Agilent 1260 Infinity Binary LC System User Guide
Technical Reference Summary
Covered Models & Document Scope
Covers the Agilent 1260 Infinity Binary LC system and its typical modules: 1260 Infinity Binary Pump (G1312B), 1260 Infinity High Performance Degasser (G4225A), 1260 Infinity High Performance Autosampler (G1367E), 1290 Infinity Thermostatted Column Compartment (G1316C), 1260 Infinity Diode Array Detector (G4212B), and 1200 Infinity Series Quick Change Valves. Valid for Agilent OpenLAB CDS revision A.01.04. System configurations include low, medium and standard delay volume variants.
Section-by-Section Technical Breakdown
Describes the typical modules of the 1260 Infinity Binary LC: binary pump, degasser, autosampler, thermostatted column compartment, diode array detector and quick change valves. Each module's principle of operation and key design features are outlined.
- Binary pump uses two identical pumps with high-pressure mixing
- Degasser uses four vacuum chambers with semipermeable tubing
- Autosampler uses flow-through design for precision and low carryover
- TCC provides two independent temperature zones
- DAD uses Max-Light cartridge cell with optofluidic waveguide technology
Provides physical and performance specifications for the 1260 Infinity Binary Pump (G1312B). Includes electrical, environmental, hydraulic and flow performance data.
- Flow range 0.05 – 5.0 mL/min
- Pressure up to 600 bar
- Delay volume 120 µL (low) or 600–800 µL (standard)
- Line voltage 100–240 VAC
- Weight 15.5 kg
Explains delay volume and extra-column volume concepts and provides configurations for 2.1 mm, 3 mm and 4.6 mm i.d. columns. Includes tables of component delay volumes and recommendations for low, medium and standard delay volume setups.
- Low delay volume configuration (120 µL) for 2.1 mm i.d. columns
- Medium delay volume configuration with 200 µL mixer for highest UV sensitivity
- Standard delay volume configuration for 3 and 4.6 mm i.d. columns and method transferability
- Low dispersion kit (G1316-68744) required for 2.1 mm columns
- Enhanced Solvent Compressibility function must be selected in pump auxiliary screen
Describes how to increase injection volume using the Multidraw upgrade kit and considerations for method scaling and injection solvent strength.
- Standard maximum injection volume 100 µL
- Multidraw upgrade kit (G1313-68711) adds 400 µL or 1400 µL
- Total volume up to 500 µL or 1500 µL with 100 µL analytical head
- Delay volume increases when using extended seat capillaries
- Injection volume should be scaled with column volume
Provides rules of thumb for short run times, including column selection, flow rates, gradients and temperature. Also describes alternating column regeneration.
- Use short columns with good selectivity
- 3.0 mm i.d. columns ideal for UV detection
- Use standard delay volume for 3 and 4.6 mm i.d. columns
- Use low delay volume for 2.1 mm i.d. columns
- Alternating column regeneration with two columns, two pumps and a 2-position 10-port valve
Explains carryover measurement and autosampler features to minimize carryover, including needle wash and injection valve cleaning.
- Carryover figure of 0.002 % achievable
- Internal and external needle wash available
- Needle seat backflush and injection valve cleaning
- Automated delay volume reduction (ADVR) may increase carryover
- Use injection valve in mainpass position
Discusses parameters affecting resolution and provides optimum instrument configuration for high resolution separations.
- Resolution depends on efficiency, selectivity and retention factor
- Selectivity has the biggest impact on resolution
- Column temperature and mobile phase composition are critical
- Low extra-column volume is essential for small columns
Covers configuration for high sensitivity, including flow cell selection and detector settings.
- Max-Light high sensitivity cell with 60 mm path length available
- Standard Max-Light cartridge cell has 10 mm path length
- Low volume flow cell recommended for 2.1 mm i.d. columns
- Electronic temperature control in DAD reduces baseline drift
Provides recommendations to avoid column blockages, such as filtering mobile phases and using guard columns.
- Filter mobile phases and samples
- Use guard columns
- Avoid precipitation by maintaining solvent compatibility
Covers installation of Agilent Lab Advisor software and installation of the modules. Lab Advisor provides monitoring, diagnostics and maintenance feedback.
- Agilent Lab Advisor replaces and extends ChemStation diagnostic functions
- Lab Advisor monitors LC module status and Early Maintenance Feedback
- Automates tests and identifies LAN-based instruments
- Suggests replacement parts and troubleshooting tasks
Provides instructions for preparing the system, data acquisition in Method and Run Control view, and data analysis.
- Preparing the system includes solvent preparation and purging
- Data acquisition in Method and Run Control view
- Data analysis procedures
Contains safety information, solvent information, Agilent Technologies on Internet, and setting up a method using Edit Entire Method.
- Safety information and solvent compatibility
- Setting up a method using Edit Entire Method
- Agilent Technologies web resources
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| The binary pump flow range is 0.05 – 5.0 mL/min with settable increments of 0.001 mL/min. p.24 ↗ | Enables precise flow control for both narrow-bore and standard-bore applications. | 1260 Infinity Binary Pump (G1312B) |
| Maximum operating pressure is 600 bar (60 MPa) up to 5 mL/min. p.24 ↗ | Supports UHPLC methods with sub-2 µm particles and high flow rates. | 1260 Infinity Binary Pump (G1312B) |
| Standard delay volume configuration is 600 – 800 µL (includes 400 µL mixer); low delay volume configuration is 120 µL. p.24 ↗ | Low delay volume minimizes gradient delay for fast separations; standard configuration ensures method transferability. | 1260 Infinity Binary Pump (G1312B) |
| For flow rates below 500 µL/min or for use without damper and mixer, a vacuum degasser is required. p.25 ↗ | Prevents bubble formation and ensures reliable pump operation at low flows. | 1260 Infinity Binary LC system |
| The 1290 Infinity TCC controls temperature from 10 °C below ambient up to 100 °C at 2.5 mL/min and 80 °C at up to 5 mL/min. p.17 ↗ | Allows high-temperature separations and reduced solvent viscosity for faster analysis. | 1290 Infinity Thermostatted Column Compartment (G1316C) |
| TCC temperature stability is ±0.05 °C and accuracy is ±0.5 °C (with calibration). p.17 ↗ | Provides precise and reproducible column temperatures for retention time stability. | 1290 Infinity Thermostatted Column Compartment (G1316C) |
| The autosampler injection volume range is 0.1 up to 100 µL without changing sample loops. p.8 ↗ | Flexible injection volumes for various applications without hardware changes. | 1260 Infinity High Performance Autosampler (G1367E) |
| Carryover of 0.002 % should be achievable even with a triple quadrupole mass spectrometer. p.51 ↗ | Suitable for sensitive LC/MS analyses requiring low carryover. | 1260 Infinity High Performance Autosampler (G1367E) |
| The DAD wavelength range is 190 to 640 nm with a 1024-diode array and 4 nm slit. p.19 ↗ | Covers UV-Vis detection needs with high spectral resolution. | 1260 Infinity Diode Array Detector (G4212B) |
| The Max-Light cartridge cell has a 10 mm path length; a high sensitivity cell with 60 mm path length is available. p.19 ↗ | Choose the 60 mm cell for higher sensitivity, but consider dispersion effects. | 1260 Infinity Diode Array Detector (G4212B) |
| The DAD can output signals at 80 Hz and full-range spectra at 80 Hz simultaneously. p.20 ↗ | Enables accurate recording of narrow peaks and spectral data for peak purity. | 1260 Infinity Diode Array Detector (G4212B) |
| For 2.1 mm i.d. columns, use the low delay volume configuration and install the low dispersion kit (G1316-68744). p.43 ↗ | Minimizes extra-column volume and gradient delay for optimal performance with narrow-bore columns. | 1260 Infinity Binary LC system |
| For highest UV sensitivity with 2.1 mm i.d. columns, install the 200 µL low volume mixer (5067-1565). p.44 ↗ | Reduces mixing noise and improves baseline stability under gradient conditions. | 1260 Infinity Binary LC system |
| The Multidraw upgrade kit (G1313-68711) increases injection volume by 400 µL or 1400 µL, up to 500 µL or 1500 µL total. p.47 ↗ | Allows large-volume injections for trace analysis or preparative work. | 1260 Infinity High Performance Autosampler (G1367E) |
| When using extended seat capillaries from the multidraw kit, the delay volume of the autosampler is extended; double the volume of the extended capillaries for calculations. p.47 ↗ | Account for increased delay volume in method development to avoid retention time shifts. | 1260 Infinity High Performance Autosampler (G1367E) |
| Recommended pH range for the binary pump is 1.0 – 12.5; solvents with pH < 2.3 should not contain acids that attack stainless steel. p.24 ↗ | Ensures compatibility and longevity of pump components. | 1260 Infinity Binary Pump (G1312B) |
| The binary pump requires 100 – 240 VAC, 50 or 60 Hz, and consumes 220 VA maximum. p.23 ↗ | Can be used worldwide with standard power supplies; ensure adequate power availability. | 1260 Infinity Binary Pump (G1312B) |
| The pump weight is 15.5 kg and dimensions are 180 x 345 x 435 mm. p.23 ↗ | Plan bench space and ensure proper lifting. | 1260 Infinity Binary Pump (G1312B) |
| Up to three TCCs can be clustered to switch between eight columns for automated method development. p.18 ↗ | Enables high-throughput method scouting and column selection. | 1290 Infinity Thermostatted Column Compartment (G1316C) |
| The TCC door open sensor allows methods to require the door to be closed, important for low or high-temperature methods. p.18 ↗ | Ensures temperature stability and safety. | 1290 Infinity Thermostatted Column Compartment (G1316C) |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Composition precision | < 0.15 % RSD or < 0.04 min SD whatever is greater % RSD / min SD p.25 ↗ | at 0.2 and 1 mL/min; based on retention time at constant room temperature |
| Composition accuracy | ± 0.35 % absolute % p.25 ↗ | at 2 mL/min, at 10 MPa (100 bar) (water/caffeine tracer) |
| Analog output | 1.33 mV/bar, one output mV/bar p.25 ↗ | For pressure monitoring |
| Weight | 15.5 kg (34 lbs) kg p.23 ↗ | 1260 Infinity Binary Pump (G1312B) |
| Dimensions (height × width × depth) | 180 x 345 x 435 mm (7 x 13.5 x 17 inches) mm p.23 ↗ | 1260 Infinity Binary Pump (G1312B) |
| Line voltage | 100 – 240 VAC, ± 10 % VAC p.23 ↗ | Wide-ranging capability |
| Line frequency | 50 or 60 Hz, ± 5 % Hz p.23 ↗ | |
| Power consumption | 220 VA, 74 W / 253 BTU VA, W, BTU p.23 ↗ | Maximum |
| Ambient operating temperature | 4–55 °C (39–131 °F) °C p.23 ↗ | |
| Ambient non-operating temperature | -40 – 70 °C (-40 – 158 °F) °C p.23 ↗ | |
| Humidity | < 95 % r.h. at 40 °C (104 °F) % r.h. p.23 ↗ | Non-condensing |
| Operating altitude | Up to 2000 m (6562 ft) m p.23 ↗ | |
| Non-operating altitude | Up to 4600 m (15091 ft) m p.23 ↗ | For storing the module |
| Safety standards | IEC, CSA, UL p.23 ↗ | Installation category II, Pollution degree 2; For indoor use only |
| Hydraulic system | Two dual piston in series pumps with servo-controlled variable stroke drive, power transmission by gears and ball screws p.24 ↗ | 1260 Infinity Binary Pump (G1312B) |
| Settable flow range | 0.001 – 5 mL/min, in 0.001 mL/min increments mL/min p.24 ↗ | Set points |
| Flow range | 0.05 – 5.0 mL/min mL/min p.24 ↗ | |
| Flow precision | ≤0.07 % RSD or ≤0.02 min SD, whatever is greater % RSD / min SD p.24 ↗ | based on retention time at constant room temperature |
| Flow accuracy | ± 1 % or 10 µL/min, whatever is greater % / µL/min p.24 ↗ | pumping degassed H2O at 10 MPa (100 bar) |
| Pressure operating range | 0 – 60 MPa (0 – 600 bar, 0 – 8700 psi) MPa / bar / psi p.24 ↗ | up to 5 mL/min |
| Pressure pulsation | < 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar), whatever is greater % / MPa p.24 ↗ | at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi) |
| Pressure pulsation (low delay volume configuration) | < 5 % amplitude (typically < 2 %) % p.24 ↗ | Low delay volume configuration |
| Compressibility compensation | Pre-defined, based on mobile phase compressibility p.24 ↗ | |
| Recommended pH range | 1.0 – 12.5 pH p.24 ↗ | solvents with pH < 2.3 should not contain acids which attack stainless steel |
| Gradient formation | High-pressure binary mixing p.24 ↗ | |
| Delay volume (standard configuration) | 600 – 800 µL (includes 400 µL mixer) µL p.24 ↗ | Standard delay volume configuration; dependent on back pressure; measured with w |
| Delay volume (low configuration) | 120 µL µL p.24 ↗ | Low delay volume configuration; measured with water at 1 mL/min (water/caffeine |
| Composition range | settable range: 0 – 100 %; recommended range: 1 – 99 % or 5 µL/min per channel, whatever is greater % p.24 ↗ | |
| Delay volume (Binary Pump, low configuration) | 120 µL p.42 ↗ | in low delay volume configuration with bypassed damper and mixer |
| Delay volume (Binary Pump, standard configuration) | 600 – 800 µL p.42 ↗ | in standard delay volume configuration |
| Delay volume (Low volume mixer) | 200 µL p.42 ↗ | |
| Delay volume (Mixer) | 400 µL p.42 ↗ | |
| Delay volume (Autosampler) | 270 µL p.42 ↗ | |
| Delay volume (Low dispersion heat exchanger) | 1.6 µL p.42 ↗ | |
| Delay volume (Built-in heat exchanger) | 3 and 6 µL p.42 ↗ | |
| System delay volume (low delay volume configuration) | Pump: 120 µL p.42 ↗ | Low delay volume configuration |
| System delay volume (medium delay volume configuration) | Pump: 320 µL p.42 ↗ | Medium delay volume configuration |
| System delay volume (standard delay volume configuration) | Pump: 600 – 800 µL p.42 ↗ | Standard delay volume configuration |