Agilent 2200 TapeStation System Troubleshooting Manual
Technical Reference Summary
Covered Models & Document Scope
Covers the Agilent 2200 TapeStation System family: 2200 TapeStation (G2964AA), 2200 Nucleic Acid TapeStation (G2965AA), 2200 ScreenPlex TapeStation (G2966AA), and legacy Lab901 TapeStation variants (ST007, ST008, ST009, ST010, ST017, ST019). Includes DNA, RNA and Protein application troubleshooting, ScreenTape compatibility, USB communication, maintenance and error messages.
Section-by-Section Technical Breakdown
Covers intended use, performance limitations, tools and handling, mixing recommendations, reagent and sample handling, ScreenTape handling, instrument LED and temperature limits, consumables and manual marker assignment. These practices are prerequisites for reliable electrophoretic separation of nucleic acids and proteins.
- Maximum 16 samples per ScreenTape device; more samples require a 96-well plate and multiple ScreenTape.
- Use reverse pipetting for RNA and High Sensitivity RNA assays.
- Vortex mix with IKA MS3 default setting (1 minute) or manually 10 seconds at maximum speed.
- Equilibrate reagents and samples to room temperature for 30 minutes before use.
- Store ScreenTape at 2 – 8 °C; partially used ScreenTape max 2 weeks stored vertically.
- Blue LED: idle steady, slow flash during run, fast flash indicates an issue.
- Minimum 3 µL sample volume must be at the bottom of the tube without air bubbles.
Provides the minimum TapeStation software version required for each ScreenTape assay and instrument variant. Software is downloaded from the Agilent Genomics website; the instrument must not be connected to the laptop during installation or upgrade.
- P200 ScreenTape (5067-5371) requires A.01.01 on G2964AA and Lab901-ST010; not compatible with G2965AA, G2966AA and other legacy systems.
- Genomic DNA ScreenTape (5067-5365) requires A.01.03 on G2964AA, G2965AA, G2966AA, Lab901-ST010, ST017, ST019.
- D1000, HS D1000, RNA and HS RNA ScreenTape (5067-5582, 5067-5584, 5067-5576, 5067-5579) require A.01.04 on all compatible systems.
- Legacy Lab901-ST007, ST008 and ST009 are not compatible with any listed ScreenTape.
Describes power-up order, USB connection, USB driver update procedure, barcode reader requirements and log file retrieval. Each instrument exposes three USB devices: barcode reader, USB-to-serial converter and camera.
- Power up in order: laptop, instrument, controller.
- Power down laptops and instruments at night; save run files to a local folder on the TapeStation laptop.
- Barcode reader identified as 5180 Area Imager; USB-to-serial converter as USB↔Serial Converter.
- Camera: uEye UI-154x Series for G2965AA; uEye UI-224x Series for G2964AA.
- Place instrument away from direct light sources to avoid barcode recognition failure.
- Insert ScreenTape with label towards the front and barcode facing right.
- Export error logs via Help > Export Error Logs in the TapeStation Analysis Software.
Explains causes of blank lanes in the analysis file, ordered by probability, and their corrective actions. Blank lanes appear either as white lanes with no visible peaks or as black grainy lanes with very low sample signal.
- Ensure all 16 tips are loaded into the tip rack at the start of each run.
- Centrifuge sample strips or 96-well plates after preparation to remove bubbles at the bottom of wells.
- Ensure a minimum of 3 µL of solution in the sample tubes; do not re-use sample strips.
- Ensure positions selected in the controller software match the sample strip locations.
- Remove sample strip lids before starting the run to avoid instrument damage.
- If all causes are ruled out, a hardware fault may exist; contact local support.
Covers good measurement practices for DNA quantification, sizing and molarity, including mixing effects, peak integration, marker identification, Genomic DNA assay specifics and troubleshooting tables for incorrect quantification, incorrect sizing and missing marker peaks.
- Insufficient mixing dramatically lowers reported concentration: 68.5 ng/µL with vortex mixing vs 21.2 ng/µL with pipette mixing only vs 6.5 ng/µL with no mixing (theoretical 70 ng/µL).
- In DNA assays concentration is calculated from sample peak area relative to the known upper marker concentration; RNA, HS RNA and Genomic DNA use the lower marker.
- Electropherogram view reports the highest point of the peak; Region view reports the center of the region's mass.
- Genomic DNA reagents must equilibrate to room temperature for 30 minutes; cold reagents overestimate size (+43 % at 4 °C vs +11 % at RT).
- Prepare Genomic DNA ladder fresh for each run and run it in the first available position.
- Do not shake the Genomic DNA ladder vial; inversion 30 times degrades the top 48,500 bp fragment and can prevent DIN scoring.
- For SureSelect, increase magnetic plate incubation to 10 minutes to remove residual AMPure XP beads that run with the upper marker.
Covers troubleshooting of RNA assays including incorrect quantification, incorrect sizing, missing marker peaks, degraded RNA ladder and/or samples, incorrect or missing RINe value, incorrect or missing peak annotation and unexpected migration profile.
- RNA and High Sensitivity RNA assays require reverse pipetting.
- RNA ScreenTape and HS RNA ScreenTape require software A.01.04.
- RNA assays contain only a lower marker.
Covers buffer compatibility, quantification of protein samples, incorrect quantification and sizing results and unexpected migration profiles for protein assays on the 2200 TapeStation.
- Protein assays require a heating block or PCR machine (not supplied).
- P200 ScreenTape is used for protein applications and requires software A.01.01 on G2964AA.
Provides general maintenance information, the procedure for changing the needle and the preventative maintenance interval.
- Needle replacement procedure is described in this section.
- Preventative maintenance interval is specified in this section.
Lists analysis software warning messages, communication error messages and consumable error messages with their meanings and corrective actions.
- Communication error: 'Failed to connect to the barcode scanner or camera' indicates a USB connection or driver issue.
- Temperature out of range error appears when the instrument is outside the recommended range for the inserted ScreenTape.
- Consumable error messages relate to ScreenTape and tip handling.
Lists ScreenTape products and the parts and consumables required for the 2200 TapeStation System, including loading tips, sample strips, plates, foil seals and the shipping box.
- Loading tips: 5067-5152 or 5067-5153.
- Optical Tube 8x Strip 401428 and Optical Cap 8x Strip 401425, or 96-well Sample Plates 5067-5150 and 96-well Plate Foil Seal 5067-5154.
- Shipping box part number G2960-60100.
- IKA MS3 vortexer part number 4674100 (obtainable from IKA quoting 'Agilent Technologies').
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| The 2200 TapeStation can analyze a maximum of 16 samples on a single ScreenTape device. p.8 ↗ | Higher throughput requires a 96-well plate and multiple ScreenTape devices. | whole series |
| The instrument will load a sample from a minimum of 3 µL onto ScreenTape. p.12 ↗ | Sample tubes must contain at least 3 µL at the bottom without air bubbles to avoid blank lanes. | whole series |
| Optimal instrument operating temperature is 20 °C; assay-specific ranges are 12–37 °C (D1000), 17–37 °C (HS D1000), 15–30 °C (Genomic DNA), 14–30 °C (RNA/HS RNA) and 10–33 °C (P200). p.15 ↗ | Operating outside the assay-specific range triggers a 'Temperature out of Range' error and invalidates the run. | whole series |
| ScreenTape must be stored at 2 – 8 °C and can be used straight from the fridge with no equilibration time. p.13 ↗ | No warm-up time is needed, but partially used ScreenTape must be returned to 2 – 8 °C vertically and used within 2 weeks. | whole series |
| Partially used ScreenTape must be used with the same laptop that performed previous runs because run data is stored locally on the instrument laptop. p.14 ↗ | Changing or updating the laptop can cause lane usage information to be lost and partially used lanes to be reused. | whole series |
| Vortex mixing with the IKA MS3 at 2,000 rpm for 1 minute followed by brief centrifugation yields 68.5 ng/µL vs 21.2 ng/µL for pipette mixing only and 6.5 ng/µL for no mixing (theoretical 70 ng/µL). p.41 ↗ | Insufficient mixing dramatically underestimates concentration; the recommended vortex protocol is mandatory for accurate quantification. | D1000 ScreenTape assay |
| Genomic DNA reagents equilibrated at 4 °C overestimate sizing by +43 % (24369 bp vs expected 17000 bp), while RT-equilibrated reagents give +11 % (18867 bp). p.50 ↗ | Genomic DNA reagents must be equilibrated to room temperature for 30 minutes before use. | Genomic DNA ScreenTape assay |
| Shaking the Genomic DNA ladder vial by inverting 30 times degrades the top 48,500 bp fragment and can prevent DIN scores from being presented. p.51 ↗ | Handle the ladder vial carefully before adding sample buffer; after mixing with sample buffer vortex gently for 5 seconds. | Genomic DNA ScreenTape assay |
| Genomic DNA reagents are shipped frozen on dry ice and must be kept at 2 – 8 °C in the refrigerator upon receipt. p.51 ↗ | Storage at 2 – 8 °C preserves reagent integrity and sizing accuracy. | Genomic DNA ScreenTape assay |
| In the SureSelect workflow, residual AMPure XP beads run with the upper marker and lower the reported sample concentration; increasing magnetic plate incubation to 10 minutes removes a higher percentage of beads. p.46 ↗ | Extend magnetic plate incubation to 10 minutes when using SureSelect to avoid concentration bias. | SureSelect workflow on 2200 TapeStation |
| In DNA assays concentration is calculated from sample peak area relative to the known upper marker concentration; RNA, HS RNA and Genomic DNA assays use the lower marker. p.43 ↗ | Peak integration must be manually adjusted when necessary to encompass the whole peak, otherwise concentration is biased. | whole series |
| Electropherogram view reports the highest point of the peak, while Region view reports the center of the region's mass. p.47 ↗ | Select the analysis mode appropriate to the sample type (discrete peaks vs smears/NGS libraries) to obtain the intended sizing value. | whole series |
| Power-up order must be laptop, then instrument, then controller; laptops and instruments must be powered down at night. p.24 ↗ | Following the power sequence avoids communication failures between laptop and instrument. | whole series |
| Each instrument exposes three USB devices: 5180 Area Imager (barcode reader), USB↔Serial Converter, and uEye UI-154x (G2965AA) or uEye UI-224x (G2964AA) camera. p.27 ↗ | A missing or exclamation-marked device in Device Manager indicates a driver update is required. | G2964AA, G2965AA |
| The instrument must be placed away from direct light sources because excess light interferes with barcode recognition. p.33 ↗ | Direct sunlight or strong lighting can cause a non-functioning instrument due to barcode read failure. | whole series |
| Before transport, enable 'transport mode' in the controller software via the Transport drop-down menu and remove the sample plate, tip holder and ScreenTape. p.16 ↗ | Failure to enable transport mode or remove components before transport can damage the instrument. | whole series |
| Shipping boxes for transport can be purchased using part number G2960-60100. p.16 ↗ | Use the correct packaging when transporting the instrument. | whole series |
| The IKA MS3 vortexer (part number 4674100) is bundled with 2200 TapeStation instruments (G2965AA or G2964AA) and includes a 96-well plate adaptor for 96-well PCR plates and 8-way strips. p.10 ↗ | Use the IKA MS3 with adaptor for the recommended vortex mixing protocol; it is not sold separately by Agilent. | G2964AA, G2965AA |
| Loading tips are 5067-5152 or 5067-5153; sample strips are Optical Tube 8x Strip 401428 with Optical Cap 8x Strip 401425, or 96-well Sample Plates 5067-5150 with 96-well Plate Foil Seal 5067-5154. p.17 ↗ | Only Agilent 2200 TapeStation approved consumables must be used to avoid instrument damage and performance impact. | whole series |
| The Genomic DNA ladder must be prepared fresh for each run and run in the first available position because the run profile decreases as the ladder warms up within the instrument. p.50 ↗ | Reusing or repositioning the ladder affects sizing results. | Genomic DNA ScreenTape assay |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Temperature out of Range p.15 ↗ | Instrument temperature outside the recommended range for the inserted ScreenTape. | If above range, move the system out of direct sunlight and away from windows and check air conditioning. If below range, allow the instrument to equilibrate to ambient temperature and avoid using in a cooled area. | Figure 3 |
| Failed to connect to the barcode scanner or camera p.26 ↗ | Laptop fails to connect to the instrument or loses connection; USB cable or driver issue. | Check that power and USB cables are attached to the instrument; if issues remain, re-install the USB drivers. | Updating USB Drivers |
| Blank lane (white lane with no visible peaks) p.37 ↗ | Tips in the selected positions were missing from the tip rack. | Always ensure that all 16 tips are loaded into the tip rack at the start of each run. | Table 2 |
| Blank lane (white lane with no visible peaks) p.37 ↗ | A bubble in the sample tube prevented the sample from being picked up. | Always ensure that sample strips or 96-well plates are centrifuged after preparation to remove any bubbles present at the bottom of wells. | Table 2 |
| Blank lane (white lane with no visible peaks) p.37 ↗ | Insufficient sample present. | Always ensure a minimum of 3 µL of solution is present in the sample tubes. Do not re-use sample strips from previous runs. | Table 2 |
| Blank lane (white lane with no visible peaks) p.37 ↗ | The samples were not selected correctly in the controller software. | Always ensure that the positions selected in the controller software match the location of the sample strips within the instrument. | Table 2 |
| Blank lane (white lane with no visible peaks) p.37 ↗ | The sample lids were not correctly removed before starting the run. | Always ensure that sample strip lids are removed before starting the run. Failure to do so can result in damage to the instrument. | Table 2 |
| Blank lane (black grainy lane with very small sample intensity values) p.37 ↗ | Hardware fault. | If the above causes have been ruled out, send all relevant data to your local support representative for investigation. | Table 2 |
| Incorrect quantification results p.53 ↗ | Insufficient mixing. | Ensure correct mixing by reading the mixing recommendations. | Table 5 |
| Incorrect quantification results p.53 ↗ | Sample concentration outside recommended range for application. | Either dilute or concentrate your sample until it is within the recommended range for application as stated in the assay Quick Guide, then prepare a new run. | Table 5 |
| Incorrect quantification results p.53 ↗ | Incorrect peak integration. | Ensure that all sample peaks and markers are integrated correctly in the TapeStation Analysis Software, by clicking and dragging so that the whole peak is encompassed. | Table 5 |
| Incorrect quantification results p.53 ↗ | Some sample has run concurrently with the upper marker. | Ensure your sample is within the recommended sizing range for the application. If using SureSelect Protocol, ensure that all AMPure beads are removed by increasing the time your sample is on the magnetic plate. | Table 5 |
| Incorrect quantification results p.53 ↗ | Incorrect protocol used. | Ensure the correct sample protocol for the application is used. Differences exist between standard and High Sensitivity assay protocols; consult the User Manual or appropriate Quick Guides. | Table 5 |
| Incorrect quantification results p.53 ↗ | Sample prepared too long before analysis. | To avoid evaporation or settling, ensure that your run is started immediately after your sample preparation. If using a 96-well plate, always cover with the recommended foil seal. | Table 5 |
| Incorrect quantification results p.53 ↗ | Incorrect pipetting technique or pipette calibration. | Follow guidelines in the essential measurement practices section. | Table 5 |
| Incorrect quantification results p.53 ↗ | Incorrect marker peaks picked up. | Use the Aligned button in the TapeStation Analysis Software to detect the correct markers, then assign them by right clicking. Ensure that all sample peaks are within the recommended sizing range for the application. | Table 5 |
| Incorrect sizing results p.55 ↗ | Incorrect marker peaks picked up. | Ensure the correct marker peaks are picked up in the TapeStation Analysis Software. Ensure that all sample peaks are within the recommended sizing range for the application. | Table 6 |
| Incorrect sizing results p.55 ↗ | Incorrect analysis software mode. | Reported sizing can differ between electropherogram and region modes. Select the correct mode for your samples. | Table 6 |
| Incorrect sizing results p.55 ↗ | Issues with the ladder lane. | Any issues with the run ladder lane will be annotated within the TapeStation Analysis Software sample table, under Observations. Manually insert or delete affected peaks, or insert a software saved ladder to regain sizing information (if available). | Table 6 |
| Incorrect sizing results p.55 ↗ | Insufficient migration due to presence of salt. | Dilute your samples to ensure low levels of buffer salt. True migration profiles can be seen by unaligning your gel image by pressing the Aligned button. Refer to the salt tolerance guidelines for your assay. | Table 6 |
| Missing marker peaks p.57 ↗ | Insufficient mixing. | Ensure correct mixing by reading the mixing recommendations. | Table 7 |
| Missing marker peaks p.57 ↗ | Sample concentration outside recommended range for application. | Either dilute or concentrate your sample until it is within the recommended range for application as stated in the assay Quick Guide, then prepare a new run. | Table 7 |
| Missing marker peaks p.57 ↗ | Blank lane. | See Blank lane section. | Table 7 |
| Missing marker peaks p.57 ↗ | Insufficient migration. | To check migration true distance, unalign the gel image using the Aligned button. | Table 7 |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Maximum samples per ScreenTape device | 16 samples p.8 ↗ | |
| Minimum sample volume loaded onto ScreenTape | 3 µL p.12 ↗ | |
| Optimal instrument operating temperature | 20 °C p.15 ↗ | |
| Assay operating temperature | 12 – 37 °C p.15 ↗ | D1000 ScreenTape |
| Assay operating temperature | 17 – 37 °C p.15 ↗ | High Sensitivity D1000 ScreenTape |
| Assay operating temperature | 15 – 30 °C p.15 ↗ | Genomic DNA ScreenTape |
| Assay operating temperature | 14 – 30 °C p.15 ↗ | RNA and High Sensitivity RNA ScreenTape |
| Assay operating temperature | 10 – 33 °C p.15 ↗ | P200 ScreenTape |
| ScreenTape storage temperature | 2 – 8 °C p.13 ↗ | |
| Maximum storage time for partially used ScreenTape | 2 weeks p.14 ↗ | stored vertically at 2 – 8 °C |
| Reagent equilibration time to room temperature | 30 minutes p.11 ↗ | |
| Sample equilibration time to room temperature | 30 minutes p.12 ↗ | |
| Vortex mixing time with IKA MS3 default setting | 1 minute p.10 ↗ | D1000, HS D1000, RNA, HS RNA ScreenTape assays |
| Manual vortex mixing time at maximum speed | 10 seconds p.10 ↗ | if IKA MS3 not available |
| Vortex mixing speed for recommended protocol | 2,000 rpm p.41 ↗ | IKA vortexer with adaptor, 1 minute |
| Genomic DNA ladder vortex mixing time after addition of sample buffer | 5 seconds p.51 ↗ | gentle vortex |
| Magnetic plate incubation time for SureSelect AMPure bead removal | 10 minutes p.46 ↗ | SureSelect workflow |
| Genomic DNA sizing accuracy with reagents at RT | +11 % p.50 ↗ | expected 17000 bp, measured 18867 bp |
| Genomic DNA sizing accuracy with reagents at 4 °C | +43 % p.50 ↗ | expected 17000 bp, measured 24369 bp |
| Measured concentration with vortex mixing then centrifugation | 68.5 ng/µL p.42 ↗ | D1000 assay, theoretical 70 ng/µL |
| Measured concentration with pipette mixing only | 21.2 ng/µL p.42 ↗ | D1000 assay, theoretical 70 ng/µL |
| Measured concentration with no mixing | 6.5 ng/µL p.42 ↗ | D1000 assay, theoretical 70 ng/µL |