6850 Series Control Module User Information
Technical Reference Summary
Covered Models & Document Scope
Covers the Agilent 6850 Gas Chromatograph Control Module (Part No. G2629-90329, first edition March 2004, replaces G2629-90327 and G2629-90328). Includes operation with split/splitless, purged packed, programmable temperature vaporization (PTV), and cool on-column inlets; thermal conductivity (TCD), flame ionization (FID), microcell electron capture (ECD), and flame photometric (FPD) detectors; auxiliary EPC channels; valves; and LAN cards J2552B, J4100A, and Lantronics.
Section-by-Section Technical Breakdown
Describes the physical Control Module: display, keyboard, connecting cable to the 6850 GC, and a PCMCIA flash memory card slot. Explains that the module is operated by entering instructions from the keyboard into screens, which can be stored as a named method.
- Display, keyboard, and cable to 6850 GC
- PCMCIA flash memory card slot on left side
- Instructions entered via keyboard into screens
- Screens can be stored as a named method
Explains the Status screen as the starting point for all operations and the function of F1-F8 keys. Provides a worked example of setting up the column so the instrument can convert flows to pressures and vice versa.
- Status screen is the starting point; press Esc repeatedly to return
- F1-F5 are navigation keys; F6-F8 are action keys
- Example: Status/Setup/Column Setup to configure column source and outlet
- Table 2 lists all Control Module screens
Covers global configuration required before first use and whenever hardware changes: display contrast, display and keyboard functions, time and date, RS-232 port, IP address settings, column, automation, inlet, detector, and auto prep run. Notes that configuration settings directly affect methods.
- Configure before first use and when changing/adding hardware
- Unconfigured devices may be unavailable or have no settable values
- Configure carrier gas first, then column, then inlet; detectors anytime
- Includes display contrast, Local UI locks, clock, RS-232, and LAN settings
Describes the active method concept, saving methods as named methods or the SERVICE method, restoring the default method, and using PC cards to store and transfer methods. Five named methods and a SERVICE method can be stored in GC memory.
- Active method is the set of control values presently operating the GC
- Five named methods and a SERVICE method can be stored
- Configure carrier gas first, then column, then inlet
- PC card must be inserted before connecting to GC; disconnect before changing cards
Covers hydrogen and column shutdown behavior, electronic pneumatic control, turning gas flows on and off, interpreting flow and pressure displays, configuring the column, column modes, initial flow/pressure, flow/pressure programming, auxiliary channels, and solving flow and pressure problems.
- Hydrogen shutdown closes carrier supply valve, opens split valve, turns off oven and heated zones
- Column shutdown turns off oven heater and opens oven flaps halfway
- All flow and pressure displays corrected to NTP (25°C, 1 atm)
- Gas control valves may leak up to 0.2 mL/min in OFF state
Describes injector control parameters including plunger pumps, sample size, viscosity delay, slow plunger, wash parameters, needle depth, dwell times, and solvent parameters. Also covers sequence parameters, controlling a sequence, Run Table, and Clock Table.
- Plunger Pumps expel bubbles before drawing sample
- Slow plunger reduces injection speed from about 100 µL/sec to about 5 µL/sec with a 10 µL syringe
- Wash order: solvent A pre-washes, solvent B pre-washes, sample pre-washes, injection, solvent A post-washes, solvent B post-washes
- Default needle depth 0 mm includes safety factor to avoid vial bottom
Covers using hydrogen, options, inlet modes, inlet and column relationship, inlet setup, prep run, setting the inlet mode, split/splitless terms, pressure pulse modes, split mode, splitless mode, and gas saver.
- Gas saver mode reduces carrier gas consumption between runs
- Direct entry of split ratios provided column is configured
- Pressure pulse modes available
- Prep run functionality described
Covers using hydrogen, inlet and column controls, inlet setup, and using a purged packed inlet.
- Using hydrogen with purged packed inlet
- Inlet and column controls
- Inlet setup procedure
- Operating a purged packed inlet
Covers using hydrogen, inlet modes, inlet and column, inlet setup, setting the inlet mode, heating the inlet, PTV terms, pulsed modes, split mode, splitless mode, and solvent vent mode.
- PTV inlet modes and setup
- Heating the inlet
- Pulsed modes
- Solvent vent mode
Covers inlet temperature and operating the cool on-column inlet.
- Inlet temperature control
- Operating procedure for cool on-column inlet
Covers using hydrogen, operation conditions, TCD parameters, makeup gas, polarity, signal selection, analog output, and using the TCD.
- TCD parameters
- Makeup gas
- Polarity and signal selection
- Analog output
Covers using hydrogen, detector operation notes, jets, electrometer, makeup gas, signal selection, analog output, automatic reignition—Lit Offset, FID parameters, and using the FID.
- Jets and electrometer
- Makeup gas
- Automatic reignition with Lit Offset
- FID parameters
Covers linearity, detector gas, temperature, electrometer, analog output, and operating the detector.
- Linearity
- Detector gas and temperature
- Electrometer
- Analog output
Covers oven capabilities, oven safety, oven setup, creating an isothermal run, temperature programming, and column compensation.
- Oven capabilities and safety
- Oven setup
- Isothermal run creation
- Temperature programming and column compensation
Covers using hydrogen, general information, using lit offset, igniting the flame, using the electrometer, signal selection, selecting the makeup gas mode, heater configuration, FPD parameters, and using the FPD.
- Using lit offset
- Igniting the flame
- Selecting makeup gas mode
- Heater configuration and FPD parameters
Covers valve types, configuring valves, sample valves, multi valve with sample valve, controlling valves manually, and setting the valve box temperature.
- Valve types
- Configuring valves
- Sample valves and multi valve with sample valve
- Controlling valves manually and setting valve box temperature
Covers the Service Screen, Log Book, Diagnostics, Calibration, Maintenance, and Update functions.
- Service Screen and Log Book
- Diagnostics
- Calibration
- Maintenance and Update functions
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Gas supply pressure should be at least 10 psi greater than the desired setpoint. p.49 ↗ | Ensure supply regulators are set high enough to avoid inlet/detector shutdown from failure to reach setpoint. | whole series |
| When a gas control valve is driven to the OFF state, there may still be a small flow of as much as 0.2 mL/min through it. p.37 ↗ | Do not rely on OFF state for absolute zero flow; consider additional shutoff if zero flow is critical. | whole series |
| All flow and pressure displays are corrected to Normal Temperature and Pressure (NTP): 25°C and 1 atmosphere. p.37 ↗ | Setpoints and displayed flows are referenced to NTP, not local conditions; the GC measures atmospheric pressure and temperature to compensate. | whole series |
| For capillary columns, recommended carrier gas flow rates are proportional to column diameter and are 20% lower for helium than for hydrogen. p.42 ↗ | When switching from hydrogen to helium carrier, reduce flow setpoints by approximately 20% for the same column. | capillary columns |
| Recommended carrier gas flow for a 250-µm id capillary column is 2.5 mL/min hydrogen or 2.0 mL/min helium. p.42 ↗ | Use these as starting points for method development with 250-µm columns. | capillary columns |
| If the column is not defined (length or ID entered as 0), only pressure setpoints are available with a split/splitless inlet, or total flow with a purged packed inlet. p.39 ↗ | Define column dimensions to enable flow control, split ratio entry, and column modes. | whole series |
| Set column parameters with the oven at its initial temperature because pneumatic setpoints change with oven temperature due to column resistance and gas viscosity changes. p.41 ↗ | Configure column at initial oven temperature to avoid confusion from changing setpoints during temperature programs. | whole series |
| Slow plunger reduces plunger speed during injection from normal (about 100 µL/sec with a 10 µL syringe) to about 5 µL/sec. p.52 ↗ | Use slow plunger for viscous samples or to reduce boil-out from a hot needle. | 6850 automatic injector |
| Default needle depth offset is 0 mm, which includes a small safety factor to avoid striking the bottom of the vial. p.53 ↗ | Adjust depth offset to sample headspace instead of liquid or solid sample. | 6850 automatic injector |
| For most uses, both dwell times will be zero, giving a fast injection with a minimum of boil-out from a hot needle. p.54 ↗ | Set dwell times to zero unless method requires otherwise. | 6850 automatic injector |
| Five named methods and a method named SERVICE can be stored in GC memory. p.28 ↗ | Plan method storage accordingly; use PC cards for additional method library. | whole series |
| PC card must be inserted into the Control Module before connecting to the GC; to change cards, disconnect the Control Module first. p.31 ↗ | Follow this sequence to avoid card or communication errors. | whole series |
| If a J2552B or J4100A LAN card is installed, the message “Supports BootP Control Mode Only” is displayed; setting LAN address from the Control Module is not recommended because the connection will be very slow. p.18 ↗ | Use BootP for J2552B/J4100A; use Lantronics card for locally entered IP settings via Control Module. | LAN cards J2552B, J4100A |
| Hydrogen safety shutdown closes the carrier supply valve to the inlet, opens the split valve in the split/splitless inlet, turns off the oven heater and opens oven flaps fully, and turns off inlet, detector, and auxiliary heated zones. p.36 ↗ | To recover, fix the cause (tank valve closed, serious leak), turn instrument off, then back on. | whole series when hydrogen is configured |
| Column shutdown turns off the oven heater and opens oven flaps halfway to avoid column damage from heat without carrier gas. p.36 ↗ | To recover, fix the cause and turn the oven and offending inlet or auxiliary channel back on. | whole series |
| Do not use or select air as the gas for the column input channel. p.47 ↗ | Selecting air for column input can cause unsafe or incorrect operation. | auxiliary channels |
| When hydrogen is used, dangerously high flows are possible if insufficient flow resistance is provided downstream of the supply tube. p.47 ↗ | Ensure adequate downstream flow restriction when using hydrogen on auxiliary channels. | auxiliary channels |
| To disable a pressure ramp and all further ramps, enter a rate of 0.0. p.48 ↗ | Use rate 0.0 to terminate ramp programming in auxiliary pressure ramps. | auxiliary channels |
| The GC has an internal clock used for logging methods and errors; the clock resets when Done is pressed after setting date and time. p.17 ↗ | Set time and date to ensure accurate method and error logs. | whole series |
| The Control Module can lock local keyboard, local start key, remote start, sequence execution, clock table access, clock table execution, and method loading via the Locks screen. p.16 ↗ | Use locks to prevent unauthorized or accidental changes when GC is used as a dedicated analyzer or controlled remotely. | whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| A gas does not reach the setpoint pressure or flow p.49 ↗ | The gas supply pressure is too low to reach the setpoint. | The pressure at the supply should be at least 10 psi greater than the desired setpoint. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | A large leak is present somewhere in the system. | Use an electronic leak detector to find leaks; correct them. Check the column—a broken column is a very large leak. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | Gas saver flow rate is not high enough to maintain the highest column-head pressure during a run. | Be sure that the gas saver flow rate is high enough to maintain the highest column-head pressure during a run. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | The flow is too low for the column in use. | Increase flow or use appropriate column. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | The column is plugged or mis-installed. | Check and correct column installation or replace column. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | The inlet or detector pressure sensor is not operating correctly. | Contact your Agilent service representative. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | Split/splitless inlet: The split ratio is too low. | Increase the amount of split flow. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | Split/splitless inlet: The inlet proportional control valve is stuck because of contamination or other fault. | Contact your Agilent service representative. | |
| A gas does not reach the setpoint pressure or flow p.49 ↗ | Purged packed inlet: The inlet control valve is stuck closed because of contamination or other fault. | Contact your Agilent service representative. | |
| A gas exceeds the setpoint pressure or flow p.50 ↗ | The pressure sensor for that device is not operating properly. | Contact your Agilent service representative. | |
| A gas exceeds the setpoint pressure or flow p.50 ↗ | Split/splitless inlet: The split ratio is too high. | Decrease the split ratio. | |
| A gas exceeds the setpoint pressure or flow p.50 ↗ | Split/splitless inlet: The proportional control valve is stuck closed. | Contact your Agilent service representative. | |
| A gas exceeds the setpoint pressure or flow p.50 ↗ | Split/splitless inlet: The trap on the split vent line is contaminated. | Contact your Agilent service representative. | |
| A gas exceeds the setpoint pressure or flow p.50 ↗ | Purged packed inlet: The inlet proportional control valve is stuck open. | Contact your Agilent service representative. | |
| The inlet pressure or flow fluctuates p.50 ↗ | A small leak is present in the flow system. | Use an electronic leak detector to find leaks; correct them. Also check for leaks in the gas supply plumbing. | |
| The inlet pressure or flow fluctuates p.50 ↗ | Large restrictions are present in the split/splitless inlet, such as a blockage in a liner or the split vent trap. | Make sure you are using the correct liner. Replace liners with large pressure drops caused by design or tight packaging. If the liner does not appear to be causing the problem, the split vent trap may be blocked. Contact your Agilent service representative. | |
| The inlet pressure or flow fluctuates p.50 ↗ | Extreme changes in room temperature during runs. | Correct laboratory temperature problem or move the instrument to a more suitable location. | |
| The inlet pressure or flow fluctuates p.50 ↗ | Large volumes have been added to the system (this may occur if you are using a sampling valve). | Decrease the sample volume. |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Carrier gas flow rate | 0.5 mL/min p.42 ↗ | Capillary 50-µm id, hydrogen |
| Carrier gas flow rate | 0.4 mL/min p.42 ↗ | Capillary 50-µm id, helium |
| Carrier gas flow rate | 1.0 mL/min p.42 ↗ | Capillary 100-µm id, hydrogen |
| Carrier gas flow rate | 0.8 mL/min p.42 ↗ | Capillary 100-µm id, helium |
| Carrier gas flow rate | 2.0 mL/min p.42 ↗ | Capillary 200-µm id, hydrogen |
| Carrier gas flow rate | 1.6 mL/min p.42 ↗ | Capillary 200-µm id, helium |
| Carrier gas flow rate | 2.5 mL/min p.42 ↗ | Capillary 250-µm id, hydrogen |
| Carrier gas flow rate | 2.0 mL/min p.42 ↗ | Capillary 250-µm id, helium |
| Carrier gas flow rate | 3.2 mL/min p.42 ↗ | Capillary 320-µm id, hydrogen |
| Carrier gas flow rate | 2.6 mL/min p.42 ↗ | Capillary 320-µm id, helium |
| Carrier gas flow rate | 5.3 mL/min p.42 ↗ | Capillary 530-µm id, hydrogen |
| Carrier gas flow rate | 4.2 mL/min p.42 ↗ | Capillary 530-µm id, helium |
| Carrier gas flow rate | 30 mL/min p.42 ↗ | Packed metal 1/8-in. id, hydrogen |
| Carrier gas flow rate | 30 mL/min p.42 ↗ | Packed metal 1/8-in. id, helium |
| Carrier gas flow rate | 60 mL/min p.42 ↗ | Packed metal 1/4-in. id, hydrogen |
| Carrier gas flow rate | 30-60 mL/min p.42 ↗ | Packed metal 1/4-in. id, helium |
| Carrier gas flow rate | 30 mL/min p.42 ↗ | Packed glass 2-mm id, hydrogen |
| Carrier gas flow rate | 30 mL/min p.42 ↗ | Packed glass 2-mm id, helium |
| Carrier gas flow rate | 30-60 mL/min p.42 ↗ | Packed glass 4-mm id, hydrogen |
| Carrier gas flow rate | 30-60 mL/min p.42 ↗ | Packed glass 4-mm id, helium |
| Recommended gas pressure | 25 (170) psi (kPa) p.43 ↗ | Capillary 0.10 µm film, 10 m length |
| Recommended gas pressure | 6 (40) psi (kPa) p.43 ↗ | Capillary 0.20 µm film, 10 m length |
| Recommended gas pressure | 3.7 (26) psi (kPa) p.43 ↗ | Capillary 0.25 µm film, 10 m length |
| Recommended gas pressure | 2.3 (16) psi (kPa) p.43 ↗ | Capillary 0.32 µm film, 10 m length |
| Recommended gas pressure | 0.9 (6.4) psi (kPa) p.43 ↗ | Capillary 0.53 µm film, 10 m length |
| Recommended gas pressure | 39 (270) psi (kPa) p.43 ↗ | Capillary 0.10 µm film, 15 m length |
| Recommended gas pressure | 9 (61) psi (kPa) p.43 ↗ | Capillary 0.20 µm film, 15 m length |
| Recommended gas pressure | 5.6 (39) psi (kPa) p.43 ↗ | Capillary 0.25 µm film, 15 m length |
| Recommended gas pressure | 3.4 (24) psi (kPa) p.43 ↗ | Capillary 0.32 µm film, 15 m length |
| Recommended gas pressure | 1.4 (9.7) psi (kPa) p.43 ↗ | Capillary 0.53 µm film, 15 m length |
| Recommended gas pressure | 68 (470) psi (kPa) p.43 ↗ | Capillary 0.10 µm film, 25 m length |
| Recommended gas pressure | 15 (104) psi (kPa) p.43 ↗ | Capillary 0.20 µm film, 25 m length |
| Recommended gas pressure | 9.5 (65) psi (kPa) p.43 ↗ | Capillary 0.25 µm film, 25 m length |
| Recommended gas pressure | 5.7 (40) psi (kPa) p.43 ↗ | Capillary 0.32 µm film, 25 m length |
| Recommended gas pressure | 2.3 (16) psi (kPa) p.43 ↗ | Capillary 0.53 µm film, 25 m length |
| Recommended gas pressure | 83 (570) psi (kPa) p.43 ↗ | Capillary 0.10 µm film, 30 m length |
| Recommended gas pressure | 18 (126) psi (kPa) p.43 ↗ | Capillary 0.20 µm film, 30 m length |
| Recommended gas pressure | 12 (80) psi (kPa) p.43 ↗ | Capillary 0.25 µm film, 30 m length |
| Recommended gas pressure | 7 (48) psi (kPa) p.43 ↗ | Capillary 0.32 µm film, 30 m length |
| Recommended gas pressure | 2.8 (19) psi (kPa) p.43 ↗ | Capillary 0.53 µm film, 30 m length |
| Recommended gas pressure | 32 (220) psi (kPa) p.43 ↗ | Capillary 0.20 µm film, 50 m length |
| Recommended gas pressure | 20 (135) psi (kPa) p.43 ↗ | Capillary 0.25 µm film, 50 m length |
| Recommended gas pressure | 12 (81) psi (kPa) p.43 ↗ | Capillary 0.32 µm film, 50 m length |
| Recommended gas pressure | 4.7 (32) psi (kPa) p.43 ↗ | Capillary 0.53 µm film, 50 m length |
| Recommended gas pressure | 39 (267) psi (kPa) p.43 ↗ | Capillary 0.20 µm film, 60 m length |
| Recommended gas pressure | 24 (164) psi (kPa) p.43 ↗ | Capillary 0.25 µm film, 60 m length |
| Recommended gas pressure | 14 (98) psi (kPa) p.43 ↗ | Capillary 0.32 µm film, 60 m length |
| Recommended gas pressure | 5.6 (39) psi (kPa) p.43 ↗ | Capillary 0.53 µm film, 60 m length |