An Integrated Workflow for Automated Calculation of
Technical Reference Summary
Covered Models & Document Scope
Application note describing an integrated automated workflow for ADC DAR determination using the Agilent AssayMAP Bravo, Agilent MassHunter Walkup, and Agilent MassHunter BioConfirm DAR Calculator software with an Accurate-Mass TOF LC/MS system; covers intact and reduced ADC, glycosylated and deglycosylated sample conditions.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| LC/MS system | Accurate-Mass TOF LC/MS p.3 ↗ | |
| Ion source | Dual Agilent Jet Stream p.3 ↗ | Intact ADC |
| Ion source | Dual Agilent Jet Stream p.3 ↗ | Reduced ADC |
| Gas temperature | 325 °C p.3 ↗ | Intact ADC |
| Gas flow | 13 L/min p.3 ↗ | Intact ADC |
| Gas flow | 13 L/min p.3 ↗ | Reduced ADC |
| Nebulizer | 60 psig p.3 ↗ | Intact ADC |
| Sheath gas temperature | 350 °C p.3 ↗ | Intact ADC |
| Sheath gas temperature | 350 °C p.3 ↗ | Reduced ADC |
| Sheath gas flow | 12 L/min p.3 ↗ | Intact ADC |
| Sheath gas flow | 12 L/min p.3 ↗ | Reduced ADC |
| VCap | 4,000 V p.3 ↗ | Intact ADC |
| VCap | 4,000 V p.3 ↗ | Reduced ADC |
| Nozzle voltage | 2,000 V p.3 ↗ | Intact ADC |
| Nozzle voltage | 2,000 V p.3 ↗ | Reduced ADC |
| Fragmentor | 400 V p.3 ↗ | Intact ADC |
| Skimmer | 65 V p.3 ↗ | Intact ADC |
| Mass range | 800–8,000 m/z p.3 ↗ | Intact ADC |
| Scan rate | 1.0 spectra/sec p.3 ↗ | Intact ADC |
| Scan rate | 1.0 spectra/sec p.3 ↗ | Reduced ADC |
Operating Limits
- Autosampler thermostat temperature: 5 °C p.3 ↗
- Sample incubation temperature: 50 °C for 10 minutes p.2 ↗
- ADC storage temperature: –80 °C p.2 ↗