Eppendorf epMotion ® 96 Flex - 96-Well Liquid Handler
Eppendorf ·Fabriqué en Germany · En production active

Eppendorf epMotion ® 96 Flex - 96-Well Liquid Handler

Eppendorf epMotion ® 96 Flex - 96-Well Liquid Handler — Systèmes de pipetage automatisés.

The epMotion 96 Flex is an electronic air-cushion liquid handler for 96-well plates, with two SLAS/ANSI positions and a 0.5–1 µL pipetting range. It is the 96-well modification of the epMotion series, distinct from the 5070/5073/5075 single-channel platforms. It serves cell culture and molecular biology workflows requiring automated low-volume plate pipetting.

Télécharger la fiche (PDF)

Spécifications complètes

Prestaciones y funciones principales
Application Cell culture, Molecular biology
Autoclavable no
Consommables epT.I.P.S. ® Motion Reload System
Dimensions (l × P × H) 34,5 x 41,5 x 62 cm
Mode de fonctionnement electronic
Type de pipetage Air-cushion system
Volume de pipetage 0.5...1 µL
SLAS/ANSI positions mfr 2
Dimensiones, peso y diseño mecánico
Poids du produit 19 kg
Eléctrico, alimentación y accionamiento
Consommation électrique max. 175 W
Tension 100 – 240 V

Description

Construction and positioning

Behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Size the mains circuit for 175 W maximum draw

The instrument accepts 100–240 V and draws up to 175 W. Provide a grounded outlet on a circuit rated for that load and avoid sharing it with high-inrush equipment; the wide input range means the same unit can be deployed on 120 V or 230 V sites without a voltage-specific variant.

Factory Specification Sheet
Process Limits
Pipetting range is limited to 0.5–1 µL

The air-cushion system is specified for 0.5–1 µL only. Protocols requiring larger volumes or single-tube transfers fall outside this modification and belong to the 5070/5073/5075 platforms. Verify accuracy and precision in-house because systematic and random error are not published for this model.

Factory Specification Sheet
Maintenance
Do not autoclave; use the non-autoclave decontamination procedure

The unit is marked non-autoclavable. Sterilizing it in an autoclave risks damage to the electronics and air-cushion system. Follow the manufacturer’s wipe-down/decontamination instructions instead, and use the epT.I.P.S. Motion Reload System tips as the consumable path.

Factory Specification Sheet

Plans dimensionnels et documentation technique

Documentation technique

PDF
Automated Liquid Handling Brochure epMotion 5070 5073 5075 96 flex Unl
Brochure ·5,0 МБ ·19 pp.
Voir le PDF