Metallic water cooling kit
9699823
Plastic water cooling kit
9699824
Air cooling kit (0.5 m cable)
9699290
Air cooling kit extension cable (5 m)
9699940
Pump extension cable
9699942
Vibration damper DN 63
9699375
Vibration damper CFF 4.5"
9699376
Mechanical pump DS 102
9499315
Mechanical pump DS 42
9499309
Dry scroll SH 100
SH01001 UNIV
Dry scroll SH 110
SH01101 UNIV
Rack controller 81-AG base
9698988
Rack controller 81-AG RS232/485
9698989
Rack controller 81-AG Profibus
9698990
PCB 24 V
9699538
Turbo-V 81-AG Navigator Controller 24 Vdc
9698995
Turbo-V 81-AG Navigator Controller 100-240 Vac
9698996
Tab. 4 Vent valve model selection table
CONTROL UNIT MODEL
VENT VALVE PART NUMBER
Turbo-V 81-AG Navigator Controller (cable 0.7 m) (any version)
N.O. 9699844
Turbo-V 81-AG Rack Controller (cable 0.7 m) (any version)
N.O. 9699844
Controller TV70 Navigator (P/N 9698970, 9698971)
9699834
Turbo-V 70 Rack controller (P/N 9699405, 9699505)
9699843
Tab. 1
FLANGE
FIXING DEVICE
N.
FIXING TORQUE
ISO 63
M10 clamps
4
22 Nm
Engineering Facts
14 facts · every claim linked to a page
Installation
The protective cap and adhesive must remain in place until the pump is actually connected to the vacuum system, preventing contamination during handling.
Removing the cap early exposes the pump interior to dust and moisture, which degrades ultimate vacuum and can damage bearings.
Source
“Do not remove the adhesive and protective cap before connecting the turbopump to the system.”
Operation
Operating ambient temperature is limited to +5 °C to +35 °C, and relative humidity must be 0–95 % non-condensing.
Exceeding these limits can overheat the pump or cause condensation, leading to electrical faults or bearing damage.
Source
“temperature: from +5 °C to +35 °C (see the diagram pressure-temperature in the appendix “Technical Information”)”
Installation
The pump can be mounted in any orientation but must be fixed to a counter-flange able to withstand 250 Nm torque around its axis.
Insufficient fixing torque risks the pump rotating or detaching under rotor crash torque, causing equipment damage or injury.
Source
“Fix the Turbo-V 81-T in a stable position connecting the inlet flange of the turbopump to a fixed counter-flange capable of withstanding a torque of 250 Nm around its axis.”
Installation
The Turbo-V 81-T cannot be fixed by means of its base; it must be supported via the inlet flange connection.
Attempting base mounting will not provide adequate support and may lead to vibration, misalignment, or pump damage.
Source
“The Turbo-V 81-T cannot be fixed by means of its base.”
Operation
During bakeout, inlet flange temperature must not exceed 120 °C for ConFlat or 80 °C for ISO flanges, and water cooling must always be used.
Exceeding these temperatures can damage the pump's internal components or seals; water cooling is required to protect the pump during bakeout.
Source
“While heating the vacuum chamber, the temperature of the inlet flange must not exceed 120 °C for a ConFlat flange and 80 °C for a ISO flange.”
Operation
Never operate the pump with the inlet flange open to atmosphere; it must be connected to the vacuum chamber or closed with a blank flange.
Running the pump without a sealed inlet causes overspeed, overheating, and potential rotor damage.
Source
“Never use the turbopump when the inlet flange is not connected to the vacuum chamber.”
Operation
Vent pressure at the vent port must be less than 2 bar above atmospheric pressure, and vent gas must be free of dust or particles.
Excessive vent pressure can damage the rotor or bearings; particulate contamination can degrade vacuum performance.
Source
“The pressure at the vent port must be less than 2 bar (above atmospheric pressure).”
Maintenance
Before any maintenance, disconnect mains power, vent the pump, wait for the rotor to stop, and wait until surface temperature falls below 50 °C.
Ignoring these steps risks electric shock, burns, or injury from a spinning rotor.
Source
“Before carrying out any work on the system, disconnect it from the mains, vent the pump by opening the appropriate valve, wait until the rotor has stopped turning and wait until the surface temperature of the pump falls below 50 °C.”
Selection
The pump requires a suitable Agilent controller and must be connected to a primary (backing) pump for proper operation.
Without the correct controller and backing pump, the turbopump cannot achieve its specified performance and may be damaged.
Source
“The Turbo-V 81-T pumps must be used in conjunction with one of the suitable Agilent controller and they must be connected to a primary pump (see "Technical Information").”
Selection
The inlet screen reduces pumping speed by about 10 % and protects blades from debris larger than 0.7 mm.
When selecting the pump for a process, account for this 10 % speed loss if an inlet screen is required for debris protection.
Source
“The inlet screen, however, does reduce the pumping speed by about 10 %.”
Operation
The pump must be soft-started when first received and operated by the customer.
Soft-starting prevents excessive inrush current and mechanical stress on bearings during initial run-in.
Source
“the turbomolecular pumps must be always soft-started when received and operated for the first time by the customer”
Selection
The shelf life of the turbomolecular pump is 10 months from the shipping date; exceeding this requires return to factory.
Planning installation within the shelf life avoids costly returns and ensures bearing lubricant integrity.
Source
“the shelf life of a turbomolecular pump is 10 months from the shipping date.”
Installation
The pump belongs to installation category II per EN 61010-1 and must be connected to a mains line meeting that category.
Connecting to an inadequate mains category risks electrical safety and compliance with EN 61010-1.
Source
“The Turbo-V 81-T belongs to the second installation (or overvoltage) category as per directive EN 61010-1.”
Operation
When pumping aggressive gases, a special purge port must be used to supply inert gas (nitrogen or helium) to protect the bearings.
Without inert gas purge, aggressive gases can corrode bearings and shorten pump life.
Source
“For pumping aggressive gases, these pumps are fitted with a special port to allow a steady flow of inert gas (like Nitrogen or Helium) for pump bearing protection (see the appendix "Technical Information").”