Application note describing intact monoclonal antibody characterization by LC/MS using the Agilent 6545XT AdvanceBio LC/Q-TOF with an Agilent 1290 Infinity II UHPLC and PLRP-S column; covers NIST mAb RM 8671, Herceptin (Trastuzumab) and a T-DM1 ADC. No product model variants or ordering codes are given.

Technical Data

Parameter Value Condition
Column Agilent PLRP-S, 1000 Å, 2.1 × 50 mm, 5 µm (p/n PL1912-1502) p.2 ↗
Autosampler thermostat temperature 4 °C p.2 ↗
Solvent A 0.1 % Formic acid in DI water p.2 ↗
Solvent B 0.1 % Formic acid in 100 % acetonitrile p.2 ↗
Column temperature 60 °C p.2 ↗
Flow rate 0.5 mL/min p.2 ↗
Injection volume 0.5 µL p.2 ↗
Ion source Dual Agilent Jet Stream p.2 ↗
Gas temperature 350 °C p.2 ↗
Gas flow 12 L/min p.2 ↗
Nebulizer 60 psig p.2 ↗
Sheath gas temperature 400 °C p.2 ↗
Sheath gas flow 11 L/min p.2 ↗
Capillary voltage 5,500 V p.2 ↗
Nozzle voltage 2,000 V p.2 ↗
Fragmentor 380 V p.2 ↗
Skimmer 140 V p.2 ↗
Quad AMU 500 m/z p.2 ↗
Mass range 100–10,000 m/z p.2 ↗
Acquisition rate 1.0 spectra/s p.2 ↗
Reference mass 922.0098 p.2 ↗
Acquisition mode Positive, Extended (10,000 m/z) Mass Range p.2 ↗
Deconvoluted mass range 140,000 to 160,000 Daltons p.2 ↗
Sample concentration 1.0 µg/µL p.2 ↗ diluted with DI water

Operating Limits

  • Mass deconvolution averaging range: 2,000 to 5,000 m/z p.2 ↗
  • Deconvoluted mass range: 140,000 to 160,000 Daltons p.2 ↗