Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS Quick Start Guide

Technical Reference Summary

Covered Models & Document Scope

Covers Agilent 6200 Series TOF LC/MS (6224, 6230) and 6500 Series Q-TOF LC/MS (6520B, 6530A/B, 6538A, 6540A/B, 6545, 6545XT, 6546, 6550, 6560 Ion Mobility) systems with MassHunter Workstation Software versions B.09.00 and 10.0; compatible LC modules 1200, 1260, 1260 Infinity II, 1290 Infinity, 1290 Infinity II and ion sources ESI, Dual ESI, AJS ESI, Dual AJS ESI, APCI, APPI, Multimode, nanospray, dual nanospray, MALDI, GC-APCI.

Section-by-Section Technical Breakdown

pp. 23
Supported sources for all Autotunes

Lists which ion sources are supported for all automated tune procedures on each TOF and Q-TOF model. NanoESI, Dual nanoESI, MALDI and GC-QTOF sources cannot be used for any automated tune. The Dual ESI source is supported for all Autotunes on all instruments.

  • 6224 TOF: ESI, Dual ESI
  • 6230 TOF: ESI, Dual ESI, AJS ESI, Dual AJS ESI
  • 6520B Q-TOF: ESI, Dual ESI
  • 6530A/B Q-TOF: ESI, Dual ESI, AJS ESI, Dual AJS ESI
  • 6538 Q-TOF: ESI, Dual ESI
  • 6540A/B Q-TOF: ESI, Dual ESI, AJS ESI, Dual AJS ESI
  • 6545, 6545XT, 6546 Q-TOF: ESI, Dual ESI, AJS ESI, Dual AJS ESI
  • 6550: ESI, Dual ESI, AJS ESI, Dual AJS ESI
  • 6560: Dual ESI, Dual AJS ESI
  • 6545XT ships with dielectric capillary; resistive capillary required for Fast Polarity Switching
pp. 21-28
SWARM Autotune

Describes the SWARM autotune algorithm used on 6224/6230 TOF and 6530, 6545, 6545XT, 6546, 6550, 6560 Q-TOF instruments. It combines Particle Swarm Optimization and a simplex algorithm, interchanging them between coarse and fine adjustment, with parameters bounded by ion physics. System Tune is the SWARM Initial Auto Tune including detector tune and requires Standard (3200 m/z) mass range.

  • System Tune requires Standard (3200 m/z) Mass Range
  • Transmission Tune does not include detector tune; run as frequently as needed
  • Low (1700 m/z) Mass Range disables System Tune
  • High Resolution Slicer Mode must be selected before System Tune
  • Set Detector Gain and System Tune only with mass range 3200 m/z
  • Fragile Ions checkbox reduces Fragmentor Voltage to minimize fragmentation
pp. 29-34
Classic Autotune

Covers the classic autotune workflow for TOF and Q-TOF instruments other than the SWARM models. Initial Tune, Standard Tune and Set Detector Gain require Extended Dynamic Range instrument mode and Standard (3200 m/z) mass range. After changing mass range or instrument mode, recalibration is required.

  • Instrument Mode must be Extended Dynamic Range for Set Detector Gain, Standard Tune, Initial Tune
  • Mass Range must be Standard (3200 m/z) for Set Detector Gain and Initial Auto Tune
  • High Resolution Slicer Mode required before Initial Auto Tune
  • For 6530 with Dual AJS ESI, 6538 or 6540, do not click Both; tune TOF then Quadrupole with tuning mix change
  • Software B.05.01 SP1 or later: additional Quad Tune in 1700 m/z range after Initial Tune improves molecules < 300 m/z
pp. 24, 33
Diluting the ESI-L Tune Mix

Specifies tune mix dilution ratios per instrument model, source and polarity. Autotune can fail on 6530A, 6540A, 6545A, 6545XT, 6546, 6550A and 6560A with AJS ESI or Dual AJS ESI unless the tuning mix is diluted. Negative ion polarity generally requires dilution for proper calibration.

  • 6545, 6545XT, 6550A, 6560A with Dual AJS ESI: 1:10 dilution both polarities
  • 6530 with Dual AJS ESI: 1:10 positive, 1:40 negative
  • 6538A: undiluted tune mix
  • 6230 TOF with AJS/Dual AJS ESI: 1:10 positive, 1:40 negative
  • 6224 TOF with Dual ESI: do not dilute for positive ion
  • 6550 Q-TOF always uses 1:10 dilution
pp. 28-29, 33
Calibrate the mass-axis

Explains mass-axis calibration: a sample with known masses is infused and flight times measured to calculate updated calibration coefficients. Recalibration and tune check are required after changing Instrument Mode, Mass Range or Slicer Mode. Mass Calibration is triggered by Mass Calibration/Check, Quick Tune, Standard Tune or Initial Auto Tune (Classic) or by Mass Calibration/Check, Transmission Tune or System Tune (SWARM).

  • Recalibrate after changing Instrument Mode, Mass Range or Slicer Mode
  • 6545 with 50-250 m/z option uses calibrant ions below 100 m/z
  • Repeat Mass Calibration/Check if calibrant abundances exceed 480,000 (6538/6540) or ~650,000 (other TOF/Q-TOF)
  • Agilent recommends regular mass-axis calibration
pp. 17-19
Step 2. Prepare the LC modules

Describes preparing the LC before MS acquisition: switching the LC stream to waste while conditioning/equilibrating the column, purging the LC pump, conditioning the column, setting up real-time actuals monitoring with min/max limits, and configuring real-time plot displays.

  • Switch LC Stream to Waste in Method Editor General tab before conditioning column
  • Apply Now sends only the current LC eluent value; Apply sends all method parameters
  • Purge the LC pump per the pump User Guide
  • Actuals window: set min/max values; out-of-range values shown with red background
  • Monitor baseline to confirm column equilibration
pp. 21-36
Step 3. Prepare the TOF and Q-TOF instrument

Covers tuning the TOF or Q-TOF instrument: when tuning is required, SWARM vs Classic autotune selection, supported sources, tune mix dilution, mass-axis calibration, and switching the LC stream from Waste to MS after tuning. Also covers monitoring MS baseline and using the System Logbook Viewer.

  • Tune first use or after maintenance, service or pump-down and restart
  • Switch LC stream to MS after conditioning column and tuning
  • Monitor TOF/Q-TOF baseline stability and absence of interfering spectra
  • System Logbook Viewer: click Log on Data Acquisition toolbar or enable task bar notification
  • Manual tune recommended only for ion source and optics 1; not after collision cell
pp. 12-14
Data Acquisition windows

Describes the seven Data Acquisition windows: Instrument Status, Actuals, Chromatogram Plot, Spectrum Pane, Method Editor, Sample Run and Worklist. Instrument Status shows device status by text and color coding; Method Editor contains acquisition parameters and, for the 6560, the Acquisition Mode section.

  • Instrument Status window shows each device status as text and color-coded
  • Actuals window shows current values of selected parameters with conditional formatting
  • Spectrum Pane shows spectra from different devices on different tabs
  • Sample Run window supports Override DA Method and Method Type Both Acquisition and DA or DA Only
  • Worklist window runs samples in listed order; Part of method to run options Acquisition Only, Both Acquisition and DA, DA only
pp. 15
Tune window

The Tune window is used to tune the mass spectrometer with automated algorithms or manually, select instrument mode, and calibrate. For the 6560 Ion Mobility Q-TOF it contains the Acquisition Mode section (IM-QTOF or QTOF-Only). Additional small molecule autotune options exist for 6545, 6545XT, 6550 and 6560; 6545XT also supports large molecule autotune.

  • Manual tuning recommended only for ion source and optics 1
  • 6560: select IM-QTOF or QTOF-Only in Acquisition Mode
  • 6545, 6545XT, 6550, 6560: additional small molecule autotune options
  • 6545XT: large molecule autotune available
  • Cannot do small molecule tune after large molecule autotune; upload older system tune file or do system tune first

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
System Tune (SWARM Initial Auto Tune) can only be performed with Mass Range set to Standard (3200 m/z). p.25 ↗ Select Standard mass range before System Tune; Low (1700 m/z) disables System Tune. 6224/6230 TOF and 6530, 6545, 6545XT, 6546, 6550, 6560 Q-TOF
High Resolution Slicer Mode must be selected before performing a System Tune or Initial Auto Tune. p.22 ↗ Set Slicer Mode to High Resolution before these tunes to obtain best tuning results. Instruments with Slicer Mode option
Set Detector Gain and System Tune (TOF or Quadrupole) require mass range 3200 m/z. p.22 ↗ These buttons are grayed out or missing with other mass ranges or unsupported sources. SWARM and Classic autotune instruments
Classic Initial Tune, Standard Tune and Set Detector Gain require Instrument Mode Extended Dynamic Range. p.22 ↗ Select Extended Dynamic Range before running these autotunes; buttons gray out otherwise. Classic autotune TOF/Q-TOF instruments
After changing Instrument Mode, Mass Range or Slicer Mode, the TOF or Q-TOF must be recalibrated. p.22 ↗ Plan a Mass Calibration/Check after any such change before acquiring data. All TOF and Q-TOF models
Repeat Mass Calibration/Check if calibrant ion abundances exceed 480,000 for 6538/6540 Q-TOF or approximately 650,000 for other TOF/Q-TOF instruments. p.33 ↗ Dilute the calibration solution and repeat the check when abundances exceed these limits. 6538/6540 Q-TOF and other TOF/Q-TOF
The 6550 Q-TOF always uses 1:10 diluted calibration solution. p.33 ↗ Do not repeat Quick Tune for dilution on the 6550; dilution is always applied. 6550 Q-TOF
For 6550 Q-TOF small molecule applications, set Quad AMU to 100 in positive polarity and 90 in negative polarity before autotune. p.31 ↗ Click Apply after each polarity change to save the Quad AMU value. 6550 Q-TOF
For 6530 with Dual AJS ESI, 6538 or 6540 in classic autotune, do not click Both; tune TOF first then Quadrupole with a tuning mix change. p.32 ↗ Tuning mix must be changed between TOF and Quadrupole tunes on these models. 6530 with Dual AJS ESI, 6538, 6540
Fast Polarity Switching Enabled restricts autotune to Mass Calibration/Check or FPS System Tune (SWARM) or Mass Calibration/Check or FPS Initial Auto Tune (Classic). p.23 ↗ Polarity and Quadrupole/TOF/Both selections are unavailable when FPS is enabled. 6224/6230 TOF and 6530, 6545, 6545XT, 6546, 6550, 6560 Q-TOF
The 6545XT ships with a dielectric capillary; a resistive capillary must be ordered from Agilent to use Fast Polarity Switching. p.23 ↗ Order the resistive capillary if FPS is required on a 6545XT. 6545XT Q-TOF
NanoESI, Dual nanoESI, MALDI and GC-QTOF sources cannot be used for any automated tune. p.23 ↗ Only Mass Calibration/Check and Quick Tune (if available) are possible with these sources. 6200 Series TOF and 6500 Series Q-TOF
The Dual ESI source is supported for all Autotunes on all instruments. p.34 ↗ Use Dual ESI when a source must support every autotune procedure. All TOF and Q-TOF models
Transmission Tune takes approximately six to ten minutes and does not include detector tune. p.28 ↗ Run Transmission Tune regularly, e.g. when a different sample class is analyzed. SWARM autotune instruments
Check tune duration is approximately 30 seconds. p.27 ↗ A quick check can be used to decide whether further tuning is needed. SWARM autotune instruments
Quick Tune takes 7 to 10 minutes and cannot be run if Quadrupole or Both is selected as the part to tune. p.33 ↗ Select TOF only when running Quick Tune. Classic autotune instruments
Standard Tune takes 10 to 15 minutes for the TOF part and 10 to 15 minutes for the Quadrupole part. p.34 ↗ Allow sufficient time when running a Standard Tune on both parts. Classic autotune instruments
If the Fragile Ions checkbox is marked, Fragmentor Voltage is reduced to minimize fragmentation. p.28 ↗ Expect lower calibrant ion signal but increased transmission of fragile ions such as metabolites and glycans. SWARM autotune instruments
For 6545XT with High (10K - 30K m/z) Mass Range, a large molecule transmission tune is performed during a transmission tune. p.21 ↗ Select the specific mass range with the slider in the Instrument State tab. 6545XT Q-TOF
Manual tuning is recommended only for the front part of the instrument: ion source and optics 1. p.15 ↗ Do not manually tune parameters after the collision cell; manual tuning may give less than optimal results. All TOF and Q-TOF models
MassHunter 10.0 supports only the Agilent 6546; no other TOF or Q-TOF mass spectrometers are supported in that release. p.5 ↗ Use MassHunter B.09.00 for other TOF/Q-TOF models. MassHunter Workstation 10.0

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Autotune can fail p.24 ↗ Tuning mix not diluted on 6530A, 6540A, 6545A, 6545XT, 6546, 6550A or 6560A with Agilent Jet Stream ESI or Dual AJS ESI source Dilute the tuning mix as specified (e.g. 1:10 for 6545/6545XT/6550A/6560A with Dual AJS ESI) Diluting the ESI-L Tune Mix
Calibration not proper in negative ion polarity p.24 ↗ Tuning mix not diluted Dilute the tuning mix to calibrate properly Diluting the ESI-L Tune Mix
Not getting proper calibration in positive ion polarity p.24 ↗ Tuning mix not diluted Dilute the tuning mix Diluting the ESI-L Tune Mix
Problems occur if you dilute the tune mix and tune for positive ion p.24 ↗ Tune mix diluted on 6224 TOF with Dual ESI source Do not dilute the tune mix for the Dual ESI source on the 6224 TOF Diluting the ESI-L Tune Mix
Calibration masses detected out of the 50 to 650K range p.30 ↗ Calibrant ion abundance too high; fragmentor voltage cannot be adjusted low enough System tells you to dilute the calibrant and then try the calibration or autotune again Adjust abundances for optimal calibration (6530A/B)
Check tune results are not acceptable p.27 ↗ Mass calibration/optimization not within acceptance Continue with Transmission Tune; if still unacceptable, run System Tune; if System Tune results are not acceptable, contact your local field service engineer SWARM autotune steps 15-18
Transmission tune results are not acceptable p.28 ↗ Tune not within acceptance Click System Tune (select Standard (3200 m/z) Mass Range and Apply if option not available); if System Tune results are not acceptable, contact your local field service engineer SWARM autotune step 17
Quick Tune results are not acceptable p.33 ↗ Limited parameter set tune not within acceptance Continue with Standard Tune Classic autotune step 19-20
Standard Tune produces unacceptable results p.34 ↗ Standard Tune not within acceptance Run an Initial Auto Tune; after performing an Initial Auto Tune, adjust the Collision Cell Gas Pressure; if this also fails, contact Agilent Field Support Classic autotune step 20
FPS Initial Auto Tune results not acceptable p.33 ↗ Fast Polarity Switching tune not within acceptance Contact Agilent Field Support Classic autotune step 18
Cannot do a small molecule tune after a large molecule autotune p.15 ↗ Large molecule autotune performed on 6545XT Upload an older system tune file first, or do a system tune Tune window NOTE
Error you want to troubleshoot during instrument preparation p.36 ↗ Instrument or software event/error View the System Logbook Viewer: click Log on the Data Acquisition toolbar, or right-click the system task bar, click Enable Notification, then right-click LOG and click Configure View the system logbook for events and errors

Technical Data

Parameter Value Condition
Mass Range (Standard) 3200 m/z p.22 ↗ System Tune / Initial Tune / Set Detector Gain
Mass Range (Low) 1700 m/z p.21 ↗ 6530, 6545, 6545XT, 6546, 6550, 6560 Q-TOF small molecule autotune
Mass Range (High) 10K - 30K m/z p.21 ↗ 6545XT large molecule transmission tune
Mass Range (High) 10000 m/z p.27 ↗ 6546
Low mass autotune range 50 - 1700 m/z p.27 ↗ Low (1700 m/z) Mass Range, Quadrupole or Both
Calibrant abundance limit 480,000 counts p.33 ↗ 6538/6540 Q-TOF; repeat Mass Calibration/Check above this
Calibrant abundance limit approximately 650,000 counts p.33 ↗ other Q-TOF and TOF instruments
Calibrant abundance range 50 to 650K counts p.30 ↗ 6530A/B Adjust abundances for optimal calibration
Check tune duration approximately 30 seconds p.27 ↗ SWARM Check tune
Transmission Tune duration approximately six to ten minutes p.28 ↗ SWARM Transmission Tune
Mass Calibration / Check duration (TOF) a few minutes p.33 ↗ Classic autotune
Mass Calibration / Check duration (Quadrupole) 2 to 5 minutes p.33 ↗ Classic autotune
Quick Tune duration 7 to 10 minutes p.33 ↗ Classic autotune
Standard Tune duration (TOF part) 10 to 15 minutes p.34 ↗ Classic autotune
Standard Tune duration (Quadrupole part) 10 to 15 minutes p.34 ↗ Classic autotune
Quad AMU (Positive polarity) 100 p.31 ↗ 6550 Q-TOF small molecule applications
Quad AMU (Negative polarity) 90 p.31 ↗ 6550 Q-TOF small molecule applications
Tune mix dilution (positive ion TOF tune) 1:10 diluted tune mix p.24 ↗ 6530A with Dual AJS ESI or 6540
Tune mix dilution (negative ion TOF tune) 1:40 diluted tune mix p.24 ↗ 6530A with Dual AJS ESI or 6540
Tune mix dilution (positive ion TOF tune) Undiluted tune mix p.24 ↗ 6520B or 6530 with Dual ESI
Tune mix dilution (negative ion TOF tune) Undiluted tune mix p.24 ↗ 6520B or 6530 with Dual ESI
Tune mix dilution (polarity switching tunes) Undiluted tune mix p.24 ↗ 6530A with Dual AJS ESI or 6540; 6520B or 6530 with Dual ESI
Tune mix dilution (Quad tunes) Undiluted tune mix p.24 ↗ 6530A with Dual AJS ESI or 6540; 6520B or 6530 with Dual ESI
Tune mix dilution (SWARM, both polarities) 1:10 diluted tune mix p.24 ↗ 6545, 6545XT, 6550A, 6560A with Dual AJS ESI
Tune mix dilution (positive mode) 1:10 diluted tune mix p.24 ↗ 6530 with Dual AJS ESI
Tune mix dilution (negative mode) 1:40 diluted tune mix p.24 ↗ 6530 with Dual AJS ESI
Tune mix dilution undiluted tune mix p.24 ↗ 6538A
Tune mix dilution (positive ion) 1:10 dilution p.24 ↗ 6230 TOF with AJS ESI or Dual AJS ESI
Tune mix dilution (negative ion) 1:40 dilution p.24 ↗ 6230 TOF with AJS ESI or Dual AJS ESI
Tune mix dilution do not dilute p.24 ↗ 6224 TOF with Dual ESI, positive ion
Tune mix dilution 1:10 dilution p.33 ↗ 6550 Q-TOF (always)
LC driver version A.02.19 SR2 p.5 ↗ MassHunter 10.0
LC driver version A.02.19 p.5 ↗ MassHunter B.09.00
Supported operating systems Windows 7 and Windows 10 p.5 ↗ MassHunter B.09.00