Automated Plasma Processing for Quantitative, Targeted LC/MS Analysis

Technical Reference Summary

Covered Models & Document Scope

Application note on automated plasma processing for quantitative targeted LC/MS protein analysis using the Agilent Bravo Automated Liquid Handling Platform; compares automated trypsin digestion in 96-well plates to manual digestion in microcentrifuge tubes.

Technical Data

Parameter Value Condition
Samples processed per day up to 96 samples/day p.1 ↗ Bravo Platform, automated
Shaking speed 800 rpm p.2 ↗ digestion
Digestion time 30 minutes p.2 ↗ 37 °C
Digestion temperature 37 °C p.2 ↗
TCEP concentration 25 mM p.2 ↗ reduction
IAA concentration 80 mM p.2 ↗ alkylation
Centrifugation speed 5,000 x g p.3 ↗ 5 minutes
Reconstitution solvent 3 % acetonitrile, 5 % acetic acid p.3 ↗ 3 % ACN, 5 % HOAc
Injection volume 1 μL p.3 ↗ HPLC-Chip
Median recovery (PAR) 50 % p.3 ↗ manual (tubes)
Median precision (CV) 5.9 % p.3 ↗ manual (tubes)

Operating Limits

  • maximum samples per day: 96 samples/day (automated) vs 24 samples/day (manual) p.1 ↗
  • digestion temperature: 37 °C p.2 ↗
  • centrifugation: 5,000 x g for 5 minutes p.3 ↗