Quantification of Crystallinity Using Transmission Raman Spectroscopy

Technical Reference Summary

Covered Models & Document Scope

Application note covering the Agilent TRS100 Raman system for bulk quantification of crystalline API content (0–9.4 % w/w) in spray dried solid dispersions; comparison with pXRD and ssNMR.

Technical Data

Parameter Value Condition
Limit of detection (LOD) 0.9 % w/w p.2 ↗ TRS100, crystalline API in amorphous solid dispersion
Crystalline content range analyzed 0–9.4 % w/w p.2 ↗ nine powder mixtures
Correlation fit (R2) 0.99 p.2 ↗ PLS chemometric model
Data acquisition time 1–5 minutes p.2 ↗ TRS
Sensitivity (LOD) 0.2–1 % p.2 ↗ TRS
Sensitivity (LOD) 2–5 % p.2 ↗ pXRD
Sensitivity (LOD) 0.3–1 % p.2 ↗ ssNMR
Data acquisition time ~1 hour p.2 ↗ pXRD
Data acquisition time 6–24 hours p.2 ↗ ssNMR

Operating Limits

  • Limit of detection (LOD): 0.9 % w/w p.2 ↗
  • Crystalline content analyzed (max): 9.4 % w/w p.2 ↗