System overview, abbreviations, and quadrupole analyzer description
Overviewpp. 15-22
Vacuum gauge, safety warnings, symbols, and intended use
Procedurepp. 23-34
Column installation, conditioning, and interface tip seal procedures
Procedurepp. 35-58
EI setup, pumping down, temperature and collision cell settings, autotune
Procedurepp. 59-76
CI setup, reagent gas flow control, CI autotune, and mode switching
Procedurepp. 77-108
Tools, safety, vacuum and analyzer maintenance, LE-EI ion source procedures
Procedurepp. 109-126
Axial CI ion source removal, cleaning, assembly, and filament replacement
p. 67table, 13 rows
p. 24table, 9 rows
p. 45table, 2 rows
p. 78table, 14 rows
p. 90table, 18 rows
p. 95table, 18 rows
p. 116table, 17 rows
p. 12table, 26 rows
p. 42table, 5 rows
p. 63table, 2 rows
Table 7 Typical CI application settings
Parameter
Methane Positive
Methane Negative
Isobutane Positive
Isobutane Negative
Ammonia Positive
Ammonia Negative
Ion polarity
Positive
Negative
Positive
Negative
Positive
Negative
Emission
150 A
50 A
150 A
50 A
150 A
50 A
Electron energy
150 eV
150 eV
150 eV
150 eV
150 eV
150 eV
Filament
1
1
1
1
1
1
Repeller
3 V
3 V
3 V
3 V
3 V
3 V
Ion focus
130 V
130 V
130 V
130 V
130 V
130 V
Entrance lens offset
20 V
20 V
20 V
20 V
20 V
20 V
Shutoff valve
Open
Open
Open
Open
Open
Open
Suggested flow
20%
40%
20%
40%
20%
40%
Source temp
250 °C
150 °C
250 °C
150 °C
250 °C
150 °C
Quad temp
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
Interface temp
280 °C
280 °C
280 °C
280 °C
280 °C
280 °C
Autotune
Yes
Yes
No
Yes
No
Yes
Emission values printed as '150 A' and '50 A' (microampere symbol lost in text extraction).
Table 2 Gas flows
Feature
Gas flows
High-vacuum pump 1
Split-flow turbo
High-vacuum pump 2
Turbo
High-vacuum pump 3
Turbo
Optimal carrier gas flow, mL/min
1.0 to 1.5
Collision cell Nitrogen gas flow, mL/min
1
Collision cell Helium gas flow, mL/min
4
Maximum recommended gas flow, mL/min*
2.0
Maximum gas flow, mL/min†
2.4
Maximum column id
0.32 mm (30 m length)
* Total gas flow into the MS = column flow + collision cell gas flow. † Expect degradation of spectral performance and sensitivity.
Table 4 Recommended temperature settings
Zone
EI operation
MS Source
230 °C default. High limit 350 °C. Application dependent.
MS Quad
150 °C
Table 9 Maintenance schedule
Task
Every week
Every 6 months
Every year
As needed
Check the foreline pump oil level
X
Check the calibration vial(s)
X
Replace the foreline pump oil*
X
Replace the tip seals on the optional IDP
X
Replace the filters on the optional IDP
X
Clean the ion source
X
Check the carrier gas trap(s) on the GC and MS
X
Replace the worn out parts
X
Replace GC gas supplies
X
Replace CI reagent gas supplies
X
Check the foreline pump
X
Tune the MS
X
Leak check the system
X
Lubricate side plate or vent valve O-rings†
X
* Or as needed, every 3 months if ammonia is being used as CI reagent gas. † Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals can interfere with their correct function.
Table 10 Parts list for LE-EI ion source (Figure 21)
Item number
Item description
Part number
1
Source finger grip
G7002-20008
2
Filament block
G7002-20019
3
Extractor lens (5)*, with 3 mm opening
G7004-20061
4
Ceramic insulator for extractor
G7002-20064
5
Entrance lens assembly (1)*
G7250-20075
6
Ion focus lens (2)*
G7004-20068
7
Lens insulator/holder
G7002-20074
8
M2 × 0.4 screw × 12 mm long gold plated screw
G7002-20083
9
Source body
G7002-20084
10
Post extractor lens 2 (3)*
G7004-20090
11
Post extractor lens 1 (4)*
G7004-20004
12
M2 × 6 mm gold plated screw
G7002-20109
13
Locking ring lens insulator
G7002-20126
14
Filament assembly, HE, Single
G3850-60021
15
Ring heater/sensor assembly
G7002-60043
16
Source mount 1.5 mm
G7002-60053
17
Repeller assembly
G7002-60057
Complete assembly
Ion Source Assy, LE-EI 7250A
G7250-67170
*The number in parenthesis is the number engraved on the lens.
Table 11 Parts list for LE-EI ion source (Figure 24)
Item number
Item description
Part number
1
Source finger grip
G7002-20008
2
Filament block
G7002-20019
3
Extractor lens (5)*, with 3 mm opening
G7004-20061
4
Ceramic insulator for extractor
G7002-20064
5
Entrance lens assembly (1)*
G7250-20075
6
Ion focus lens (2)*
G7004-20068
7
Lens insulator/holder
G7002-20074
8
M2 × 0.4 screw × 12 mm long gold plated screw
G7002-20083
9
Source body
G7002-20084
10
Post extractor lens 2 (3)*
G7004-20090
11
Post extractor lens 1 (4)*
G7004-20004
12
M2 × 6 mm gold plated screw
G7002-20109
13
Locking ring lens insulator
G7002-20126
14
Filament assembly, HE, single
G3850-60021
15
Ring heater/sensor assembly
G7002-60043
16
Source mount 1.5 mm
G7002-60053
17
Repeller assembly
G7002-60057
Complete assembly
Ion Source Assy, LE-EI 7250A
G7250-67170
*The number in parenthesis is the number engraved on the lens.
Table 12 Parts list for Axial CI ion source
Item number
Item description
Part number
1
Source finger grip
G7002-20008
2
Filament block
G7002-20019
3
Filament, HES CI
G7250-60075
4
5973 Gold Fil Screw (M2 × 0.4 10MM-LG)
G1999-20021
5
in-WSHR-Shoulder 0.25 0.37 0.06 Alumina
3050-2670
6
HES CI source mount assembly
G7250-60009
7
HES CI repeller assy
G7250-60102
8
GCMS Ring heater/sensor assembly
G7002-60058
9
HES CI Body
G7250-20097
10
Drawout plate
G1999-20446
11
HES CI Drawout cylinder
G7250-20098
12
HES Ramp CI Ion focus
G7250-20096
13
Entrance lens assy, STD, Half-Ramped
G7250-20075
14
HES CI Lens insulator
G7250-20095
15
Locking ring for lens insulator
G7002-20126
16
Screw, M2 × 6 Pan Head Torx, gold plated
G7002-20109
Complete assembly
Axial CI Ion Source Assy (Support)
G7250-67404
Table 1 Abbreviations
Abbreviation
Definition
AC
Alternating current
ALS
Automatic liquid sampler
CC
Collision cell
CI
Chemical ionization
DC
Direct current
EI
Electron impact
EPC
Electronic pneumatic control
eV
Electron volt
GC
Gas chromatograph
GC/MS
Gas chromatograph mass spectrometer
id
Inside diameter
IDP
Isolated dry pump
LAN
Local Area Network
LE-EI
Low energy EI
m/z
Mass-to-charge ratio
MFC
Mass flow controller
MS
Mass spectrometer
OFN
Octafluoronaphthalene (sample)
PFTBA
Perfluorotributylamine (calibrant)
Q-TOF
Quadrupole time-of-flight
Quad
Quadrupole mass filter
RFI
Radio frequency interference
RFPA
Radio frequency power amplifier
TOF
Time-of-flight
Torr
Unit of pressure, 1 mm Hg
Turbo
Turbomolecular vacuum pump
Table 3 Influence of carrier and collision cell gas flows on vacuum
Column flow (mL/min)
CC Gas On N2 = 1 mL/min He = 4 mL/min Rough Vac (mTorr)
CC Gas On N2 = 1 mL/min He = 4 mL/min Quad Vac (Torr)
CC Gas On N2 = 1 mL/min He = 4 mL/min TOF Vac (Torr)
CC Gas Off Rough Vac (mTorr)
CC Gas Off Quad Vac (Torr)
CC Gas Off TOF Vac (Torr)
0.7
1.25 E+02
3.50 E-05
2.61 E-07
2.70 E+01
5.78 E-07
1.58 E-07
1
1.30 E+02
3.49 E-05
2.61 E-07
3.61 E+01
6.21 E-07
1.58 E-07
1.2
1.34 E+02
3.49 E-05
2.61 E-07
4.13 E+01
6.55 E-07
1.58 E-07
2
1.48 E+02
3.51 E-05
2.61 E-07
6.18 E+01
8.09 E-07
1.59 E-07
3
1.63 E+02
3.52 E-05
2.61 E-07
8.45 E+01
9.99 E-07
1.60 E-07
Pressures are approximate and will vary from instrument to instrument by as much as 30%.
Table 5 Temperatures for CI operation
Ion source
Analyzer
GC/MS interface
PCI
300 °C
150 °C
280 °C
NCI
150 °C
150 °C
280 °C
Engineering Facts
14 facts · every claim linked to a page
Safety
Hydrogen gas must not be used with this instrument under any circumstances.
Carrier gas selection must exclude hydrogen; using it violates the manufacturer's safety requirements.
Source
“The use of hydrogen gas is specifically prohibited with this product.”
Installation
The analyzer side plate top thumbscrew must be finger-tight only, and the collision cell top plate shipping brackets must remain fastened during normal operation.
Improper torque causes vacuum leaks, and removing the shipping brackets compromises explosion protection.
Source
“You MUST ensure the top thumbscrew on the analyzer side plate is fastened finger-tight. Do not over-tighten the thumbscrews; this can cause air leaks.”
Safety
Combustible or absorbent materials must not be left in the oil pan under the foreline pump.
Housekeeping around the foreline pump is a fire-safety requirement during installation and operation.
Source
“Combustible materials (or flammable/nonflammable wicking material) placed under, over, or around the foreline (roughing) pump constitutes a fire hazard.”
Installation
The GC/MS interface temperature should be slightly above the maximum GC oven temperature but must never exceed the maximum column temperature.
This sets the allowable interface setpoint relative to oven and column limits to avoid column damage.
Source
“the interface temperature should be slightly higher than the maximum GC oven temperature, but never higher than the maximum column temperature.”
Operation
The quadrupole setpoint is software-limited to 200 °C and the ion source to 350 °C.
Defines the maximum allowable heated-zone setpoints when configuring the method.
Source
“The software will not allow you to exceed 200 °C for the quadrupole, or 350 °C for the ion source.”
Operation
Carrier gas flow must be turned on before any GC heated zones are energized.
Prevents column damage during pumpdown and startup sequencing.
Source
“Do not turn on any GC heated zones until carrier gas flow is on. Heating a column with no carrier gas flow will damage the column.”
Operation
After the 'Okay to run' message, wait 2 hours for thermal equilibrium before acquiring data.
Data acquired before thermal equilibrium may not be reproducible, so this wait is required for valid results.
Source
“After the message Okay to run appears, wait 2 hours for the MS to reach thermal equilibrium.”
Operation
Turbo pump speed should reach 80% within 10–15 minutes and eventually 95%; otherwise the electronics shut off the foreline pump and a power cycle is required.
Provides the acceptance criterion for a successful pumpdown and the recovery action if it fails.
Source
“Within 10 to 15 minutes, the turbo pump speed should be up to 80%.”
Operation
The MS must never be vented by admitting air through either end of the foreline hose.
Defines the only acceptable venting method to avoid contaminating the analyzer.
Source
“Never vent the MS by allowing air in through either end of the foreline hose.”
Maintenance
When running ammonia five or more hours per day, the foreline pump must be ballasted with air at least one hour per day, and the MS purged with methane after ammonia use.
Ammonia CI imposes a mandatory daily maintenance routine to protect pump seals.
Source
“When running ammonia for five or more hours a day, the foreline pump must be ballasted (flushed with air) for at least one hour a day to minimize damage to pump seals.”
Operation
The ammonia supply pressure must be kept below 5 psig and the tank kept upright below the flow module level.
Higher pressure can condense ammonia to liquid and damage the flow module.
Source
“The pressure of the ammonia supply must be less than 5 psig.”
Maintenance
Do not use compressed air to blow dust out of the optional IDP foreline pump when replacing its tip seal.
Compressed air would spread potentially hazardous chemical-laden dust into the laboratory.
Source
“When replacing the pump's tip seal, do NOT use compressed air to blow out this dust.”
Selection
The instrument mainframe weighs 159 kg, with the foreline pump adding 25 kg (DS202 rotary vane) or 45 kg (IDP).
Bench/site load-bearing and rigging requirements depend on these masses.
Source
“The weight of the instrument mainframe is 159 kg.”
Selection
The flight tube extends 136 cm above the top of the instrument, which must be accounted for in lab ceiling/clearance planning.
Vertical clearance above the bench is a site-preparation constraint.
Source
“The flight tube extends 136 cm over the top of the instrument.”