Trace Analysis of Ammonia in Ethylene by Gas Chromatography

Technical Reference Summary

Covered Models & Document Scope

Application note describing a GC/NCD configuration for trace ammonia in ethylene: Agilent 7890B GC with Agilent 8255 NCD, 6-port gas sample valve with 1-mL sample loop, megabore column p/n CP8590 (Select Low Ammonia), UltiMetal Plus deactivated stainless steel surfaces, OpenLAB CDS.

Technical Data

Parameter Value Condition
Valve/valve box temperature 125 °C p.2 ↗
Column oven temperature program 100 (hold 0.25 minutes), ramp 25 °C/min to 150 (hold 2.75 minutes) °C p.2 ↗
NCD base temperature 200 °C p.2 ↗
NCD burner temperature 900 °C p.2 ↗
Oxidizer flow during analysis 30 mL/min p.2 ↗ maximum flow, during matrix hydrocarbon elution
Oxidizer flow at ammonia elution 4 mL/min p.2 ↗ just before ammonia eluted to detector
Sample loop volume 1 mL p.2 ↗ gas sample valve
Valve switch-off time after injection 0.25 minutes p.2 ↗
Ammonia run area 264.08 p.3 ↗ run 6
Ammonia standard deviation 7.20 p.3 ↗ run 6
Ammonia relative standard deviation (RSD) 2.73 % p.3 ↗

Operating Limits

  • Detection capability: low ppbv levels p.2 ↗
  • Oxidizer flow maximum: 30 mL/min p.2 ↗
  • Oxidizer flow minimum at ammonia elution: 4 mL/min p.2 ↗