Automated Online SPE for LC/MS/MS Analysis of Trace

Technical Reference Summary

Covered Models & Document Scope

Application note describing an automated online SPE LC/MS/MS method for trace organic contaminants (TOrCs) in water using the Agilent 1290 Infinity Flexible Cube module with two online SPE cartridges and an Agilent 6460 Triple Quadrupole LC/MS; covers 34 indicator compounds, sample volume <2 mL, cycle time <15 min.

Technical Data

Parameter Value Condition
Analytical column Agilent Poroshell 120 EC C18 p.3 ↗
Column temperature 30 °C p.3 ↗
Injection volume 2 × 850 µL, 1.7 mL total µL p.3 ↗
Injection draw speed 500 µL/min p.3 ↗
Draw position 0.0 mm p.3 ↗
Flow rate 1.0 mL/min p.3 ↗
Mobile phase A Water (95 %) + acetonitrile (5 %) + 0.1 % acetic acid p.3 ↗
Solvent delay 0.7 minutes p.3 ↗
Sheath gas flow rate 12 L/min p.3 ↗
Drying gas flow rate 11 L/min p.3 ↗
Nebulizer pressure 45 psig p.3 ↗
Nozzle voltage 4,000 positive; 3,500 negative V p.3 ↗
ESI mode simultaneous with fast polarity switching p.3 ↗
Sample volume less than 2 mL p.2 ↗
Cycle time less than 15 minutes p.2 ↗
Number of indicator compounds 34 p.2 ↗

Operating Limits

  • Sample volume: less than 2 mL p.2 ↗
  • Cycle time: less than 15 minutes p.2 ↗
  • Detection limits: low ng/L p.2 ↗