Fast Hydrocarbon and Sulfur Simulated Distillation Using the
Technical Reference Summary
Covered Models & Document Scope
Application note describing a fast multielement simulated distillation (SimDis) method for hydrocarbon and sulfur boiling point distribution using the Agilent Low Thermal Mass (LTM) system on the 7890A GC with FID and the 355 Sulfur Chemiluminescence Detector (SCD) in series configuration; no product model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| LTM column temperature | 350 °C p.2 ↗ | |
| Column | DB-1 5 m × 320 mm × 1 µm p.2 ↗ | LTM column module |
| FID temperature | 350 °C p.2 ↗ | |
| H2 flow | 40 mL/min p.2 ↗ | |
| Data rate | 5 Hz p.2 ↗ | |
| Burner temperature | 800 °C p.2 ↗ | SCD |
| Air flow | 8.3 SCCM p.2 ↗ | |
| Temperature program rate | 200 °C/min p.2 ↗ | example high program rate |
Operating Limits
- Run time for hydrocarbon and sulfur boiling point distribution vs conventional ASTM D2887 procedure: at least six times shorter p.1 ↗