Application Note

Technical Reference Summary

Covered Models & Document Scope

Application note covering the handheld Agilent 4100 ExoScan FTIR spectrometer for measuring epoxy primer thickness on aluminum; also mentions the related 4200 FlexScan FTIR variant (same optical components, optics/electronics separated by cable).

Technical Data

Parameter Value Condition
Calibration band center 1610 cm-1 p.3 ↗ epoxy primer thickness calibration
Absorbance of thinnest sample approximately 0.07 absorbance units p.3 ↗ 1610 cm-1 band
Absorbance of thickest sample 0.7 absorbance units p.3 ↗ 1610 cm-1 band
Linear range of IR absorbance typically 0.05 to 1 absorbance units p.3 ↗
Alternative stronger band 1510 cm-1 p.3 ↗ usable for thicker calibration ranges
Alternative weaker band 1285 cm-1 p.3 ↗ usable for thinner calibration ranges
Calibration thicknesses 0.06, 0.09, 0.17, 0.30 and 0.39 mils p.3 ↗ epoxy primer calibration samples
Optical head weight 3 pound p.3 ↗ 4200 FlexScan
Battery and electronics pack weight 4 pound p.3 ↗ 4200 FlexScan

Operating Limits

  • Linear range of IR absorbance: typically 0.05 to 1 absorbance units p.3 ↗