PowerPoint Presentation
Technical Reference Summary
Covered Models & Document Scope
This document is a conference poster (MSACL 2018, Poster #30A) comparing immunoassay and rapid LC-TOF-MS for presumptive drug screening in clinical research; it is not a product datasheet and covers no models, sizes, or ordering codes.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column dimensions | 2.1x50 mm p.1 ↗ | Agilent Poroshell SB-C8, 2.7 µm |
| Column particle size | 2.7 µm p.1 ↗ | Agilent Poroshell SB-C8 |
| Column temperature | 75 °C p.1 ↗ | |
| Injection volume | 2 µL p.1 ↗ | Positive mode |
| Flow rate | 1 mL/minute p.1 ↗ | Positive and negative mode |
| Mobile phase A (positive mode) | 5 mM ammonium formate, pH 3.5 p.1 ↗ | Positive mode |
| Mobile phase B (positive mode) | 0.1% formic acid in ACN p.1 ↗ | Positive mode |
| Mobile phase A (negative mode) | 0.1% acetic acid in water p.1 ↗ | Negative mode |
| Gradient | 5%-95% B in 1.25 minutes, 0.5 minute re-equilibration p.1 ↗ | Positive mode |
| Gas temperature | 250 °C p.1 ↗ | Positive and negative mode |
| Drying gas flow | 15 L/min p.1 ↗ | Positive and negative mode |
| Nebulizer pressure | 60 psig p.1 ↗ | Positive and negative mode |
| Sheath gas temperature | 400 °C p.1 ↗ | Positive and negative mode |
| Sheath gas flow | 12 L/min p.1 ↗ | Positive and negative mode |
| Vcap | 3500 V p.1 ↗ | Positive and negative mode |
| Fragmentor | 125 V p.1 ↗ | Positive and negative mode |
| Scan speed | 4 spectra/sec p.1 ↗ | Positive and negative mode |
Operating Limits
- Creatinine inter- and intra-assay imprecision: <3% at two QC levels p.1 ↗
- Carryover: 0.02% p.1 ↗
- Creatinine comparison to Jaffe method slope: 0.91 p.1 ↗
- Creatinine comparison to Jaffe method correlation coefficient: 0.96 p.1 ↗