Sensitive Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

Technical Reference Summary

Covered Models & Document Scope

Application Note describing an online SPE UHPLC method for determination of 16 PAHs in drinking water on an Agilent 1200 Infinity Series Online SPE system (Agilent 1260 Infinity LC with autosampler); no product model/size variants or ordering codes are covered.

Technical Data

Parameter Value Condition
Sample volume required less than 2 mL p.2 ↗
Overall run time (enrichment + analytical run) 30 minutes p.2 ↗
Number of PAHs monitored 16 p.2 ↗
Autosampler sample capacity 2 × 18, 6-mL vials p.2 ↗ Agilent 1260 Infinity autosampler
Fluorescence detector acquisition Multisignal acquisition, set at λex= 260 nm p.3 ↗ Agilent 1290 Infinity fluorescence detector
Retention time RSD 0.011–0.090 % p.4 ↗ n = 7, 16 PAHs
Area RSD 0.3–6.0 % p.4 ↗ n = 7, 16 PAHs
Linearity R² 0.99834–0.99996 p.5 ↗ 16 PAHs
Limit of detection (LOD, S/N = 3) 0.0003–0.0233 µg/L p.5 ↗ 16 PAHs
Limit of quantification (LOQ, S/N = 10) 0.0009–0.0775 µg/L p.5 ↗ 16 PAHs
LOD range 0.2–23 ng/L p.5 ↗

Operating Limits

  • Minimum sample volume sufficient for analysis: less than 2 mL p.2 ↗
  • Overall run time: 30 minutes p.2 ↗