Cannabinoid Profiling and Quantitation in Hemp Extracts using

Technical Reference Summary

Covered Models & Document Scope

Application brief covering cannabinoid profiling and quantitation in hemp extracts using the Agilent 1290 Infinity II UHPLC with Agilent 6230B LC-TOF MS; comparison with 1290 Infinity II UHPLC with DAD. No instrument model variants or ordering codes are given.

Technical Data

Parameter Value Condition
Column Agilent ZORBAX Bonus RP. 2.1 × 50 mm, 1. p.2 ↗
Stop time 12.5 minutes p.2 ↗
Post time 1.0 minutes p.2 ↗
Overall run time 20.0 minutes p.2 ↗ including re-equilibration
Nebulizer gas pressure 40 psi p.2 ↗
Capillary voltage 4,000 V p.2 ↗
Fragmentor 175 V p.2 ↗
Skimmer 65 V p.2 ↗
Ref mass enabled Yes p.2 ↗
Detection limit 1.0 ng/mL p.1 ↗ as low as
Linear regression coefficient 0.999 p.1 ↗ or better
Calibration range 10 to 1,000 µg/mL p.1 ↗
THC concentration in hemp products <0.3 % p.1 ↗

Operating Limits

  • Detection limit (lowest): 1.0 ng/mL p.1 ↗
  • Calibration range: 10 µg/mL to 1,000 µg/mL p.1 ↗
  • THC content in hemp products: <0.3 % p.1 ↗