Rapid and Accurate Forensics Analysis using High Resolution All

Technical Reference Summary

Covered Models & Document Scope

Application note describing the Agilent All Ions MS/MS approach for high resolution accurate mass screening of forensic samples using TOF or Q-TOF LC/MS instruments; covers HPLC conditions, ion source parameters, MS parameters (MS-mode and All Ions MS/MS mode), and sample preparation for a 30-compound test standard and spiked blood sample.

Technical Data

Parameter Value Condition
UHPLC column Acquity UPLC, HSS T3, 2.1 mm × 150 mm, 1.8 µm p.2 ↗
Mobile phase A 10 mM ammonium formate + 0.05% formic acid in water p.2 ↗
Mobile phase B 0.05% formic acid in acetonitrile p.2 ↗
Flow rate 0.5 mL/min p.2 ↗
Stop time 12.0 min p.2 ↗
Gas temperature 300 °C p.2 ↗
Gas flow 6 L/min p.2 ↗
Nebulizer 35 psi p.2 ↗
Sheath gas temperature 375 °C p.2 ↗
Sheath gas flow 10 L/min p.2 ↗
Scan range (MS-mode) 50–1,000 amu p.2 ↗ MS-mode
Scan rate (MS-mode) 1.5 spectra/s p.2 ↗ MS-mode
Scan segments (All Ions MS/MS mode) Three scan segments at 0 eV, 20 eV, and 40 eV p.2 ↗ All Ions MS/MS mode
Scan range (All Ions MS/MS mode) 50–1,000 amu p.2 ↗ All Ions MS/MS mode
Scan rate (All Ions MS/MS mode) 3 spectra/s p.2 ↗ All Ions MS/MS mode
Internal reference purine (m/z 121.0509) and HP 921 (m/z 922.0098) p.2 ↗ Automatic internal reference mass correction
Injection volume 2 µL p.2 ↗
Sample preparation liquid/liquid extraction with acetonitrile/ethanol 90/10 (250 mg blood + 500 µL extraction solvent) p.2 ↗