Analysis of Arsine and Phosphine in Ethylene and Propylene Using the

Technical Reference Summary

Covered Models & Document Scope

Agilent Arsine Phosphine GC/MS Analyzer, model G3440B with option SP1 7890-0667, comprising Agilent 7890B GC and Agilent 5977B Series GC/MSD System HES with hydrogen cleaning; application for arsine and phosphine (plus H2S, COS) in ethylene and propylene.

Technical Data

Parameter Value Condition
Analyzer model G3440B p.2 ↗ with option SP1 7890-0667
Gas chromatograph 7890B GC p.2 ↗
Mass spectrometer detector 5977B Series GC/MSD System HES p.2 ↗ with hydrogen cleaning
Column 120 m × 0.32 mm, 8.0 µm Select Olefins p.2 ↗
Carrier gas Helium, constant flow at 1.2 mL/min p.2 ↗
Oven program 35 °C p.2 ↗ for 22 minutes
Hydrogen flow into source 150 µL/min p.2 ↗ H2 at 40 psi
Linearity (R²), Phosphine (PH3) 0.9996 p.3 ↗
Linearity (R²), Hydrogen sulfide (H2S) 0.9995 p.3 ↗
Linearity (R²), Arsine (AsH3) 0.9999 p.3 ↗
Linearity (R²), Carbonyl sulfide (COS) 0.9987 p.3 ↗
RSD, Phosphine (PH3) 5.14 % p.3 ↗
RSD, Hydrogen sulfide (H2S) 3.96 % p.3 ↗
RSD, Arsine (AsH3) 0.62 % p.3 ↗
RSD, Carbonyl sulfide (COS) 5.06 % p.3 ↗
IDL, Phosphine (PH3) 0.715 ppb p.3 ↗
IDL, Hydrogen sulfide (H2S) 0.456 ppb p.3 ↗
IDL, Arsine (AsH3) 0.063 ppb p.3 ↗
IDL, Carbonyl sulfide (COS) 0.575 ppb p.3 ↗

Operating Limits

  • Detection level (IDL) for arsine: 0.063 ppb p.3 ↗
  • Detection level (IDL) for phosphine: 0.715 ppb p.3 ↗
  • Detection level (IDL) for hydrogen sulfide: 0.456 ppb p.3 ↗
  • Detection level (IDL) for carbonyl sulfide: 0.575 ppb p.3 ↗