Agilent 1100 Series Valves User's Guide
Technical Reference Summary
Covered Models & Document Scope
Covers Agilent 1100 Series valves: G1157A (2 position/10 port), G1158A (2 position/6 port), G1159A (6 position selection), G1160A (12 position/13 port selection), G1162A (2 position/6 port micro), G1163A (2 position/10 port micro). Includes installation, operation, maintenance, applications, parts and specifications.
Section-by-Section Technical Breakdown
Defines physical and electrical site requirements for the valves, including weight, dimensions, 24 V DC power, ambient temperature and humidity. Also lists Agilent 1100 modules that can supply 24 V DC power via DC-CAN adapter and the external power supply option.
- Weight 1.9 kg (4.2 lbs)
- Dimensions 92 x 84 x 200 mm
- Power supply 24 V DC (1.7 A)
- Ambient operating temperature 4 to 55 °C
- Humidity < 95%, non-condensing
- External power supply P/N 0950-4422 or listed 1100 modules with DC-CAN adapter
Describes how to check packaging for damage and lists the delivered parts for standard valves (G1157A, G1158A, G1159A, G1160A) and micro valves (G1162A, G1163A). Includes part numbers for rail assembly, CAN cables, fitting screws, hex keys and wrenches.
- Do not install valve if damage is visible
- Standard valves include rail assembly 5065-4450, CAN cable 5181-1519, DC-CAN cable 5181-1533
- Micro valves include M4 micro valve fittings 5022-2186 and plug 5022-2187
- M4 micro valve fittings support PEEK coated fused silica capillaries O.D. 0.8 mm or 1/32"
Explains two installation options: free bench space or mounting on the G1383A Column Organizer using the Organizer Rail Assembly. Describes CAN and DC-CAN connectors on the back panel and how to set up CAN and power connections.
- Valve can be placed on bench or mounted on G1383A Column Organizer
- Organizer Rail Assembly must be installed on mounting poles using clips
- Two CAN connectors and one DC-CAN connector on back panel
- Connect CAN cable 5181-1519 to a free CAN port on another 1100 module
- Connect DC-CAN cable 5181-1533 to a compatible 1100 module or external power supply 0950-4422
Describes how to configure the valves in Agilent ChemStation Rev. A.09.03 or higher. Includes the maximum module count limit and the steps to add valves in the configuration dialog.
- ChemStation Rev. A.09.03 or higher required
- Maximum number of all modules including CAN-slave valves must not exceed 13
- Use Configure 1100 Access in Instrument menu to add valves
- After restart, Setup Valve menu item appears
Explains valve parameter setup in ChemStation, including position selection, valve name, position descriptions, and the Time Table for valve programming during sequences. Also covers instrument actuals display.
- Position dropdown selects valve position on method load; Use Current keeps current position
- Valve Name limited to 20 characters
- Position Description limited to 19 characters
- Time Table limited to 20 lines with columns Line, Time, Position
- Next position after run option switches valve after run
Describes valve functions available with the Agilent 1100 Series handheld controller G1323B with firmware B.03.11 or higher, including switching the valve, synchronization, and setting EMF limits.
- Handheld controller firmware B.03.11 or higher required
- Switch valve via System > Control > Valve
- Synchronize valve via System > Tests > Valve
- Set EMF limit via System > Records > Valve > EMF > Setup Limits
- Reset switch counter after maintenance
Covers maintenance procedures for standard valves (G1157A, G1158A, G1159A, G1160A) and micro valves (G1162A, G1163A). Includes replacing stator face and rotor seal, and lists rebuild kits and repair parts.
- Standard valves: exchange stator face and rotor seal; stator head may need replacement if threads worn
- Micro valves: only rotor seal needs exchange; no stator face
- Always rinse valve with water after using aqueous buffers or salt solutions
- Tighten stator ring screws and stator screws finger-tight then another half turn
- Rebuild kits and part numbers listed in Table 6
Describes the EMF feature in ChemStation that monitors valve switch count and provides feedback when a user-settable limit is exceeded. Includes setting and resetting the EMF counter.
- EMF monitors number of switches
- EMF limit set in Diagnostics view
- Reset counter via Maintenance menu > Valve > Reset Counter
- EMF limit should be optimized over one or two maintenance cycles
Lists system error conditions with causes and actions, including no line power, CAN failure, switch failure, and unknown position. Also describes valve synchronization procedure.
- No line power: yellow Notready status, check power cable and module power
- CAN failure: red Pow. fail status, check CAN connections and module count
- Switch failure: yellow failure status, synchronize valve and restart ChemStation
- Unknown position: Act. Pos.: 0=-, synchronize valve
- Valve synchronization via Diagnostics view > Maintenance > Valve
Describes selected applications for the valves: alternating column regeneration, sample enrichment and stripping, column selection, and solvent selection. Includes flow diagrams and part numbers for capillary kits.
- Alternating column regeneration with G1157A using capillary kit G1156-68711 (analytical) or G1156-68713 (preparative)
- Sample enrichment/stripping with G1158A using capillary kit G1156-68714
- Column selection with G1159A using capillary kit G1156-68712 for up to 6 columns
- Solvent selection with G1160A using tubing kit G1160-68706 for 4 solvents
Lists general parts, exchange valve assemblies, rebuild kits, repair parts, capillaries and fittings for analytical, preparative and micro flow rates. Includes part numbers for CAN cables, power supply, rail assembly, valve assemblies, stator heads, rotor seals, and capillary kits.
- General parts: CAN cable 5181-1519, DC-CAN cable 5181-1533, external power supply 0950-4422, rail assembly 5065-4450
- Exchange valve assemblies listed for all valve models
- Rebuild kits and rotor seals for standard and micro valves
- Capillary and tubing kits for applications
- PEEK coated fused silica capillaries for micro valves with fitting types
Provides technical specifications for each valve model, including liquid contacts, port size, flow passage diameters, volume in flow passage, maximum pressure, and recommended flow range.
- G1157A: max pressure 41 MPa, flow range 0.2-100 ml/min
- G1158A: max pressure 41 MPa, flow range 0.2-100 ml/min
- G1159A: max pressure 35 MPa, flow range 0.3-40 ml/min (up to 100 ml/min without switching)
- G1160A: max pressure 21 MPa, flow range 0.2-100 ml/min (high pressure) or 0.2-10 ml/min (low pressure)
- G1162A: max pressure 41 MPa, flow range 0.1-100 µl
- G1163A: max pressure 41 MPa (specifications partially listed)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Maximum number of all modules (including CAN-slave valves) in an Agilent 1100 Series LC system must not exceed 13. p.8 ↗ | Plan system configuration carefully; exceeding 13 modules is not supported and may require verification with Agilent. | whole series |
| Valve power supply is 24 V DC (1.7 A), supplied either by external power supply 0950-4422 or by specific 1100 Series modules with DC-CAN adapter. p.2 ↗ | Ensure compatible power source is available; use listed modules or external supply. | whole series |
| Ambient operating temperature range is 4 to 55 °C (39 to 131 °F); humidity < 95%, non-condensing. p.2 ↗ | Install in a controlled laboratory environment within these limits. | whole series |
| G1157A maximum pressure is 41 MPa (408 bar, 6000 psi); recommended flow range 0.2 - 100 ml/min. p.56 ↗ | Select valve for applications within these pressure and flow limits. | G1157A |
| G1158A maximum pressure is 41 MPa (408 bar, 6000 psi); recommended flow range 0.2 - 100 ml/min. p.56 ↗ | Suitable for similar pressure/flow as G1157A but with 6 ports. | G1158A |
| G1159A maximum pressure is 35 MPa (345 bar, 5000 psi); recommended flow range 0.3 - 40 ml/min. Can be used up to 100 ml/min without valve switching. p.56 ↗ | For column selection, switch valve during postrun or prerun when flow can be reduced. | G1159A |
| G1160A maximum pressure is 21 MPa (207 bar, 3000 psi); recommended flow range 0.2 - 100 ml/min at high pressures (after pump) and 0.2 - 10 ml/min at low pressures (in front of pump). p.57 ↗ | Observe different flow limits depending on installation position relative to pump. | G1160A |
| G1162A maximum pressure is 41 MPa (408 bar, 6000 psi); recommended flow range 0.1 - 100 µl; volume in flow passages 70 nl port to port. p.57 ↗ | Suitable for micro-flow applications with very low volumes. | G1162A |
| Always rinse the valve with water after using aqueous buffers or salt solutions to prevent crystal formation which may damage the valve. p.21 ↗ | Include a water rinse step in maintenance procedures when using buffers/salts. | whole series |
| When reassembling valve head, tighten stator ring screws and stator screws finger-tight, then tighten another half turn. p.22 ↗ | Ensures proper sealing without over-tightening; follow this torque procedure. | standard valves G1157A, G1158A, G1159A, G1160A |
| For micro valves (G1162A, G1163A), only the rotor seal needs to be exchanged during maintenance; no stator face. p.21 ↗ | Simplifies maintenance and reduces parts inventory for micro valves. | G1162A, G1163A |
| ChemStation software revision A.09.03 or higher is required for control; 1100 firmware A.05.04 and higher; handheld controller G1323B firmware B.03.11 and higher. p.2 ↗ | Verify software/firmware versions before installation. | whole series |
| Valve Name is limited to 20 characters; Position Description limited to 19 characters; Time Table limited to 20 lines. p.13 ↗ | Plan naming and time table entries within these limits. | whole series |
| Capillary kit G1156-68711 is for alternating column regeneration with G1157A (analytical scale, flow rates < 10 ml/min); G1156-68713 for preparative scale (flow rates < 100 ml/min). p.35 ↗ | Select appropriate kit based on flow rate scale. | G1157A |
| Capillary kit G1156-68712 allows column selection with G1159A for up to 6 columns. p.42 ↗ | Enables multi-column method development or validation. | G1159A |
| Tubing kit G1160-68706 is for solvent selection (4 solvents) with G1160A, degasser and isocratic pump (flow rate < 10 ml/min). p.45 ↗ | Use for solvent selection applications within flow limit. | G1160A |
| M4 micro valve fittings support Agilent PEEK coated fused silica capillaries with O.D. 0.8 mm or 1/32". p.4 ↗ | Ensure capillary outer diameter matches fitting type for micro valves. | G1162A, G1163A |
| The G1159A 6 position selection valve can be used at flow rates up to 100 ml/min, but without valve switching. p.57 ↗ | For high flow rates, avoid switching during run; switch during postrun/prerun. | G1159A |
| EMF monitors number of valve switches and provides feedback when user-settable limit is exceeded; limit should be optimized over one or two maintenance cycles. p.27 ↗ | Use EMF to schedule preventive maintenance based on actual usage. | whole series |
| Valve synchronization is required if valve failed to switch or current position is unknown; use Diagnostics view > Maintenance > Valve > Synchronize to position. p.32 ↗ | Perform synchronization to recover from position errors. | whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| No line power p.35 ↗ | 24 volt DC line power is not available; yellow Notready status displayed; if trying to switch valve, status changes to Power. | 1. Check power cable to external power supply or to another 1100 Series module. 2. If valve power is supplied by another 1100 module, check if that module is switched on. 3. Restart the ChemStation. | |
| CAN failure p.36 ↗ | Red Pow. fail status visible; CAN communication is interrupted. | 1. Check the CAN connection at the valve and at all other modules. 2. Check that the maximum number of 13 Agilent 1100 Series modules (including CAN slaves) is not exceeded. 3. Restart the ChemStation. | |
| Switch failure p.36 ↗ | Yellow failure status indicates that the valve didn't switch correctly. | 1. Synchronize the valve as described on page 32. 2. Restart the ChemStation. 3. Disconnect and replug the CAN cable. | |
| Unknown Position p.37 ↗ | An unknown position is indicated Act. Pos.: 0=- | 1. Synchronize the valve as described on page 32. 2. Restart the ChemStation. 3. Disconnect and replug the CAN cable. |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Weight | 1.9 Kg p.8 ↗ | |
| Dimensions (height x width x depth) | 92 x 84 x 200 mm p.8 ↗ | |
| Power supply | 24 Volts DC p.8 ↗ | 1.7 amps |
| Ambient operating temperature | 4 to 55 °C p.8 ↗ | |
| Humidity | < 95 % p.8 ↗ | non-condensing |
| Agilent 1100 Firmware | A.05.04 and higher p.8 ↗ | |
| Control Module Firmware G1323B | B.03.11 and higher p.8 ↗ | |
| Agilent ChemStation Software | A.09.03 and higher p.8 ↗ | |
| Liquid contacts | Stainless Steel and PEEK p.62 ↗ | G1157A |
| Port size | Accepts 10-32 male threaded fittings p.62 ↗ | G1157A |
| Flow passage diameters | Stator and stator face assembly 0.6-mm (0.024”), rotor seal 0.6-mm (0.024”) mm p.62 ↗ | G1157A |
| Volume in flow passage | Stator (includes stator face seal) 2.1 µl/hole, rotor seal 0.7 µl/groove µl p.62 ↗ | G1157A |
| Maximum pressure | 41 MPa (408 bar, 6000 psi) MPa p.62 ↗ | G1157A |
| Recommended flow range | 0.2 - 100 ml/min p.62 ↗ | G1157A |
| Liquid contacts | Stainless steel, PEEK, and alumina ceramic p.62 ↗ | G1158A |
| Port size | Accepts 10-32 male threaded fittings p.62 ↗ | G1158A |
| Flow passage diameters | Stator and stator face assembly 0.4-mm (0.015”), rotor seal 0.5-mm (0.018”) mm p.62 ↗ | G1158A |
| Volume in flow passage | Stator (includes stator face seal) 0.7 µl/hole, rotor seal 0.3 µl/groove µl p.62 ↗ | G1158A |
| Maximum pressure | 41 MPa (408 bar, 6000 psi) MPa p.62 ↗ | G1158A |
| Recommended flow range | 0.2 - 100 ml/min p.62 ↗ | G1158A |
| Liquid contacts | Stainless steel and PEEK p.62 ↗ | G1159A |
| Port size | Accepts 10-32 male threaded fittings p.62 ↗ | G1159A |
| Flow passage diameters | Stator 0.6-mm (0.024”), stator face assembly and rotor seal 0.4-mm (0.015”) mm p.62 ↗ | G1159A |
| Volume in flow passage | Angled ports 1, 2, 5 (15.6 µl) Radial ports 2, 4, 6 (18,8 µl) µl p.62 ↗ | G1159A |
| Maximum pressure | 35 MPa (345 bar, 5000 psi) MPa p.62 ↗ | G1159A |
| Recommended flow range | 0.3 - 40 ml/min p.62 ↗ | G1159A |
| Liquid contacts | Stainless steel and PEEK p.63 ↗ | G1160A |
| Port size | Accepts 10-32 male threaded fittings p.63 ↗ | G1160A |
| Flow passage diameters | 1.0-mm (0.040”) mm p.63 ↗ | G1160A |
| Volume in flow passage | Stator (includes stator face seal) 6.4 µl/hole, rotor seal 4.0 µl/groove µl p.63 ↗ | G1160A |
| Maximum pressure | 21 MPa (207 bar, 3000 psi) MPa p.63 ↗ | G1160A |
| Recommended flow range | 0.2 - 100 (at high pressures, after the pump); 0.2 - 10 (at low pressures, in front of the pump) ml/min p.63 ↗ | G1160A |
| Liquid contacts | DuraLife processed stainless steel (stator) and vespel (rotor seal) p.63 ↗ | G1162A |
| Port size | Accepts M4 male threaded fittings p.63 ↗ | G1162A |
| Flow passage diameters | 0.20 mm (0.008") mm p.63 ↗ | G1162A |
| Volume in flow passages | 70 nl p.63 ↗ | port to port, G1162A |
| Maximum pressure | 41 MPa (408 bar, 6000 psi) MPa p.63 ↗ | G1162A |
| Recommended flow range | 0.1 - 100 µl p.63 ↗ | G1162A |
| Liquid contacts | DuraLife processed stainless steel (stator) and vespel (rotor seal) p.63 ↗ | G1163A |