Agilent 1260 Infinity Fraction Collectors User Manual
Technical Reference Summary
Covered Models & Document Scope
Covers the Agilent 1260 Infinity Fraction Collectors G1364B (preparative-scale, flow rates up to 100 ml/min, vials, deep well plates and test tubes up to 100 mm height) and G1364C (analytical-scale, flow rates below 10 ml/min, vials, test tubes up to 75 mm height, well-plates and a 40-funnel tray). A thermostatted analytical-scale variant is created by adding the G1330B Fraction Collector Thermostat. Accessory kits G1364-68725 and G1330-68705 are listed.
Section-by-Section Technical Breakdown
Defines power, bench space and environmental requirements for the fraction collector and the optional thermostat. The power supply is wide-ranging with no external voltage selector or accessible fuses; the thermostatted configuration uses two separate power supplies and a control cable. Bench space must allow air circulation and the instrument must be installed horizontally.
- Line voltage 100 – 240 VAC ±10 %, 50/60 Hz ±5 %, no voltage selector, automatic electronic fuses.
- Thermostatted fraction collector comprises G1364B/G1364C plus G1330B thermostat, each with its own power supply and power plug.
- Additional 2.5 cm (1.0 inch) space on either side and approximately 8 cm (3.1 inches) at the rear for the fraction collector.
- Thermostatted fraction collector requires 25 cm (10 inches) on either side for air circulation.
- Do not store, ship or use under conditions causing condensation; allow cold-shipped units to warm slowly in the box.
Describes inspection of shipping containers, the delivery checklist for the G1364B/G1364C and the optional G1330B thermostat, and the contents of the accessory kits with part numbers. Three model configurations are defined by flow range and vessel compatibility.
- G1364B preparative-scale: flow rates up to 100 ml/min, vials, deep well plates and test tubes up to 100 mm height.
- G1364C analytical-scale: flow rates below 10 ml/min (modifiable for higher flow rates), vials, test tubes up to 75 mm height, well-plates and a 40-funnel tray.
- G1364C thermostatted analytical scale created by adding G1330B Fraction Collector Thermostat.
- Fraction Collector Accessory Kit G1364-68725 includes wrenches, hex key, fittings, waste tubing, Y-fitting, CAN cable, air channel adapter, clamps, ESD wrist strap and trays G1313-44513 and G1313-44512.
- Thermostat Module Accessory Kit G1330-68705 includes waste tubing 5062-2463 and cable G1330-81600.
Provides recommended front and rear stack arrangements for analytical and preparative systems to achieve optimum performance and minimum delay volume. Notes that a G1330B thermostat must always be installed underneath the thermostatted module and that the fraction collector must not sit on top of heat-generating modules.
- Recommended analytical stack: solvent compartment, degasser, pump, autosampler, column compartment, UV-Vis detector, fraction collector.
- Recommended preparative stack: solvent compartment, preparative pumps A/B, autosampler, column organizer, UV-Vis detector, fraction collector.
- G1330B thermostat must always be installed underneath the thermostatted injector or fraction collector.
- Never position the fraction collector on top of a heat-generating module (e.g. G1316A TCC or diode array detector) to avoid evaporation of fractions.
- CAN bus and GPIB or LAN connections to the ChemStation are shown in the rear-view figures.
Step-by-step installation of the standalone fraction collector, including LAN board installation, removal of transport protection, waste tubing routing, side door installation, power and CAN cabling, and tray installation. The side door presence is sensed by a hall sensor and a missing door causes a NOT-READY state.
- Install corrugated waste tube in the plastic port at the front bottom center and route to a waste container.
- Re-install the left side door; a missing side door results in a NOT-READY state of the instrument.
- Place the instrument horizontally, ensure power switch is OFF before connecting the power cable.
- Connect CAN cable to other Agilent 1260 Infinity LC modules; ChemStation connects to the LAN interface (normally on the detector).
- Optional APG remote cable for non-Agilent 1260 Infinity instruments.
- Test tube trays can be adjusted in height depending on test tube height.
Installation sequence for combining the fraction collector with the G1330B thermostat, including condensation drain routing, air channel adapter and plug channel placement, tray compartment divider use, and the thermostat control cable. The exhaust fan runs for 2 minutes before the front door is closed and hardware initialization starts.
- Route the condensation drain tube downward into a waste container without upward bends; keep the tube outlet above the liquid level.
- Place the fraction collector on top of the thermostat and ensure it engages in the thermostat locks.
- Install the air channel adapter and plug channel fully pressed down to guide the cold airstream to the tray area.
- Tray compartment divider should only be used with the thermostatted fraction collector when a half or standard tray is installed in the left and center positions; remove it if a tray is installed on the right side.
- Do not disconnect or reconnect the fraction collector to thermostat cable while power cords are connected.
- After power-on the exhaust fan runs; close the front door after 2 minutes, then hardware initialization starts and the status LED should be green.
Describes routing of the waste outlet around the instrument into corrugated waste tubing and fixing with sticking holders from the accessory kit. Warns about solvent leakage, regular inspection of inlet/waste and valve-to-needle tubing, and the solvent ignition temperature limit.
- Route waste outlet around the instrument into corrugated waste tubing and fix with sticking holders from G1364-68725.
- Regularly inspect inlet/waste tubing assembly and valve-to-needle tubing and exchange if worn or damaged.
- Use only solvents with an ignition temperature higher than 200 °C.
- Observe safety procedures for toxic or hazardous solvents when opening capillary or tube fittings.
Explains how to configure custom well plates when the plate is not in the recommended list, by measuring plate dimensions and entering them in the ChemStation or Agilent Instant Pilot G4208A plate configuration table. Defines rows, columns, volume, row/column distance, plate length/width/height, offsets, column shift and well depth with limits.
- Rows up to 16; columns up to 24.
- Plate length X size 127.75 ± 0.25 mm (SBS standard); plate width Y size 85.50 ± 0.25 mm (SBS standard).
- Plate height up to 47 mm, measured from bottom to top of caps if caps/septa are used.
- Column offset measured from left edge (bottom) to center of first hole A1; row offset from back edge (bottom) to center of first hole A1.
- Column shift is the offset to Y for staggered rows.
- Well depth measured from top of plate (or top of septa) to bottom of well (or bottom of inlets).
Check-out injection verifies correct installation and configuration of the Agilent 1260 Infinity Purification System and can be used for troubleshooting. Procedures depend on flow range (analytical AS or preparative PS) and peak trigger (UV-based or mass-based); UV-triggered procedures are described here.
- Prerequisites include thorough flushing with the liquid phase, UV lamp on for at least 15 minutes for stable baseline.
- Fraction collector must be installed and configured and delay volume calibrated with a UV detector.
- 40-vial tray G1313-44512 with at least four 2 ml vials must be installed.
- All wells must be empty and stored fill levels reset by removing and reinserting the tray and confirming the dialog.
- For mass-based systems refer to Agilent 1260 Infinity Purification System User Manual G2262-90011.
Two status indicators on the front of the fraction collector show power supply status and instrument status. The power supply indicator is integrated into the main power switch and is green when power is ON. The instrument status indicator shows prerun, run, not-ready and error conditions by OFF, green, yellow and red states.
- Power supply indicator green when power is ON.
- Status indicator OFF (with power switch light on): prerun condition, ready to begin analysis.
- Green status indicator: instrument performing an analysis (run mode).
- Yellow status indicator: not-ready condition, e.g. front door not closed or self-test running.
- Red status indicator: error condition affecting correct operation, always interrupts the analysis.
Maintenance positions move the needle arm and needle carrier to specific positions for access to components. Change Parts moves the transport unit upward and the needle carrier to front center and turns off the theta motor for free rotation; Home Position moves the arm up and to the right rear; Park Arm moves the arm to the upper rear left for transport or shipping.
- Change Parts enables access to needle assembly, needle carrier assembly, valve-to-needle tubing and diverter valve.
- After maintenance click End to move the transport assembly to the Home position.
- Home Position moves the arm up and to the right rear for tray access and exchange.
- Park Arm moves the arm to the park position at the upper rear left side of the tray for transporting or shipping.
- Functions are selected from the Maintenance menu in the ChemStation Diagnosis view or the Test screens of the Agilent Instant Pilot G4208A.
Step commands allow manual execution of individual movements of the fraction collection sequence for troubleshooting and repair verification. Each step command consists of a series of individual commands moving components to predefined positions.
- Needle Up lifts the needle arm to the upper position and switches the diverter valve to waste if not already there.
- Needle into vessel lowers the needle into the specified vessel (analytical-scale G1364C only).
- Needle to rinse/flush port moves the needle to the rinse/flush port (analytical-scale G1364C only).
- Switch valve to needle switches the diverter valve to the needle.
- Switch valve to waste switches the diverter valve from needle to waste.
- Step commands are selected from the Test Selection Box in the ChemStation Diagnosis display or the Test pull-down menu of the Agilent Instant Pilot G4208A.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| G1364B preparative-scale fraction collector is designed for flow rates up to 100 ml/min and test tubes up to 100 mm height. p.14 ↗ | Select G1364B for preparative purification with high flow rates and tall test tubes. | G1364B |
| G1364C analytical-scale fraction collector is designed for flow rates below 10 ml/min (can be modified for higher flow rates) and test tubes up to 75 mm height. p.14 ↗ | Select G1364C for analytical-scale collection; higher flow rates require modification. | G1364C |
| Fraction collector line voltage is 100 – 240 VAC ±10 %, 50/60 Hz ±5 %, with no voltage selector and automatic electronic fuses. p.12 ↗ | No voltage selection or fuse replacement is required; connect to any supply in the specified range. | G1364B, G1364C |
| Thermostat G1330B line voltage is 100 – 120 or 220 – 240 VAC ±10 %, 50/60 Hz ±5 %, with two externally accessible fuses. p.13 ↗ | Thermostat supply range differs from the fraction collector and its fuses are accessible for replacement. | G1330B |
| Fraction collector requires an additional 2.5 cm (1.0 inch) on either side and approximately 8 cm (3.1 inches) at the rear for air circulation and electrical connections. p.11 ↗ | Reserve bench space accordingly to avoid overheating and allow cable access. | G1364B, G1364C |
| Thermostatted fraction collector requires 25 cm (10 inches) of space on either side for air circulation. p.11 ↗ | Plan a wider bench area when using the G1330B thermostat. | G1364C with G1330B |
| A missing left side door results in a NOT-READY state of the instrument because its presence is sensed by a hall sensor. p.23 ↗ | Ensure the side door is correctly installed after maintenance to avoid a not-ready condition. | G1364B, G1364C |
| The G1330B thermostat must always be installed underneath the thermostatted injector or fraction collector. p.19 ↗ | Stack the thermostat at the bottom of the thermostatted module to ensure correct thermal operation. | G1330B |
| The fraction collector should never be positioned on top of a heat-generating module such as the G1316A thermostatted column compartment or a diode array detector. p.19 ↗ | Avoid unwanted evaporation of fractions by keeping the fraction collector off heat sources. | G1364B, G1364C |
| Do not disconnect or reconnect the fraction collector to thermostat cable while power cords are connected to either module. p.29 ↗ | Power down both modules before connecting or disconnecting the thermostat cable to prevent electronics damage. | G1364C with G1330B |
| The condensation drain tube must run down into a waste container without upward bends and must always be above the liquid level in the container. p.26 ↗ | Blocked condensation drainage can damage the instrument electronics. | G1330B |
| After power-on the exhaust fan runs for 2 minutes before the front door should be closed; then hardware initialization starts and the status LED should be green. p.29 ↗ | Follow the 2-minute fan run before closing the door to allow solvent vapor removal and correct initialization. | G1364C with G1330B |
| Only solvents with an ignition temperature higher than 200 °C should be used with the instrument. p.31 ↗ | Verify solvent ignition temperature before use to avoid fire hazard. | G1364B, G1364C |
| Transport unit self alignment requires the 4-well-plate tray G1364-84501 with no well plates installed and takes approximately 10-15 minutes. p.97 ↗ | Schedule alignment time and remove all plates before starting; otherwise the procedure aborts without error message. | G1364B, G1364C |
| On a G1364B preparative-scale fraction collector, install the delay calibration adapter G1364-87301 for transport unit self alignment. p.98 ↗ | Use the correct adapter for preparative-scale alignment to ensure correct positioning. | G1364B |
| If the transport unit has been exchanged or is very badly misaligned, set the 8-bit configuration switch to the Forced Cold Start Configuration before alignment, then return it to the default setting afterwards. p.97 ↗ | Forced cold start is required for correct alignment after transport unit exchange. | G1364B, G1364C |
| Well-plate configuration limits: rows up to 16, columns up to 24, plate length 127.75 ± 0.25 mm, plate width 85.50 ± 0.25 mm, plate height up to 47 mm. p.39 ↗ | Custom plates must comply with SBS dimensions and height limit for reliable needle positioning. | G1364B, G1364C |
| The Needle Up step command also switches the diverter valve to waste if it is not already in that position. p.100 ↗ | Use Needle Up to safely park the needle and route flow to waste during troubleshooting. | G1364B, G1364C |
| Needle into vessel and Needle to rinse/flush port step commands are available only on the analytical-scale G1364C. p.100 ↗ | Preparative-scale G1364B does not support these manual step commands. | G1364C |
| Check-out procedure requires the 40-vial tray G1313-44512 with at least four 2 ml vials, all wells empty, and stored fill levels reset by removing and reinserting the tray. p.84 ↗ | Reset fill levels before check-out to avoid false volume errors. | G1364C |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Status indicator OFF (and power switch light is on) p.94 ↗ | Instrument is in a prerun condition and is ready to begin an analysis. | No action required; instrument is ready. | Instrument Status Indicator |
| Green status indicator p.94 ↗ | Instrument is performing an analysis (run mode). | No action required; analysis in progress. | Instrument Status Indicator |
| Yellow status indicator p.94 ↗ | Not-ready condition: instrument is waiting for a specific condition to be reached or completed (for example, front door not closed), or a self-test procedure is running. | Check and resolve the pending condition (e.g. close front door) or wait for self-test to complete. | Instrument Status Indicator |
| Red status indicator p.94 ↗ | Error condition: instrument has detected an internal problem which affects correct operation (for example, leak, defective internal components). | Attention required; error condition always interrupts the analysis. Refer to the Agilent 1260 Infinity Fraction Collectors Service Manual G1364-90111 for detailed causes and actions. | Instrument Status Indicator |
| Transport Unit Self Alignment aborted without error message p.97 ↗ | Alignment procedure was started with well plates installed on the 4-well-plate tray. | Remove all well plates from the tray and restart the alignment procedure. | Transport Unit Self Alignment |
| NOT-READY state of the instrument p.23 ↗ | Missing left side door; its presence is sensed by a hall sensor. | Re-install the left side door correctly (take care of the magnet at the back). | Installing the Fraction Collector |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Weight | 13.5 kg (29.8 lbs) p.12 ↗ | G1364B, G1364C |
| Dimensions (height × width × depth) | 200 × 345 × 440 mm (8 × 13.5 × 17 inches) p.12 ↗ | G1364B, G1364C |
| Line voltage | 100 – 240 VAC, ±10 % VAC p.12 ↗ | G1364B, G1364C |
| Line frequency | 50 or 60 Hz, ±5 % Hz p.12 ↗ | G1364B, G1364C |
| Power consumption (apparent power) | 200 VA p.12 ↗ | G1364B, G1364C, maximum |
| Power consumption (active power) | 180 W p.12 ↗ | G1364B, G1364C, maximum |
| Ambient operating temperature | 4 – 55 °C (41 – 131 °F) p.12 ↗ | G1364B, G1364C |
| Ambient non-operating temperature | -40 – 70 °C (-4 – 158 °F) p.12 ↗ | G1364B, G1364C |
| Humidity | < 95 %, at 25 – 40 °C (77 – 104 °F) % p.12 ↗ | G1364B, G1364C, non-condensing |
| Operating altitude | Up to 2000 m (6500 ft) p.12 ↗ | G1364B, G1364C |
| Non-operating altitude | Up to 4600 m (14950 ft) p.12 ↗ | G1364B, G1364C, for storing the Fraction Collector |
| Safety standards | IEC, CSA, UL p.12 ↗ | G1364B, G1364C, Installation Category II, Pollution Degree 2, indoor use only |
| Weight | 18.5 kg (40.7 lbs) p.13 ↗ | G1330B thermostat |
| Dimensions (height × width × depth) | 140 × 345 × 435 mm (5.5 × 13.5 × 17 inches) p.13 ↗ | G1330B thermostat |
| Line voltage | 100 – 120 or 220 – 240 VAC, ± 10 % VAC p.13 ↗ | G1330B thermostat |
| Line frequency | 50 or 60 Hz, ± 5 % Hz p.13 ↗ | G1330B thermostat |
| Power consumption (active power) | 210 W p.13 ↗ | G1330B thermostat, maximum |
| Power consumption (apparent power) | 260 VA p.13 ↗ | G1330B thermostat, maximum |
| Ambient operating temperature | 4 – 40 °C (41 – 131 °F) p.13 ↗ | G1330B thermostat |
| Ambient non-operating temperature | -40 – 70 °C (-4 – 158 °F) p.13 ↗ | G1330B thermostat |
| Humidity | < 95 %, at 25 – 40 °C (77 – 104 °F) % p.13 ↗ | G1330B thermostat, non-condensing |
| Operating altitude | Up to 2000 m (6500 ft) p.13 ↗ | G1330B thermostat |
| Non-operating altitude | Up to 4600 m (14950 ft) p.13 ↗ | G1330B thermostat, for storing the Fraction Collector |
| Safety standards | IEC, CSA, UL, EN p.13 ↗ | G1330B thermostat, Installation Category II, Pollution Degree 2, indoor use only |