Agilent 1200 Serie Automatische Probengeber Referenzhandbuch
Technical Reference Summary
Covered Models & Document Scope
Covers the Agilent 1200 Series automatic samplers: G1367B, G1367C (high-performance autosampler and SL version) and G1377A (micro well-plate autosampler), plus the G1330B ALS thermostat for the thermostatted configuration. Includes accessory kits G1367-68705, G1377-68705, Multi-Draw Kit G1313-68711 (G1367B/C only), and injector rinse kit G1373A.
Section-by-Section Technical Breakdown
Describes the universal power supply (100–240 VAC ±10 %) with no external voltage selector or fuses, and the separate power supplies for the autosampler and the G1330B thermostat. Specifies required free space around the instrument for air circulation and cable connections.
- Universal power supply accepts any voltage in the 100–240 VAC ±10 % range; no voltage selector switch on the rear.
- No externally accessible fuses; electronic fuses are inside the power supply.
- Thermostatted autosampler consists of autosampler (G1367B/C or G1377A) and thermostat (G1330B) with separate power cords; the thermostat has two externally accessible fuses.
- Free space around the autosampler: 2.5 cm (1.0 inch) on all sides and about 8 cm (3.1 inches) at the rear.
- Free space around the thermostatted autosampler: 25 cm (10 inches) on all sides plus 8 cm (3.1 inches) at the rear.
- Instrument must be set up in a horizontal position.
Table 1 lists weight, dimensions, mains input, frequency, power consumption, ambient temperature limits, humidity, altitude limits and safety standards for the G1367B/C and G1377A autosamplers.
- Weight 15.5 kg; dimensions 200 × 345 × 440 mm (8 × 13.5 × 17 inches).
- Mains input 100–240 VAC ±10 %, 50/60 Hz ±5 %.
- Apparent power max. 300 VA; active power max. 200 W.
- Permitted ambient temperature 4 to 55 °C; non-permitted range -40 to 70 °C.
- Humidity < 95 % at 25 to 40 °C, non-condensing.
- Max. operating altitude 2000 m; max. storage altitude 4600 m.
- Safety standards IEC, CSA, UL; installation category II, environmental protection category 2.
Covers inspection of shipping packaging, the delivery checklist, and the contents of the two accessory kits supplied with the high-performance/SL autosampler (G1367-68705) and the micro well-plate autosampler (G1377-68705).
- Check shipping packaging for visible damage and keep damaged packaging until completeness and function are verified.
- Accessory kit G1367-68705 is supplied with the G1367B/C high-performance and SL autosampler.
- Accessory kit G1377-68705 is supplied with the G1377A micro well-plate autosampler.
- Kits include capillaries, well plates, CAN cable, wrenches, ESD wrist strap, air duct adapter, leak kit and micro-seat capillary tool.
Recommends front and rear module arrangements for a complete Agilent 1200 Series system with the standard or thermostatted autosampler to achieve optimum performance and minimum delay volumes.
- Recommended front arrangement: solvent cabinet, vacuum degasser, pump, column compartment, autosampler, detector, Instant Pilot.
- Rear connections include GPIB or LAN to the LC ChemStation, analog signal to recorder, remote cable, CAN bus cable and power cables.
- For a complete system with thermostatted autosampler, Agilent recommends arranging modules in two stacks.
Step-by-step installation of the standard autosampler: removing transport foam, installing the LAN interface card if required, connecting power and CAN cables, and verifying initialization via the status LED.
- Remove tape from side and front doors, open front door, remove left side door and transport foam.
- Reinstall the waste corrugated tube into the plastic connector and reinstall the left side door (note the magnet on the rear).
- Place the autosampler horizontally on the bench or in the instrument stack.
- Connect the power cable to the rear mains inlet and the CAN cable to other Agilent 1200 modules.
- For a ChemStation controller, connect the LAN port to the LAN interface.
- After switching on and closing the front door, the exhaust fan starts and hardware initialization begins after one to two minutes; the status LED should then be green.
Installation of the G1330B thermostat together with the autosampler, including condensate drain routing, air duct adapter fitting, and interconnecting cables between the two modules.
- Place the thermostat on the lab bench and route the waste hose into the waste container.
- Ensure the condensate drain line is always above the liquid level in the collection vessel to avoid blocking the outlet and damaging electronics.
- Place the autosampler on top of the thermostat and verify it engages correctly.
- Remove the sample tray and plastic cover from the tray base and fit the air duct adapter fully down to direct cold air from the thermostat to the well-plate area.
- Do not connect or disconnect the autosampler-to-thermostat cable while either module is connected to mains power.
- Connect power cables, CAN cable, optional APG remote cable, and LAN if using a ChemStation.
Hydraulic connections between pump, injection valve, column oven, needle seat, leak overflow and peristaltic rinse pump.
- Connect the pump outlet capillary to port 1 of the injection valve.
- Connect the column oven inlet capillary to port 6 of the injection valve.
- Connect the waste line to the needle seat adapter and to the solvent waste from the leak overflow.
- Verify the waste line is in the leak channel.
- Route the line from the peristaltic rinse pump to the solvent container in the solvent cabinet.
- Select the seat capillary according to the recommendations in 'Selecting the seat capillary' (page 56).
Describes installing the sample tray, supported tray part numbers, and vial/well-plate position numbering.
- Press the underside of the right side to unlock the front door.
- Load the required well plates or sample vials into the tray.
- Slide the tray in so its rear sits close to the rear of the tray compartment, then press the front down to secure the carrier.
- If the tray pops out of position, the air duct adapter is not installed correctly.
- Supported trays: G2258-60011 for 2 well plates or vial plates plus 10 × 2 ml vials; G1329-60001 for 100 × 2 ml vials, thermostatted.
Tables of recommended well plates and cover mats with row/column counts, plate height, volume and ordering numbers, plus vial plates.
- 384-well Agilent plate: 16 rows × 24 columns, height 14.4 mm, 80 µl, order no. 5042-1388 (pack of 30).
- 96-well Agilent plate: 8 rows × 12 columns, height 14.3 mm, 400 µl, order no. 5042-1386 / 5042-1385 (pack of 10/120).
- 96-well conical Agilent plate: height 17.3 mm, 150 µl, order no. 5042-8502 (pack of 25).
- 96-deep-well Agilent 31 mm plate: height 31.5 mm, 1000 µl, order no. 5042-6454 (pack of 50).
- Sealing mat for all 96 Agilent plates: order no. 5042-1389 (pack of 50).
- Vial plates: G2255-68700 for 54 × 2 ml vials; 5022-6539 for 15 × 6 ml vials; 5022-6538 for 27 Eppendorf tubes.
Ordering tables for crimp-top, snap-top and screw-top vials and matching caps/septa.
- Crimp-top clear glass 2 ml vials: 5181-3375 (100 pcs), 5183-4491 (1000 pcs).
- Snap-top clear glass 2 ml vials: 5182-0544 (100 pcs), 5183-4504 (1000 pcs).
- Screw-top clear glass 2 ml vials: 5182-0714 (100 pcs), 5183-2067 (1000 pcs).
- Screw caps blue polypropylene with clear PTFE/red rubber septum: 5182-0717 (100 pcs).
- Crimp caps silver aluminum with clear PTFE/red rubber septum: 5181-1210 (100 pcs).
Explains how to configure a custom well plate in the ChemStation plate configuration table by measuring plate dimensions, with limits for rows, columns, volume, pitch, plate length/width/height, offsets and well diameter/depth.
- Rows: up to 16; columns: up to 24.
- Plate length C: 127.75 ± 0.25 mm (SBS standard); plate width D: 85.50 ± 0.25 mm (SBS standard).
- Plate height E: up to 47 mm.
- Well diameter I: at least 4 mm; well depth J: up to 45 mm.
- Distances must be measured extremely accurately; use a caliper.
Guidance on minimizing carryover, shortening injection cycles, reducing dead volume and achieving precise injection volumes.
- Optimization goals: lowest possible carryover, short injection cycles, low dead volume, precise injection volume.
- For SL performance optimization, refer to the Agilent 1200 Series Rapid Resolution LC System User Manual.
Describes how injector programming combined with needle rinse in the rinse port can reduce carryover by a factor of about 10 compared with conventional rinse-port-only needle rinsing, and describes the G1373A injector rinse kit.
- Injector program plus rinse-port needle rinse reduces carryover by a factor of about 10 versus conventional rinse-port-only rinsing.
- Overlapping injection together with a further injection valve switchover is not possible.
- G1373A injector rinse kit provides minimum carryover for viscous samples, minimum injection volumes, maximum sensitivity and short rinse times.
- Software control requires Agilent ChemStation version B.01.03 or newer and firmware A.06.01 or newer on all Agilent 1200 modules, plus firmware B.04.01 for the control module.
Explains overlapping injections and bypass switching to reduce system dead volume by about 300 µl, and gives recommendations for short injection cycles.
- Switching the injection valve to bypass reduces system dead volume by about 300 µl.
- Bypass switching can increase carryover in the system.
- Injecting 100 µl instead of 1 µl increases injection duration by about 8 seconds under identical standard conditions.
- For the last injection in an overlapping-injection sequence, add an empty run because the valve is not switched and retention times become longer.
- Set detector equilibration to OFF to shorten injection duration.
- Needle outside rinse duration at least 10 seconds.
Compares Vespel and Tefzel rotor seals for the injection valve with respect to mobile-phase pH range and expected lifetime.
- Standard Vespel seal is suitable for mobile phases in the pH range 2.3 to 9.5.
- Below pH 2.3 or above pH 9.5, the Vespel seal wears faster and has a shorter lifetime.
- Tefzel seal is available for extreme pH conditions; it is more resistant to extreme pH but is a softer material.
- Under normal conditions the Tefzel seal has a shorter expected lifetime than Vespel, but longer under extreme mobile phases.
The autosampler has two status indicators showing operating state (before analysis, not ready, analysis, error). Electronic, mechanical or hydraulic faults generate an error message on the control computer with description, possible causes and remedies.
- Two status indicators provide a quick overview of the operating state.
- Each error message includes a short description, a list of possible causes and a list of corrective actions.
- Maintenance functions position the needle unit, needle carrier, sample transport unit and metering unit for easy access.
The transport unit cannot complete a movement in one axis direction within the processor-defined time; encoders in the stepper motors monitor position. Axis mapping: Arm Movement 0 = X axis, 1 = Z axis, 2 = Theta (needle carrier rotation).
- Possible causes: mechanical obstruction, high friction in the transport unit, defective motor unit, defective board in the sample transport unit, defective MTP board.
- Remedies: ensure unobstructed transport unit movement, replace the sample transport unit, replace the MTP board.
The autosampler did not complete initialization properly; the processor checks position sensors and motor actuators during the initialization pattern.
- Possible causes: side door not properly inserted, sample transport/sampling unit not properly positioned, mechanical obstruction, defective sampling unit board (SUD board), defective board in transport unit, defective sampling unit motor, defective MTP board.
- Remedies: verify side door is properly inserted and its magnet is in place, perform automatic adjustment, ensure unobstructed transport unit movement, replace defective sampling unit motor, replace sample transport unit, replace MTP board.
One of the transport unit motors drew too much current and overheated; the processor switched it off to protect it. Motor 0 = X-axis motor, Motor 1 = Z-axis motor, Motor 2 = Theta motor (gripper rotation).
- Possible causes: mechanical obstruction, high friction in the transport unit, drive belt tension too high, defective motor, defective board in the transport unit.
- Remedies: switch off the autosampler at the power switch and wait at least 10 minutes before switching on again, ensure unobstructed transport unit movement, replace the sample transport unit, replace the MTP main board.
Block diagram of the BCD card showing processor interface, board identification, BCD register, external contacts, line drivers (12 × 250 mA), BCD connector, external contact connector and RF interference filters.
- BCD card provides 12 line drivers rated 4 × 250 mA per group as shown in the block diagram.
- Universal cable connects the BCD output and external outputs to external devices.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Mains input is 100–240 VAC ±10 %, 50/60 Hz ±5 %, with no external voltage selector and no externally accessible fuses. p.14 ↗ | The autosampler can be connected to any mains voltage in the specified range without reconfiguration; branch protection must be provided externally. | G1367B/C, G1377A |
| Maximum apparent power is 300 VA and maximum active power is 200 W. p.16 ↗ | Size the laboratory power distribution and UPS for at least 300 VA per autosampler. | G1367B/C, G1377A |
| Permitted ambient temperature is 4 to 55 °C; humidity < 95 % at 25 to 40 °C, non-condensing. p.16 ↗ | Avoid locations where temperature fluctuations cause condensation inside the instrument; allow slow warm-up in the original packaging after cold transport. | G1367B/C, G1377A |
| Free space around the autosampler must be 2.5 cm (1.0 inch) on all sides and about 8 cm (3.1 inches) at the rear. p.15 ↗ | Ensure adequate air circulation and room for cable connections when placing the instrument on the bench or in a stack. | G1367B/C, G1377A |
| The thermostatted autosampler requires 25 cm (10 inches) free space on all sides plus 8 cm (3.1 inches) at the rear. p.15 ↗ | Plan a larger bench footprint or a two-stack arrangement when using the G1330B thermostat. | Thermostatted configuration (G1367B/C or G1377A + G1330B) |
| The condensate drain line must always remain above the liquid level in the collection vessel. p.26 ↗ | If the line is submerged, condensate cannot drain, the outlet blocks and condensate remains in the instrument, potentially damaging the electronics. | Thermostatted configuration (G1330B) |
| Do not connect or disconnect the autosampler-to-thermostat cable while either module is connected to mains power. p.28 ↗ | Power down and unplug both modules before making or breaking the interconnecting cable to avoid damaging module electronics. | Thermostatted configuration (G1367B/C or G1377A + G1330B) |
| After switching on and closing the front door, hardware initialization starts after one to two minutes and the status LED should then be green. p.25 ↗ | A status LED that does not turn green after initialization indicates a fault requiring troubleshooting. | G1367B/C, G1377A |
| The standard well-plate sample tray G2258-60001 must be installed and empty for self-adjustment of the sample transport unit. p.61 ↗ | Self-adjustment cannot be performed with a loaded tray or a different tray; plan adjustment before loading samples. | G1367B/C, G1377A |
| Self-adjustment is required after disassembly or replacement of the sample transport unit, sampling unit, sample tray or MTP board. p.61 ↗ | Include a self-adjustment step in any maintenance procedure involving these components. | G1367B/C, G1377A |
| Switching the injection valve to bypass reduces system dead volume by about 300 µl. p.51 ↗ | Use bypass switching to shorten injection cycles, especially at low flow rates with narrow-bore columns, but expect increased carryover. | G1367B/C, G1377A |
| Injecting 100 µl instead of 1 µl increases injection duration by about 8 seconds under identical standard conditions. p.51 ↗ | Increase draw and dispense speeds for large injection volumes where sample viscosity permits, to keep cycle times short. | G1367B/C, G1377A |
| For overlapping-injection sequences, the last injection does not switch the valve and therefore has longer retention times. p.52 ↗ | Add an empty run as the last injection in the sequence to avoid retention-time shifts outside the calibration table. | G1367B/C, G1377A |
| The standard Vespel rotor seal is suitable for mobile phases from pH 2.3 to 9.5; Tefzel is available for pH outside this range. p.55 ↗ | Select the Tefzel seal for mobile phases below pH 2.3 or above pH 9.5 to avoid accelerated Vespel wear. | Injection valve |
| Custom well plates must have length 127.75 ± 0.25 mm and width 85.50 ± 0.25 mm (SBS standard), height up to 47 mm, well diameter at least 4 mm and well depth up to 45 mm. p.40 ↗ | Measure custom plates with a caliper and enter the dimensions in the ChemStation plate configuration table; out-of-tolerance plates may cause positioning errors. | G1367B/C, G1377A |
| The G1373A injector rinse kit requires Agilent ChemStation version B.01.03 or newer and firmware A.06.01 or newer on all Agilent 1200 modules, plus firmware B.04.01 for the control module. p.48 ↗ | Verify software and firmware versions before installing the injector rinse kit. | G1373A injector rinse kit |
| The BCD interface card provides line drivers rated 4 × 250 mA as shown in the block diagram. p.254 ↗ | External devices connected to the BCD outputs must not exceed the driver current rating. | Optional BCD interface card |
| The autosampler has no pre-piercing needle, so adhesive cover mats can clog the needle after several injections. p.35 ↗ | Do not use adhesive cover mats as a rule; use non-adhesive sealing mats such as 5042-1389. | G1367B/C, G1377A |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Arm Movement Failed / Arm Movement Timeout p.78 ↗ | Mechanical obstruction; high friction in the transport unit; defective motor unit; defective board in the sample transport unit; defective MTP board. | Ensure unobstructed movement of the transport unit; replace the sample transport unit; replace the MTP board. | Arm Movement 0 = X axis, 1 = Z axis, 2 = Theta (needle carrier rotation) |
| Initialization failed p.85 ↗ | Side door not properly inserted; sample transport/sampling unit not properly positioned; mechanical obstruction; defective sampling unit board (SUD board); defective board in the transport unit; defective sampling unit motor; defective MTP board. | Verify side door is properly inserted and its magnet is in place; perform automatic adjustment; ensure unobstructed transport unit movement; replace defective sampling unit motor; replace the sample transport unit; replace the MTP board. | |
| Motor Temperature p.87 ↗ | Mechanical obstruction; high friction in the transport unit; drive belt tension too high; defective motor; defective board in the transport unit. | Switch off the autosampler at the power switch and wait at least 10 minutes before switching on again; ensure unobstructed transport unit movement; replace the sample transport unit; replace the MTP main board. | Motor 0 = X-axis motor, Motor 1 = Z-axis motor, Motor 2 = Theta motor (gripper rotation) |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Weight | 15,5 kg p.16 ↗ | G1367B/C / G1377A |
| Dimensions (height × width × depth) | 200 × 345 × 440 mm p.16 ↗ | G1367B/C / G1377A |
| Dimensions (height × width × depth) | 8 × 13.5 × 17 inches p.16 ↗ | G1367B/C / G1377A |
| Mains input | 100 – 240 VAC, ±10 % VAC p.16 ↗ | multi-range |
| Frequency | 50 or 60 Hz, ±5 % Hz p.16 ↗ | |
| Power consumption (apparent power) | 300 VA p.16 ↗ | maximum |
| Power consumption (active power) | 200 W p.16 ↗ | maximum |
| Permitted ambient temperature | 4 bis 55 °C p.16 ↗ | |
| Non-permitted temperature range | -40 bis 70 °C p.16 ↗ | |
| Humidity | < 95 %, bei 25 bis 40 °C % p.16 ↗ | non-condensing |
| Max. altitude during operation | bis 2000 m p.16 ↗ | |
| Max. altitude during storage | bis 4600 m p.16 ↗ | for storage of the autosampler |
| Safety standards | IEC, CSA, UL p.16 ↗ | Installation category II, environmental protection category 2 |