Enabling Automated Serial Dilutions Using the Agilent Bravo BenchCel Workstation
Technical Reference Summary
Covered Models & Document Scope
Application Note describing an automated serial dilution workflow on the Agilent Bravo BenchCel Workstation, consisting of an Agilent BenchCel 4R Microplate Handler with 660-mm Front Loading Racks and an Agilent Bravo Automated Liquid Handling Platform with Bravo Pipetting Head (300 nL to 200 µL), driven by Agilent VWorks Automation Control Software.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Pipetting volume range | 300 nL to 200 µL p.1 ↗ | Agilent Bravo Pipetting Head |
| Number of standard deck positions | nine p.1 ↗ | Agilent Bravo |
| Microplate handler front loading rack size | 660 mm p.2 ↗ | Agilent BenchCel 4R |
| Tartrazine standard concentration | 0.36 mM p.2 ↗ | 0.096 g Tartrazine in 500 mL DI water |
| Mix volume per mix cycle | 150 µL p.3 ↗ | Serial Dilution Plate protocol, three mix cycles |
| Mix amplitude | 6 mm p.3 ↗ | Serial Dilution Plate protocol, three mix cycles |
| Number of mix cycles | three p.3 ↗ | Serial Dilution Plate protocol |
| Average precision | 2.7 % p.4 ↗ | Plate 1, column/row serial dilution |
| Processing time | 568 sec/plate p.4 ↗ | Plate 1, column/row serial dilution |
| Accuracy ratio | 1:1.95 p.4 ↗ | Column/row serial dilution |