Polyamide Analysis on Agilent PL HFIPgel with Gel Permeation Chromatography
Technical Reference Summary
Covered Models & Document Scope
Application Note covering analysis of polyamides by gel permeation chromatography (GPC) using Agilent PL HFIPgel columns (two-column set) with refractive index detection in HFIP + 0.02 M NaTFAc; calibration with Agilent PM-1 PMMA narrow standards.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Solvent | HFIP p.2 ↗ | GPC mobile phase |
| Salt additive concentration | 0.02 M p.2 ↗ | NaTFAc in HFIP |
| Column set | two PL HFIPgel columns p.2 ↗ | two-column set |
| Detector | refractive index p.2 ↗ | |
| Sample concentration | 0.2 % (w/v) p.2 ↗ | polyamide sample |
| Calibration standard | PM-1 PMMA narrow standards p.2 ↗ | |
| Highest molecular weight standard excluded | 1,900,000 g/mol p.2 ↗ | Mp, excluded on PL HFIPgel columns |
| Standard Mp values | 790,000; 281,700; 144,000; 59,800; 28,900; 13,300; 5,720; 1,936; 1,020 g/mol p.2 ↗ | PMMA standards |
| Analysis temperature | 40 °C p.3 ↗ |
Operating Limits
- Highest molecular weight standard usable on PL HFIPgel columns: 1,900,000 g/mol excluded p.2 ↗
- Analysis temperature: 40 °C p.3 ↗