Precise Characterization of Intact Monoclonal Antibodies by the
Technical Reference Summary
Covered Models & Document Scope
Application note describing intact monoclonal antibody characterization by LC/MS using the Agilent 6545XT AdvanceBio LC/Q-TOF with an Agilent 1290 Infinity II UHPLC and PLRP-S column; covers NIST mAb RM 8671, Herceptin (Trastuzumab) and a T-DM1 ADC. No product model variants or ordering codes are given.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column | Agilent PLRP-S, 1000 Å, 2.1 × 50 mm, 5 µm (p/n PL1912-1502) p.2 ↗ | |
| Autosampler thermostat temperature | 4 °C p.2 ↗ | |
| Solvent A | 0.1 % Formic acid in DI water p.2 ↗ | |
| Solvent B | 0.1 % Formic acid in 100 % acetonitrile p.2 ↗ | |
| Column temperature | 60 °C p.2 ↗ | |
| Flow rate | 0.5 mL/min p.2 ↗ | |
| Injection volume | 0.5 µL p.2 ↗ | |
| Ion source | Dual Agilent Jet Stream p.2 ↗ | |
| Gas temperature | 350 °C p.2 ↗ | |
| Gas flow | 12 L/min p.2 ↗ | |
| Nebulizer | 60 psig p.2 ↗ | |
| Sheath gas temperature | 400 °C p.2 ↗ | |
| Sheath gas flow | 11 L/min p.2 ↗ | |
| Capillary voltage | 5,500 V p.2 ↗ | |
| Nozzle voltage | 2,000 V p.2 ↗ | |
| Fragmentor | 380 V p.2 ↗ | |
| Skimmer | 140 V p.2 ↗ | |
| Quad AMU | 500 m/z p.2 ↗ | |
| Mass range | 100–10,000 m/z p.2 ↗ | |
| Acquisition rate | 1.0 spectra/s p.2 ↗ | |
| Reference mass | 922.0098 p.2 ↗ | |
| Acquisition mode | Positive, Extended (10,000 m/z) Mass Range p.2 ↗ | |
| Deconvoluted mass range | 140,000 to 160,000 Daltons p.2 ↗ | |
| Sample concentration | 1.0 µg/µL p.2 ↗ | diluted with DI water |
Operating Limits
- Mass deconvolution averaging range: 2,000 to 5,000 m/z p.2 ↗
- Deconvoluted mass range: 140,000 to 160,000 Daltons p.2 ↗