Induztr
Engineering Digest Document Viewer Related Models

Enabling Automated Serial Dilutions Using the Agilent Bravo BenchCel Workstation

Technical Reference Summary

Covered Models & Document Scope

Application Note describing an automated serial dilution workflow on the Agilent Bravo BenchCel Workstation, consisting of an Agilent BenchCel 4R Microplate Handler with 660-mm Front Loading Racks and an Agilent Bravo Automated Liquid Handling Platform with Bravo Pipetting Head (300 nL to 200 µL), driven by Agilent VWorks Automation Control Software.

Technical Data

Parameter Value Condition
Pipetting volume range 300 nL to 200 µL p.1 ↗ Agilent Bravo Pipetting Head
Number of standard deck positions nine p.1 ↗ Agilent Bravo
Microplate handler front loading rack size 660 mm p.2 ↗ Agilent BenchCel 4R
Tartrazine standard concentration 0.36 mM p.2 ↗ 0.096 g Tartrazine in 500 mL DI water
Mix volume per mix cycle 150 µL p.3 ↗ Serial Dilution Plate protocol, three mix cycles
Mix amplitude 6 mm p.3 ↗ Serial Dilution Plate protocol, three mix cycles
Number of mix cycles three p.3 ↗ Serial Dilution Plate protocol
Average precision 2.7 % p.4 ↗ Plate 1, column/row serial dilution
Processing time 568 sec/plate p.4 ↗ Plate 1, column/row serial dilution
Accuracy ratio 1:1.95 p.4 ↗ Column/row serial dilution