Polyamide Analysis on Agilent PL HFIPgel with Gel Permeation Chromatography

Technical Reference Summary

Covered Models & Document Scope

Application Note covering analysis of polyamides by gel permeation chromatography (GPC) using Agilent PL HFIPgel columns (two-column set) with refractive index detection in HFIP + 0.02 M NaTFAc; calibration with Agilent PM-1 PMMA narrow standards.

Technical Data

Parameter Value Condition
Solvent HFIP p.2 ↗ GPC mobile phase
Salt additive concentration 0.02 M p.2 ↗ NaTFAc in HFIP
Column set two PL HFIPgel columns p.2 ↗ two-column set
Detector refractive index p.2 ↗
Sample concentration 0.2 % (w/v) p.2 ↗ polyamide sample
Calibration standard PM-1 PMMA narrow standards p.2 ↗
Highest molecular weight standard excluded 1,900,000 g/mol p.2 ↗ Mp, excluded on PL HFIPgel columns
Standard Mp values 790,000; 281,700; 144,000; 59,800; 28,900; 13,300; 5,720; 1,936; 1,020 g/mol p.2 ↗ PMMA standards
Analysis temperature 40 °C p.3 ↗

Operating Limits

  • Highest molecular weight standard usable on PL HFIPgel columns: 1,900,000 g/mol excluded p.2 ↗
  • Analysis temperature: 40 °C p.3 ↗