Agilent 240 Ion Trap GC/MS Hybrid Ionization Users Guide
Technical Reference Summary
Covered Models & Document Scope
Covers the Hybrid Ionization configuration of the Agilent 240 Ion Trap GC/MS system, including Hybrid PCI and Hybrid NCI modes, Hybrid HPS (High Pressure Source) and LPS (Low Pressure Source) options, and mode conversion from Internal or External configurations. Requires external ionization option, chemical ionization option, and a security chip.
Section-by-Section Technical Breakdown
Describes the Hybrid configuration of the 240 Ion Trap GC/MS, where reagent ions are generated in the external source by EI of a reagent gas and drawn into the ion trap to react with analytes. Hybrid CI is a softer ionization technique than EI, producing less fragmentation and more molecular weight information. The mode requires the external ionization option, chemical ionization option, and a security chip.
- Hybrid configuration is one of three operational configurations of the 240 Ion Trap GC/MS.
- Reagent ions are generated in the external source through EI of a reagent gas.
- Advantages include avoiding ion-molecule reactions with neutral reagent and avoiding negative ion losses.
- Hybrid CI imparts less energy than EI, reducing fragmentation.
- Requires external ionization option, chemical ionization option, and a security chip.
- Supports SIS, and with optional software/equipment MS/MS, MSn, AMD, and MRM.
Explains how to use the View Reagent Ions function to tune the isolation of individual reagent ions. When no Ejection Amplitude is specified, all ions in the Reagent Ion Start and End Masses range can be viewed.
- Use View Reagent Ions to tune isolation of individual reagent ions.
- Without Ejection Amplitude, all ions in the range are viewed.
- Schematic shows E.M. end cap, filament end cap, lens structure, CI volume assembly, filament, transfer line tip, ring electrode, electron multiplier, and H.E. dynode.
Explains mass analysis by ramping RF voltage on the ring electrode, ejection of ions to the conversion dynode and electron multiplier, and AGC control of ionization time. The conversion dynode is held at -10,000 V in positive modes and +10,000 V in negative mode; signal is amplified by ~10^5.
- Ions are ejected from low to high mass by increasing RF voltage on the ring electrode.
- Supplemental dipole and quadrupole voltages improve mass resolution.
- Conversion dynode held at -10,000 V in positive modes, +10,000 V in negative mode.
- Signal amplified by ~10^5 by the electron multiplier.
- AGC prescan and up to six analytical scan segments.
- Ion trap has a maximum storage capacity beyond which resolution and spectral quality deteriorate.
Notes that no libraries of Hybrid PCI or NCI mass spectra are included with 240 MS software, but users can create libraries. Discusses selectivity advantages of Hybrid CI, including poor hydrocarbon response in methane PCI and good NCI response for high electron affinity species.
- No Hybrid PCI or NCI libraries included with 240 MS software.
- Users can create libraries of Hybrid spectra.
- Hydrocarbons have poor response in methane PCI.
- Negative CI gives good response only for species with high electron affinity such as halogenated compounds.
- Analyzing samples with multiple ionization and ion preparation options is recommended during method development.
Provides instructions for installing methane CI gas and adjusting CI gas flow. Connect the regulator through a 50 mL/min restrictor, set the second stage to 20 psi, and adjust the CI gas valve until ion gauge pressure is within 70 to 100 µTorr.
- Connect gas cylinder regulator to the back of the instrument through a 50 mL/min restrictor.
- Set methane regulator second stage to 20 psi.
- Use CI Gas Adjust Valve inside the front door of the 240 MS.
- Turn knob clockwise to increase flow, counterclockwise to decrease.
- Set ion gauge pressure within 70 to 100 µTorr.
- Liquid reagents like methanol and acetonitrile give weak responses in Hybrid PCI.
Describes starting acquisition, checking readiness of GC and autosampler, and running single samples or sample lists. The software automatically shifts the MS module into Acquisition mode. Status indicators include Run Time, End Time, and Ready/No Fault lights.
- Click Start Acquisition to start a run.
- Software automatically shifts MS module into Acquisition mode.
- Not Ready message changes to Ready when GC and AutoSampler are ready.
- Individual ready states visible under Windows pull-down menu for 240 MS, 7890 GC, and sampler modules.
- Before acquisition, Run Time is 0.00 minutes and Ready/No Fault lights are green.
Step-by-step instructions for injecting a single sample, including entering sample name, vial number, injection volume, and data file information. For manual injection, wait for the System Control title bar to read 'Waiting for Injection of Sample' and a blinking yellow Waiting light.
- Click Inject Single Sample from the Inject menu.
- Enter sample name, vial number, injection volume, and injector.
- Click Defaults to change default values.
- Click Data Files to create a name with date/time or change storage directory.
- For manual injection, wait for 'Waiting for Injection of Sample' and blinking yellow Waiting light.
Explains creating and editing a Sample List in the Automation File Editor or System Control, and injecting multiple samples. The Sample Name column does not scroll when the table is scrolled right.
- Create and edit Sample List in Automation File Editor or System Control.
- Click New Sample List or Open Sample List from File menu.
- Sample List window contains fields specific to the configured autosampler.
- Sample Name column remains visible when scrolling right.
- Click Begin in lower left corner to start the Sample List.
Describes monitoring run status in the instrument window and viewing chromatogram and spectra in System Control or MS Data Review.
- Monitor status of the run in the instrument window.
- Monitor chromatogram and spectra in System Control or MS Data Review.
- See Acquiring GC/MS Data section in the 240 Ion Trap GC/MS Software Operation help.
Lists checks before initial pump down: vacuum connections leak free, transfer line seated in trap, vent valve closed fully clockwise, column not broken. The roughing pump should stop gurgling after about 10 to 20 seconds.
- Verify vacuum connections are leak free.
- Verify transfer line is seated in the trap.
- Verify vent valve is closed fully clockwise.
- Verify column is not broken.
- Roughing pump should stop gurgling after about 10 to 20 seconds.
- If pump continues to gurgle, check analyzer assembly seating, transfer line seating, and vent valve seal.
Provides typical operating ranges for the 240 MS in External mode and explains diagnosing leaks. Large leaks are indicated by turbo speed less than 100%; small leaks by increased pump current or ion gauge pressure.
- Typical turbo speed: 100%.
- Typical pump current: 200 to 300 mAmps.
- Typical power: 9 to 13 Watts.
- Typical ion gauge pressure: < 20 µTorr.
- Typical roughing line pressure: < 50 mTorr.
- Large leaks indicated by turbo speed less than 100%.
- Small leaks show as increased pump current or ion gauge pressure.
- Start without damping gas, then increase to 0.5 mL/min to test sensitivity.
Describes monitoring instrument state via the Monitoring tab and performing hardware checks via the Diagnostics tab. Helium buffer gas flow initial choice is 3 to 4 mL/min, set in the Module Attributes tab.
- Monitor vacuum system, electron multiplier, waveform system, temperatures, and ion source.
- Perform hardware checks on the 240 MS using the Diagnostics tab.
- Initial helium buffer gas flow choice is 3 to 4 mL/min.
- Set helium buffer gas flow rate in the Module Attributes tab.
Covers analysis temperatures, system bakeout, and temperature effects on calibration. Ion trap temperature must be high enough to prevent condensation. Transfer line temperature should be 20 °C below the maximum column temperature. Default manifold temperature is typically 50 °C.
- Ion trap temperature must prevent condensation of analyte eluting into the trap.
- Transfer line temperature should be 20 °C below maximum column temperature.
- Default manifold temperature is typically 50 °C.
- Perform mass calibration and trap function calibration shortly after source temperature is reached and again several hours later.
- Wait at least two hours after bakeout before AutoTune or running the 240 MS.
- Transfer line temperature should not exceed maximum isothermal temperature of the column.
Describes using Startup/Shutdown to safely start or shut down the system. During startup, software is locked until turbo pump speed reaches 100%. Shutdown reduces turbo speed to 35% of full speed; after all zones are below 80 °C, turn off main power and vent for at least 5 minutes.
- Software locked in Startup/Shutdown mode until turbo speed reaches 100%.
- Failure to reach 100% pump speed in reasonable time indicates a vacuum leak.
- Shutdown reduces turbo pump speed to 35% of full speed.
- After all temperature zones are below 80 °C, turn off main power at rear switch.
- Manually vent the system for at least 5 minutes using the front panel valve.
Covers running diagnostic tests, RF tune, RF ramp adjustment, CI gas adjustment, air/water check, Auto Tune, integrator zero, electron multiplier setting, electron lens tuning, ion lens tuning, RF full scale adjust, mass calibration, trap frequency calibration, and trap DC offset voltage.
- Retract transfer line before lifting analyzer assembly to avoid damage.
- Run all routines in Diagnostic Tests tab after turbo reaches 100% speed.
- Adjust RF tuning after MS maintenance, changing analyzer assembly, or changing MS configuration.
- RF ramp adjustment: turn RF Adjustment screw until tuning display shows straight line and intensity at minimum.
- Air/Water Check uses electron multiplier voltage with gain at 10^5.
- Auto Tune works the same in EI or Hybrid CI modes.
- Electron lens balance range is 200 to 300 µA.
- For CI methods in Hybrid mode, electron/repeller lens must be tuned with CI plunger in place and CI gas on.
- Ion lens tuning uses Cal Gas ions at m/z 131 and 414.
- Mass calibration uses PFTBA ions at m/z 69, 131, 264, 414, 464, and 614.
- Do not run mass calibration until ion trap temperature has stabilized for at least two hours.
- Trap frequency calibration determines parameters for MS/MS and SIS.
- Trap DC offset optimizes ion signal for m/z 414.
Describes Hybrid CI scan functions where reagent ions are created during the ionization pulse and consumed during the reaction period. AGC prescan results are used to calculate ionization time and reaction time for the analytical scan, enabling linear calibration curves.
- Ion trap operates in pulsed mode.
- Reagent ions created only during ionization pulse and consumed during reaction period.
- Number of analyte ions depends on analyte concentration, initial reagent ion intensity, and reaction time.
- AGC prescan calculates ionization time and reaction time for analytical scan.
- Spectral intensity proportional to sample concentration and reaction time.
Step-by-step instructions for creating a new method using the Method Builder wizard, including selecting instrument, detectors, data channels, and post-run processing. The method contains sections for 7890 GC Control, 240 MS Control, Standard MS Reports, and MS Data Handling.
- Click Method Builder icon on Workstation Toolbar.
- Click Create a New Method File.
- Select Instrument 1 and click Next.
- Use Custom configuration to create methods on a PC remote from the instrument.
- Select detectors and data channels for post-run processing.
- Method contains default values for all parameters.
- In Hybrid Mode, instrument configuration is Chemical Ionization (CI).
Explains that Hybrid methods can be performed only in PCI or NCI modes except when running Auto Tune methods requiring EI mode. Hybrid HPS (High Pressure Source) is performed with the CI volume inserted; LPS (Low Pressure Source) occurs in the EI source. Centroid data is default; profile data is for diagnostics and approximately 10 times larger.
- Hybrid methods only in PCI or NCI modes, except Auto Tune methods requiring EI.
- Hybrid HPS performed with CI volume inserted into ion source.
- Hybrid LPS occurs in the EI source.
- Centroid data is default acquisition data type.
- Data handling, library searching, and spectral comparison only with centroid data.
- Profile data collected at 10 points per m/z and approximately 10 times larger than centroid files.
- Up to 250 time segments for runs up to 650 minutes.
Describes editing parameters for Hybrid CI methods, including scan function settings, general parameters, ionization control, hybrid parameters, and scan parameters. Three scan modes are available: Normal (5000 u/sec), Fast (10000 u/sec), and Fastest (10000 u/sec, no prescan, Full scan only).
- CI is the only ionization choice in Hybrid configuration.
- Normal scan mode uses prescan and scans at 5000 u/sec.
- Fast scan mode uses prescan and scans at 10000 u/sec.
- Fastest scan mode uses no prescan and scans at 10000 u/sec; only available in Full scan type.
- Scan Time, Scans Averaged, and Data Rate are linked.
- Setting scans averaged to three gives best compromise between data rate and spectral averaging.
- Mass Defect allows systematic correction of nominal vs exact mass.
- Multiplier Offset adjusts EM voltage by up to 300 V relative to current setting.
- Count Threshold normally 1; value of 2 or 3 reduces low-level ions reported.
- Target TIC usually not below 10,000 for full scan; typically 20,000 to 40,000 gives best results.
- Default Target TIC for CI is 5,000; maximum 65,000.
- Reagent Low Mass must be set at least 10u below lowest reagent ion of interest.
- Default Ejection Amplitude is 15 V.
- Max Reaction Time range is 1 to 2000 µsec.
- Reaction Storage Level should not be set above m/z of CI reagent ion.
- CI Background Mass is lowest m/z counted during CI prescan.
Covers setting full scan and MS/MS parameters. Full scan allows up to six non-contiguous mass ranges separated by at least 10u. MS/MS parameters include Precursor Ion, Isolation Window (1.0 to 14.0 m/z, default 3.0), Low/High Edge Offset (-0.5 to 0.5 m/z), High Mass Ejection (default 35 V), and Waveform Type (resonant or non-resonant).
- Up to six non-contiguous mass ranges separated by at least 10u for full scan.
- MS/MS isolation window range 1.0 to 14.0 m/z; default 3.0 m/z.
- If isolation window smaller than 1.5 m/z, enter exact precursor ion mass.
- Low Edge Offset range -0.5 to 0.5 m/z; default 0.
- High Edge Offset range -0.5 to 0.5 m/z; default 0.
- Adjust edge offsets initially in 0.2 m/z increments.
- High Mass Ejection default 35 volts.
- Waveform type is either resonant or non-resonant.
- MS/MS useful only when target compounds are known.
Covers conversion from Internal to Hybrid and External to Hybrid configurations, and effects of hardware changes.
- Internal to Hybrid conversion.
- External to Hybrid conversion.
- Effects of hardware changes.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Hybrid mode requires the external ionization option, chemical ionization option, and a security chip, but does not involve any unique hardware. p.5 ↗ | Verify option licenses and security chip are installed before attempting Hybrid operation. | Hybrid configuration |
| Default RF storage level is 35u, so CI reagent ions at m/z 17 and 29 are not stored; only reagent ions above 35u react with analyte molecules. p.8 ↗ | This inherent selectivity can be used to avoid unwanted low-mass reagent ions; adjust Reagent Low Mass to change the storage cutoff. | Hybrid mode |
| Helium buffer gas flow rate of 1 mL/min is used for ion storage; initial choice for starting the instrument is 3 to 4 mL/min. p.9 ↗ | Set helium flow in Module Attributes tab; trapping efficiency is compound dependent. | Whole series |
| Methane CI gas is connected through a 50 mL/min restrictor and the regulator second stage is set to 20 psi. p.13 ↗ | Use the specified restrictor and pressure to achieve correct CI gas flow. | Methane CI installation |
| CI gas adjustment target is ion gauge pressure within 70 to 100 µTorr. p.13 ↗ | Adjust CI Gas Adjust Valve until adjustment result is OK before acquiring data in Hybrid CI mode. | Hybrid CI |
| Conversion dynode is held at -10,000 V in positive modes and +10,000 V in negative mode; signal is amplified by ~10^5 by the electron multiplier. p.11 ↗ | Understand polarity-dependent dynode voltage for method setup and troubleshooting. | Whole series |
| Typical operating ranges in External mode: turbo speed 100%, pump current 200 to 300 mAmps, power 9 to 13 Watts, ion gauge pressure < 20 µTorr, roughing line < 50 mTorr. p.20 ↗ | Use these ranges to diagnose vacuum system health; deviations indicate leaks or pump issues. | 240 MS in External mode |
| Transfer line temperature should be 20 °C below the maximum column temperature of the active method and must not exceed the maximum isothermal temperature of the column. p.22 ↗ | Set transfer line temperature to avoid cold spots while protecting the column. | Whole series |
| After bakeout, wait at least two hours before attempting AutoTune or running the 240 MS. p.22 ↗ | Allow all temperature zones to equilibrate thoroughly to avoid calibration drift. | Whole series |
| During shutdown, turbo pump speed is reduced to 35% of full speed; after all temperature zones are below 80 °C, turn off main power and vent for at least 5 minutes. p.24 ↗ | Follow this sequence to safely shut down and vent the system. | Whole series |
| Electron lens balance range is 200 to 300 µA; if outside this range, auto tune searches optimal values by changing four variables one at a time. p.28 ↗ | If auto tune fails to find best lens voltage, it generates an error and restores last values. | Whole series |
| For CI methods in Hybrid mode, the electron/repeller lens must be tuned with the CI plunger (CI volume) in place and the CI gas turned on. p.28 ↗ | Adjust CI gas flow in Manual Control before performing electron lens tuning. | Hybrid CI |
| Mass calibration uses PFTBA ions at m/z 69, 131, 264, 414, 464, and 614; do not run until ion trap temperature has stabilized for at least two hours. p.29 ↗ | Temperature changes can shift the mass calibration axis; stabilize before calibration. | Whole series |
| Normal scan mode scans at 5000 u/sec; Fast and Fastest modes scan at 10000 u/sec; Fastest uses no prescan and is only available in Full scan type. p.37 ↗ | Choose scan mode based on required data rate and spectral quality. | Whole series |
| Default Target TIC for CI is 5,000 counts; typical best results at 20,000 to 40,000 counts; maximum 65,000 counts. p.39 ↗ | Setting Target TIC too high can cause space charge distortions (loss of resolution or mass shift). | Hybrid CI |
| Reagent Low Mass must be set at least 10u below the mass of the lowest reagent ion of interest without loss of intensity. p.40 ↗ | This sets the RF storage level to exclude ions below the selected m/z. | Hybrid parameters |
| Default Ejection Amplitude for high mass isolation of CI reagent ions is 15 V; Max Reaction Time range is 1 to 2000 µsec. p.40 ↗ | Adjust these parameters to optimize reagent ion isolation and reaction time. | Hybrid parameters |
| MS/MS Isolation Window range is 1.0 to 14.0 m/z with default 3.0 m/z; if smaller than 1.5 m/z, enter exact precursor ion mass. p.42 ↗ | Adjust isolation window and edge offsets to isolate precursor ions effectively. | MS/MS |
| High Mass Ejection default amplitude is 35 volts; reducing it may help if precursor ions are lost due to dissociation. p.43 ↗ | Balance ejection amplitude to retain precursor ions while removing higher-mass ions. | MS/MS |
| Up to 250 chromatographic time segments can be created for runs up to 650 minutes in length. p.36 ↗ | Time program analysis conditions to optimize data for each segment. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Pump Spin Speed does not increase steadily p.20 ↗ | There may be a leak in the system. | Diagnose leaks by observing changes in ion gauge reading and pinpoint using leak check section in method Service.mth. See Troubleshooting section in 240 GC/MS Ion Trap Hardware Operation Manual. | Check the vacuum status |
| Turbo speed less than 100% p.20 ↗ | Large leaks in the system. | Check vacuum connections, transfer line seating, vent valve seal, and analyzer assembly seating. | Check the vacuum status |
| Increase in pump current after at 100% or in ion gauge pressure p.20 ↗ | Small leaks in the system. | Diagnose small leaks by observing changes in ion gauge reading and pinpoint using leak check section in method Service.mth. | Check the vacuum status |
| Roughing pump continues to gurgle after about 10 to 20 seconds p.19 ↗ | Analyzer assembly not seated properly, transfer line not seated in trap, or vent valve not sealed. | Verify analyzer assembly is seated properly on manifold (no gaps), transfer line is seated in trap, and vent valve is sealed. | Initial Pump down |
| Failure to reach 100% pump speed in a reasonable time p.23 ↗ | Vacuum leak. | Take corrective action; see appropriate Troubleshooting section in 240 GC/MS Ion Trap Hardware Operation Manual. | Startup and shutdown |
| Not Ready message at top of screen p.14 ↗ | GC or AutoSampler not ready. | Wait until GC and AutoSampler come to ready state; message changes to Ready. Check individual ready states under Windows pull-down menu. | Acquiring Data |
| Auto tune generates an error message and restores last values in the instrument p.28 ↗ | Electron lens tuning failed to find best voltage setting; balance outside range of 200 to 300 µA. | Check electron lens balance; algorithm searches optimal values by changing four variables one at a time. | Electron lens tuning |
| Too high a pressure of air or water in the system p.27 ↗ | Air leak or need to bake out the system. | Perform Air/Water Check; results in poor performance. Bake out system if needed. | Air/water check |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Turbo pump speed | 100 % p.20 ↗ | Typical operating range, External mode |
| Pump current | 200 to 300 mAmps p.20 ↗ | Typical operating range, External mode |
| Power | 9 to 13 Watts p.20 ↗ | Typical operating range, External mode |
| Ion gauge pressure | < 20 µTorr p.20 ↗ | Typical operating range, External mode |
| Roughing line pressure | < 50 mTorr p.20 ↗ | Typical operating range, External mode |
| Default RF storage level | 35 u p.8 ↗ | Hybrid mode |
| Helium buffer gas flow rate | 1 mL/min p.9 ↗ | Ion storage |
| Initial helium buffer gas flow rate | 3 to 4 mL/min p.21 ↗ | Starting the instrument |
| Damping gas flow rate | 0.5 mL/min p.20 ↗ | Start damping gas flow |
| CI gas restrictor flow | 50 mL/min p.13 ↗ | Methane CI installation |
| Methane regulator second stage pressure | 20 psi p.13 ↗ | Methane CI installation |
| CI gas adjustment ion gauge pressure | 70 to 100 µTorr p.13 ↗ | CI gas adjustment |
| Conversion dynode voltage, positive mode | -10,000 V p.11 ↗ | Positive modes |
| Conversion dynode voltage, negative mode | +10,000 V p.11 ↗ | Negative mode |
| Electron multiplier gain | ~10^5 p.11 ↗ | Signal amplification |
| Electron multiplier gain for Air/Water Check | 10^5 p.27 ↗ | Air/Water Check |
| Electron lens balance range | 200 to 300 µA p.28 ↗ | Electron lens tuning |
| Default manifold temperature | 50 °C p.22 ↗ | System temperatures |
| Transfer line temperature offset below max column temperature | 20 °C p.22 ↗ | Set system temperatures |
| Turbo pump speed at shutdown | 35 % of full speed p.24 ↗ | Shutting down the system |
| Temperature threshold before main power off | 80 °C p.24 ↗ | Shutting down the system |
| Minimum vent time | 5 minutes p.24 ↗ | Shutting down the system |
| Wait time after bakeout before AutoTune | 2 hours p.22 ↗ | System bakeout |
| Ion trap temperature stabilization before mass calibration | 2 hours p.29 ↗ | Mass calibration |
| Scan rate, Normal mode | 5000 u/sec p.37 ↗ | Normal scan mode |
| Scan rate, Fast mode | 10000 u/sec p.37 ↗ | Fast scan mode |
| Scan rate, Fastest mode | 10000 u/sec p.37 ↗ | Fastest scan mode, Full scan only |
| Maximum number of chromatographic time segments | 250 p.36 ↗ | Runs up to 650 minutes |
| Maximum run length for time segments | 650 minutes p.36 ↗ | Chromatographic time segments |
| Profile data collection points | 10 points per m/z p.36 ↗ | Profile data |
| Default Target TIC for CI | 5,000 counts p.39 ↗ | Positive or negative chemical ionization |
| Maximum Target TIC | 65,000 counts p.39 ↗ | Ionization control |
| Typical Target TIC for best results | 20,000 to 40,000 counts p.39 ↗ | Full scan acquisitions |
| Maximum fixed ion time | 65,000 µsec p.39 ↗ | Customize fixed ion time experiments |
| Default Ejection Amplitude | 15 V p.40 ↗ | High mass isolation of CI reagent ions |
| Max Reaction Time range | 1 to 2000 µsec p.40 ↗ | Hybrid parameters |
| Isolation Window range | 1.0 to 14.0 m/z p.42 ↗ | MS/MS parameters |
| Default Isolation Window | 3.0 m/z p.42 ↗ | MS/MS parameters |
| Low Edge Offset range | -0.5 to 0.5 m/z p.43 ↗ | MS/MS parameters |
| High Edge Offset range | -0.5 to 0.5 m/z p.43 ↗ | MS/MS parameters |
| Default High Mass Ejection amplitude | 35 volts p.43 ↗ | MS/MS parameters |
| Multiplier Offset adjustment range | up to 300 V p.38 ↗ | General parameters tab |
| Default Count Threshold | 1 counts p.38 ↗ | General parameters tab |
| Mass calibration ions (PFTBA) | 69, 131, 264, 414, 464, 614 m/z p.29 ↗ | Mass calibration |
| Ion lens tuning cal gas ions | 131 and 414 m/z p.29 ↗ | Ion lens tuning |
| Trap DC offset optimization ion | 414 m/z p.29 ↗ | Trap DC offset voltage |