Poster #47D (MSACL 2018 US) describing a fast LC/MS/MS analytical method with Alternate Column Regeneration (ACR) for the analysis of 125 drugs and metabolites in urine; developed on an Agilent 1290 Infinity II UHPLC with 6470 TQ Mass Spectrometer; two-column ACR configuration (Column 1 / Column 2, Agilent Poroshell 120 EC-C18, 2.1 x 10 mm); document number 5991-9141EN.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analysis time (standard configuration) | 7.35 min p.1 ↗ | 5 min gradient + 1.5 min equilibration + 0.85 min injection |
| Analysis time (ACR configuration) | 5.1 min p.1 ↗ | injection to injection |
| Analysis time reduction with ACR | 30 % p.1 ↗ | |
| Number of analytes | 125 p.1 ↗ | |
| Eluent pump flow rate | 0.35 mL/min p.1 ↗ | |
| Regeneration pump flow rate | 0.5 mL/min p.1 ↗ | |
| Injection volume | 10 µL p.1 ↗ | |
| Autosampler temperature | 5 °C p.1 ↗ | |
| Column temperature | 55 °C p.1 ↗ | |
| Column | Agilent Poroshell 120 EC-C18, 2.1 x 10 mm p.1 ↗ | Columns 2 |
| Mobile phase A | H2O + 5 mM Ammonium Formate + 0.01% Formic Acid p.1 ↗ | |
| Mobile phase B | Methanol + 0.01% Formic Acid p.1 ↗ | |
| Needle wash | 50:20:20:10 IPA:MeOH:ACN:H2O p.1 ↗ | |
| Gas temperature | 300 °C p.1 ↗ | positive mode |
| Gas temperature | 300 °C p.1 ↗ | negative mode |
| Gas flow | 9 L/min p.1 ↗ | positive mode |
| Gas flow | 9 L/min p.1 ↗ | negative mode |
| Nebulizer pressure | 30 psi p.1 ↗ | positive mode |
| Nebulizer pressure | 30 psi p.1 ↗ | negative mode |
| Sheath gas temperature | 380 °C p.1 ↗ | positive mode |
| Sheath gas temperature | 380 °C p.1 ↗ | negative mode |
| Sheath gas flow | 11 L/min p.1 ↗ | positive mode |
| Sheath gas flow | 11 L/min p.1 ↗ | negative mode |
| Capillary voltage | 3750 V p.1 ↗ | positive mode |
Operating Limits
- Calibration range: 1 to 1000 ng/mL p.1 ↗
- Lowest analyzed concentration: 1 ng/mL (1 pg on column) p.1 ↗
- Signal-to-noise at 1 ng/mL: better than 10 for 104 compounds p.1 ↗
- Linearity range: 1 to 1000 ng/mL for 112 curves p.1 ↗
- Reproducibility (CV): < 15% for majority of analytes p.1 ↗