Induztr
Engineering Digest Document Viewer Related Models

Analysis of Sulfur Compounds in Petroleum Gases and Natural Gas

Technical Reference Summary

Covered Models & Document Scope

Application note on analyzing sulfur compounds in petroleum gases and natural gas using an Agilent J&W DB-Sulfur SCD GC column (70 m × 0.53 mm, 4.3 µm, p/n G3903-63003) with Agilent 7890A GC and 355 SCD; two GC condition sets (Conditions 1 and 2) are described.

Technical Data

Parameter Value Condition
Column length 70 m p.2 ↗ DB-Sulfur SCD, p/n G3903-63003
Column inner diameter 0.53 mm p.2 ↗ DB-Sulfur SCD, p/n G3903-63003
Column film thickness 4.3 µm p.2 ↗ DB-Sulfur SCD, p/n G3903-63003
Carrier gas pressure 8.6 psi p.2 ↗ Helium, constant pressure mode
Inlet temperature 240 °C p.2 ↗ Conditions 1, inert flow path split/splitless inlet
Split ratio 10:1 p.2 ↗ Conditions 1
Oven initial temperature 35 °C p.2 ↗ held for 2 min
Oven ramp 1 35 to 80 at 10 °C/min p.2 ↗
Oven hold at 80 °C 6 min p.2 ↗
Oven ramp 2 80 to 220 at 15 °C/min p.2 ↗
Oven final hold 220 for 10 °C, min p.2 ↗
Sample loop volume 1 mL p.2 ↗
Burner temperature 800 °C p.2 ↗ SCD conditions
Burner vacuum 350 Torr p.2 ↗ SCD conditions
Reaction cell vacuum 5 Torr p.2 ↗ SCD conditions
Hydrogen flow 40 mL/min p.2 ↗ SCD conditions
Air flow 60 mL/min p.2 ↗ SCD conditions

Operating Limits

  • Maximum oven temperature: 220 °C p.2 ↗
  • Burner vacuum: 350 Torr p.2 ↗
  • Reaction cell vacuum: 5 Torr p.2 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Agilent J&W DB-Sulfur SCD, 70 m × 0.53 mm, 4.3 µm p.2 ↗ G3903-63003 GC column
Deactivated fused silica tubing, 1 m × 0.18 mm p.2 ↗ 160-2615-1 Tubing
Deactivated Ultimate union p.2 ↗ G3182-60580 Union
Ultra Inert liner p.2 ↗ 5190-2295 Liner