US EPA Method 624 with the Tekmar Lumin and the Agilent 7890B

Technical Reference Summary

Covered Models & Document Scope

Application note for the Teledyne Tekmar Lumin purge and trap concentrator with AQUATek 100 autosampler, used with an Agilent 7890B GC/MS for US EPA Method 624 VOC analysis; covers Lumin/AQUATek 100 operating conditions and GC/MS conditions, not a product ordering datasheet.

Technical Data

Parameter Value Condition
Valve oven temperature 150 °C p.2 ↗ Lumin, standby
Transfer line temperature 160 °C p.2 ↗ Lumin, standby
Sample mount temperature 90 °C p.2 ↗ Lumin, standby
Standby flow 20 mL/min p.2 ↗ Lumin, standby
Purge ready temperature 35 °C p.2 ↗ Lumin, standby
MCS purge temperature 20 °C p.2 ↗ Lumin, purge
Purge temperature 20 °C p.2 ↗ Lumin, purge
Purge time 11.00 min p.2 ↗ Lumin, purge
Purge flow 40 mL/min p.2 ↗ Lumin, purge
Dry purge temperature 20 °C p.2 ↗ Lumin, purge
Dry purge time No Dry Purge, 0 min p.2 ↗ Lumin, purge
Desorb preheat temperature 245 °C p.2 ↗ Lumin, desorb
Desorb temperature 250 °C p.2 ↗ Lumin, desorb
Desorb time 1.00 min p.2 ↗ Lumin, desorb
Drain flow 300 mL/min p.2 ↗ Lumin, desorb
Bake time 2.00 min p.2 ↗ Lumin, bake
Bake temperature 280 °C p.2 ↗ Lumin, bake
MCS bake temperature 180 °C p.2 ↗ Lumin, bake
Bake flow 200 mL/min p.2 ↗ Lumin, bake
Sample loop time 0.35 min p.2 ↗ AQUATek 100
Sample transfer time 0.35 min p.2 ↗ AQUATek 100
Rinse loop time 0.30 min p.2 ↗ AQUATek 100
Sweep needle time 0.30 min p.2 ↗ AQUATek 100
Presweep time 0.25 min p.2 ↗ AQUATek 100

Operating Limits

  • Calibration range: 0.5 to 200 ppb p.2 ↗
  • Trap used: #9 p.2 ↗
  • Water temperature: 90 °C p.2 ↗
  • Bake rinse drain cycles: 2 p.2 ↗
  • Bake rinse drain time: 0.35 min p.2 ↗