UHV-24/UHV-24p Ionization Gauge User Manual

Sommaire

  1. Warning p. 2
    Legal and safety notices for the manual
  2. p. 3
    Table of contents listing manual sections
  3. Overview pp. 4-5
    General instructions for operating the gauge
  4. Overview p. 6
    Purpose and structure of the manual
  5. Overview p. 7
    Key terms and definitions used in the manual
  6. Warning p. 8
    Explanation of safety warning symbols
  7. Warning pp. 10-13
    Safety precautions, grounding, combustibles, and disposal
  8. Services p. 14
    Service and support information
  9. Spec pp. 15-22
    Technical description, installation, and application notes
  10. Spec p. 19
    Gauge technical specifications and ratings
  11. Procedure p. 23
    Degassing procedure for the gauge
  12. Procedure pp. 24-25
    Maintenance and yttria-coated iridium filament replacement
  13. Spec p. 26
    Correction factors for various gases
  14. Ordering p. 29
    Ordering information for accessories and spare parts
  15. p. 35
    Customer care contact information
  16. p. 19 table, 6 rows
  17. p. 15 table, 12 rows
  18. p. 29 table, 2 rows

Table 1 Technical Specifications

ParameterSpecification
Filament Voltage3 to 4.5 V (AC or DC) - Varies with pressure and emission current
Filament Current2.5 to 3.5 A - Varies with pressure and emission current
Element VoltagesCollector voltage: Vc = 0 Filament voltage: Vfil = 28 VDC Grid voltage, Vg = +180 VDC
Degas Power Requirements (40 W)Grid Voltage: +600 VDC max Emission Current (total): 67 mA DC Filament Voltage: 7 V
Measurement SpecificationsSensitivity: • 25/Torr (UHV-24), nominal • 20/Torr (UHV-24p), nominal Emission Current: • 4 mA for widest measurement range 5x10-10 Torr to 1x10-4 Torr • 10 mA for pressures lower than 5x10-10 Torr ≤ 0.1 mA for pressure over 1x10-4 Torr
MaterialsFilament: Tungsten or yttria-coated iridium filaments Filament Supports: SST Collector - tungsten: • 0.007" diameter for UHV-24 • 0.002" for UHV-24p Grid: SST Feedthrough: SST, Alumina, Nickel Alloy

Technical Information

РазделСтраница
Description of the UHV-24/UHV-24p Ionization Gauge16
Operation16
Installation18
Technical Specifications19
Application Notes21
Degas23
Bakeout23
Maintenance: Filament Replacement24
Tungsten Filament Replacement24
Yttria-coated iridium filament Replacement25
Gas Correction Factors26
Accessories and spare parts29

Table 3 Accessories and spare parts

Part NumberDescription
X3004-64403Filament replacement kit (yttria-iridium)
9710018UHV-24 Filament Replacement kit, Tungsten

Engineering Facts

14 facts · every claim linked to a page
Selection

The UHV-24 covers 1 mTorr to 2x10-10 Torr while the UHV-24p extends the low end to 5x10-11 Torr, so choose the p-version for measurements below 2x10-10 Torr.

Defines the usable measurement range and where accuracy degrades, which drives gauge model selection.

Source
“provides reliable pressure measurement from 1 mTorr down to 2x10-10 Torr, with reduced performance at pressures lower than 2x10-10 Torr.”
Selection

The UHV-24p uses a thinner collector that lowers the x-ray limit to 5x10-12 Torr, enabling measurement down to 5x10-11 Torr.

The x-ray limit is the hard floor of UHV measurement, so it determines whether the gauge can reach the target base pressure.

Source
“utilizes an extremely thin collector which lowers its x-ray limit to 5x10-12 Torr.”
Installation

Mount the nude gauge in a relatively open space to reduce wall outgassing caused by localized filament heating.

Mounting location directly affects outgassing-induced pressure reading errors.

Source
“Install the gauge in a relatively open space to reduce wall outgassing due to localized heating by the filament.”
Installation

When mounted horizontally, place the filaments to the side of the grid so small vertical movements do not change the grid-to-filament spacing.

Orientation affects sensitivity stability and measurement repeatability.

Source
“position the filaments to the side of the grid rather than below or above it so that a slight vertical movement of the grid or filament does not change their relative spacing.”
Safety

The gauge controller and vacuum system must be separately grounded to a common ground, and the ground must support at least 10 A.

Improper grounding can expose personnel to 160-900 V and cause lethal shock.

Source
“Be certain that your UHV-24/UHV-24p Ion Gauge Controller and vacuum system are separately grounded to a common ground.”
Safety

Do not run a ground wire between the vacuum chamber and the controller chassis because large continuous currents could flow through it.

Prevents hazardous ground-loop currents and equipment damage during installation.

Source
“Do not place a ground wire between the vacuum chamber and the controller chassis; large continuous currents could flow through it.”
Safety

The gauge is not intrinsically safe; do not turn it on when combustible gases or mixtures are present because the filament and high-voltage discharge can ignite them.

Critical for process safety when pumping or backfilling flammable gas mixtures.

Source
“As with all Ionization gauges, this device is not intrinsically safe.”
Maintenance

Do not use silicone oil or silicone grease on vacuum equipment; use powder-free butyl or polycarbonate gloves and avoid Alconox on aluminum.

Silicone contamination ruins UHV performance and Alconox damages aluminum parts.

Source
“Do not use silicone oil or silicone grease.”
Operation

A +450 °C bakeout is the safe maximum for repeated or extended bakeout of the nude gauge; do not exceed it or apply thermal shock.

Exceeding the bakeout limit or thermal shock can destroy the gauge during chamber bake.

Source
“A temperature of +450 °C is a safe maximum for repeated or extended bakeout of the nude gauge.”
Operation

Degassing is generally not needed unless the goal is to reach pressures below 1x10-8 Torr, and a ~25 minute e-beam degas is sufficient.

Avoids unnecessary filament wear and chamber heating when lower base pressure is not required.

Source
“Gauge degassing is not generally needed unless the goal is to reach pressures below 1x10-8 Torr.”
Maintenance

Handle the gauge only by its ceramic base and wear nylon gloves, because fingerprint oil extends pumpdown time.

Prevents wire damage and contamination that degrade achievable vacuum.

Source
“Handle the gauge only by its ceramic base to prevent damage to the delicate wires.”
Maintenance

Set the long filament support bracket so the filament sits 0.050 to 0.070 inch from the grid structure.

Correct filament-to-grid spacing is required for proper sensitivity after replacement.

Source
“Insert the long filament support bracket into its collar so that the filament is located 0.050 to 0.070" from the grid structure.”
Performance

Long cables (100 ft or more) cause leakage currents and noise pickup; PTFE is the recommended insulator and ferrite beads may help.

Cable length and routing directly limit the lowest measurable ion current and reading accuracy.

Source
“Leakage currents are extremely difficult to avoid when using cables of 100 ft or longer.”
Performance

A grounded conductive wall near the tube can raise sensitivity by as much as 30%, typical of gauges mounted in tubulations.

Mounting geometry changes calibration, so installation position must be considered for accurate readings.

Source
“A grounded conductive wall near the tube raises the sensitivity by as much as 30%.”