Agilent 4200 FlexScan FTIR Handheld Spectrometer User's Guide

Contents

  1. Overview pp. 5-8
    Product overview, features and intended use of the spectrometer
  2. Procedure pp. 9-22
    Components, battery installation, software startup and charging procedures
  3. Quality pp. 16-20
    Diagnostic values and performance validation checks
  4. Procedure pp. 23-34
    Steps for collecting spectra and using ATR internal reflectance
  5. Procedure pp. 35-36
    Routine maintenance and care instructions
  6. Ordering pp. 37-38
    List of replaceable parts and ordering information
  7. Spec pp. 39-41
    Technical specifications and environmental operating conditions
  8. p. 19 table, 3 rows
  9. p. 20 table, 1 rows
  10. p. 18 table, 6 rows
  11. p. 37 table, 15 rows
  12. p. 13 table, 8 rows

Table 3. Minimum performance for the Signal-to-Noise test

Sample interfaceSpecification @ 2500 cm-1Specification @ 1000 cm-1
ATR25006000
External reflectance15002800
Grazing angle15002800

Table 4. Minimum performance for the Stability test

TimeSpecification @ 3000 cm-1Specification @ 1000 cm-1
30 min2%1%

Table 2. Diagnostic values

ValueOptimal (green)Marginal (yellow)Critical (red)Comment
Energy (centerburst)29,000–21,000>29,000 or <21,000> 31,000 or < 18,000Indicates the overall alignment and proper gain adjustment for the system.
Battery lifeAC or >30 min30–5 min<5 min
Source1.9 A>2.2 or <1.6>2.5 or <1.0Indicates a problem with the source control voltage or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal12,000–4,000>12,000 or <4,000>15,000 or < 3,000Can verify gross alignment errors even if the reflectance cap is not in place.
Detector temperature35–44<35 or >44<30 or >48Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
CPU temperature10–75>75>80Indicates the ambient temperature over the specified range.

5. Spare Parts

Part numberDescription
0024 - 001Agilent 4200 FlexScan FTIR system with external reflectance
0024 - 002Agilent 4200 FlexScan FTIR system with diffuse reflectance
0024 - 003Agilent 4200 FlexScan FTIR system with grazing angle
0024 - 004Agilent 4200 FlexScan FTIR system with diamond ATR
0023 - 300 -Diffuse gold reflectance reference cap for diffuse reflectance
0023-301Polystyrene frequency reference cap for diffuse reflectance
0023 - 304Diffuse gold reflectance reference cap for external reflectance
0023 - 305Polystyrene frequency reference cap for external reflectance
0023 - 307Mirrored stainless steel reflectance reference cap for external reflectance
0023 - 520ATR sample press
800 - 2002Handheld PC (PDA)
800 - 2003Handheld PC docking station (includes USB and power cables)
800 - 2004Replacement handheld PC battery
800 - 2005Replacement lithium-ion battery
430 - 0018USB cable

For sampling accessories, upgrades, service contracts, repair and refurbishment services, contact Agilent.

Table 1. Power switch LED status

StatusLED colorAction (% duty cycle)
System OFFNo LED illuminationN/A
System OFFRed100%
System startupRed / greenGreen 0.5 s on / red 0.5 s on
System ONGreen100%
Low batteryGreen50% (0.5 s on / 0.5 s off)
Critically low batteryRed50% (0.5 s on / 0.5 s off)
Dead batteryNo LED illuminationN/A
Firmware updateRed / greenGreen blinks rapidly twice then red blinks rapidly twice

Engineering Facts

14 facts · every claim linked to a page
Safety

The instrument is not intrinsically safe and must only be used in atmospheres tested for flammable materials.

Determines whether the analyzer can be deployed in hazardous or potentially explosive environments.

Source
“4200 FlexScan FTIR systems are NOT intrinsically safe. Use the system only in atmospheres that have been tested for flammable materials.”
Selection

The 45-degree specular probe is intended for films thicker than 1 micron, while the grazing angle probe is for films thinner than 1 micron.

Guides probe selection based on expected coating or film thickness.

Source
“The 45 degree probe is optimal for analysis of thin films greater than 1 micron in thickness.”
Selection

The grazing angle probe should be chosen when nanometer-level contamination must be detected.

Defines the probe choice for ultra-thin contamination analysis.

Source
“The grazing angle probe should be used when very low level contamination needs to be analyzed (nanometer thickness levels).”
Installation

After delivery, the shipping case must be allowed to reach room temperature for several hours before opening to avoid condensation.

Prevents condensation damage during initial setup and installation.

Source
“place the shipping case in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature.”
Installation

The instrument must be kept away from hot surfaces and electromagnetic interference sources.

Affects placement and mounting location to avoid measurement errors and damage.

Source
“The instrument must be kept away from hot surfaces and any sources of electromagnetic interference.”
Operation

The battery door must remain closed and locked while the system is in use.

Ensures safe and reliable battery operation during field measurements.

Source
“The door must remain closed and locked while the system is in use.”
Operation

The power button must be held for two seconds to turn the system on or off.

Prevents accidental power-off and clarifies startup procedure.

Source
“In order to turn the system on or off, hold the button down for two (2) seconds.”
Operation

The instrument must be connected via Bluetooth for MicroLab Mobile to launch.

Critical for establishing communication between the handheld PC and spectrometer.

Source
“The instrument must be connected via Bluetooth for the MicroLab Mobile software to launch.”
Performance

The system should be warmed up for at least 30 minutes before running performance validation tests.

Ensures valid performance validation results and stable measurements.

Source
“The 4200 FlexScan system should be warmed up for at least 30 minutes before conducting any of these tests.”
Operation

For reflectance measurements, the sample absorbance should be between 0.01 and 2 AU.

Helps verify proper sample contact and measurement quality in real time.

Source
“For 4200 FlexScan FTIR applications using reflectance, the value should be between 0.01 and 2 AU.”
Selection

The diamond ATR accepts samples with a pH range from 1 to 14.

Defines chemical compatibility limits for the ATR sampling interface.

Source
“Diamond can accept samples with a pH range from 1 to 14, which means hard or abrasive samples and even strong acids can be safely analyzed.”
Operation

The ATR diamond should not contact the sample or any substance other than air during background measurement.

Ensures an accurate background spectrum for ATR analysis.

Source
“The diamond should not be in contact with the sample or any substance other than the air in the atmosphere during the background measurement.”
Operation

The solid sample press delivers 15 pounds of force and is not for use with liquids, gels, or pastes.

Prevents misuse of the press and ensures proper contact for solid and powder samples.

Source
“The press is designed to deliver 15 pounds of force to the sample or over 2500 pounds per square inch of pressure. It is not for use with liquids, gels or pastes”
Maintenance

The outside of the unit may be cleaned with a damp, soft cloth and mild detergent, but must not be immersed in water.

Defines safe cleaning procedure and warns against water ingress into the electrical compartment.

Source
“Use a damp, soft cloth and a mild detergent to wipe down the outside of the unit. Do not completely immerse the 4200 FlexScan in water.”