Induztr
Engineering Digest Document Viewer Related Models

LC Troubleshooting Guide

Technical Reference Summary

Covered Models & Document Scope

Agilent LC troubleshooting guide (publication 5994-0709EN, March 1, 2019) covering best practices and troubleshooting for Agilent liquid chromatography systems; no instrument models, sizes, or ordering codes are specified.

Operating Limits

  • Recommended maximum time acetonitrile (ACN) may remain on the system: 2 to 3 days p.1 ↗
  • Recommended warm water wash temperature for ACN-related problems: 60 to 70 °C p.1 ↗
  • Recommended water content in mobile phase when using acetonitrile: 5 to 10% p.1 ↗
  • Recommended solvent filter pore size: 0.2 μm p.1 ↗
  • Recommended maximum time before using prepared solvent volume: 1 to 2 days p.1 ↗