Induztr
Engineering Digest Document Viewer Related Models

Agilent 7820A Gas Chromatograph Maintaining Your GC

Technical Reference Summary

Covered Models & Document Scope

Maintenance manual for the Agilent 7820A Gas Chromatograph (GC). Covers routine maintenance of capillary columns, Split/Splitless inlet, Purged Packed inlet, Packed Column inlet, COC inlet, FID, TCD, µECD, NPD, FPD+, G3435A/G3436A FPD, PCM, and valves. Includes procedures for removing covers and making Swagelok connections. Manual part number G4350-90014, fifth edition, August 2016.

Section-by-Section Technical Breakdown

pp. 13-14
Maintenance Methods for the 7820A GC

Describes recommended maintenance methods to prepare the GC before maintenance. The General GC Maintenance Method sets oven, inlet, and detector temperatures to 35 °C, inlet pressures to 0.0, and turns off detector flows and high voltages. The Inlet Maintenance Method leaves detectors at operating temperature while preparing the inlet. Zones must reach < 70 °C before general maintenance.

  • General GC Maintenance Method: oven 35 °C, all inlets 35 °C and 0.0 pressure, all detectors 35 °C, detector flows Off.
  • For FPD+ maintenance, turn off GC and unplug power cord.
  • For FID and NPD, turn electrometer Off to disable high voltage.
  • For TCD, turn filament Off to prevent damage if exposed to air while hot.
  • Inlet Maintenance Method: inlets Off and 0.0 pressure, detectors may remain at operating setpoints.
  • Wait for zones to reach < 70 °C before performing general GC maintenance.
  • Inlets and detectors at operating temperature may require 12 hours or longer to reach maintenance setpoints.
pp. 17-21
Removing Covers

Provides procedures for removing the detector top cover, pneumatics cover, side cover, and electronics cover as needed for routine maintenance. Only the listed covers should be removed; removing other covers can compromise safety features.

  • Detector top cover removal procedure.
  • Pneumatics cover protects flow manifolds; disconnect vent tubing from split and purge vents before removal.
  • Side cover removal depends on date of manufacture and number of screws.
  • Electronics cover removal differs depending on whether an FPD+/FPD is installed; with FPD+/FPD, remove PMT tube and heat shield first.
  • Raising the electronics cover exposes GC electronics.
pp. 24-33
Maintaining Capillary Columns

Covers consumables and parts for capillary columns, installation of column hangers and clips, conditioning a capillary column, cutting a loop from a column, and reversing a column to bake out contaminants. Includes part numbers for ferrules, nuts, and column hardware by column inner diameter.

  • Conditioning: set minimum velocity 30 cm/s, purge at room temperature 15–30 minutes, program oven at 10–15 °C/min to column maximum temperature, hold 30 minutes.
  • Do not use hydrogen as carrier for conditioning due to explosion hazard.
  • Bakeout procedure: set column flow to normal or 30 cm/s, split vent 200 mL/min, purge at least 10 minutes, inlet 300 °C or 25 °C above normal, oven 25 °C above method final temperature, bake 30 minutes.
  • Column hanger part number 1460-1914; capillary column clip kit G1530-61580.
  • Finger-tight column nut for 0.53-mm columns 5020-8293; for 0.100- to 0.320-mm columns 5020-8292.
  • No-hole ferrule 5181-3308 for testing; capillary column blanking nut 5020-8294.
pp. 36-57
Maintaining the Split/Splitless Inlet

Details consumables and parts for the Split/Splitless inlet, including liners, septa, O-rings, gold seals, and split vent traps. Provides procedures for installing a capillary column, changing the septum, cleaning the septum seat, changing liner and O-ring, replacing the gold seal, replacing the split vent filter, cleaning the inlet, and bakeout.

  • Split liner options include low-pressure drop glass wool 870 µL (5183-4647) and glass wool 990 µL (19251-60540).
  • Splitless liners include single taper glass wool 900 µL (5062-3587) and single taper no glass wool 900 µL (5181-3316).
  • Septum retainer nut 18740-60835; 11-mm high-temperature low-bleed septum 5183-4757.
  • Nonstick fluorocarbon liner O-ring for up to 350 °C: 5188-5365; graphite O-ring for above 350 °C: 5180-4168 (split) or 5180-4173 (splitless).
  • Gold-plated seal standard 5188-5367; with cross for high split flows 5182-9652.
  • Split vent trap PM kit single cartridge 5188-6495.
  • After column installation, wait 10 minutes after GC ready before igniting detector flame or bead.
  • Bakeout: split mode, column flow normal or 30 cm/s, split vent 200 mL/min, purge 10 minutes, inlet 300 °C or 25 °C above normal, oven 25 °C above final method temperature, 30 minutes.
pp. 60-67
Maintaining the Purged Packed Inlet

Lists parts for the Purged Packed inlet, including glass liners, column adapters, septa, O-rings, and packed column nuts and ferrules. Includes procedure for installing a capillary column with the Purged Packed inlet.

  • Preventative maintenance kit 5188-6498.
  • Glass liner 5080-8732 (25/pack) or deactivated 5181-3382 (5/pack).
  • 0.53-mm column adapter 19244-80540; 1/8-inch adapter 19243-80530; 1/4-inch adapter 19243-80540.
  • Viton O-ring for top insert weldment 5080-8898 (12/pk).
  • After column installation, for FPD+ immediately turn off flame; for NPD set bead voltage to 0.0.
  • Wait 10 minutes after GC ready before igniting detector flame or adjusting NPD offset.
pp. 208-231
Maintaining the NPD

Covers maintenance of the Nitrogen Phosphorus Detector (NPD), including bead assembly replacement, collector and jet maintenance, leakage current check, and bakeout. Includes bead conditioning and offset adjustment procedures.

  • After bead installation, heat detector to 150 °C for 15 minutes, then 250 °C for 15 minutes, then operating temperature 310–320 °C for 15 minutes.
  • Check NPD leakage current; if > 2.0 pA, verify bead installation or see Troubleshooting manual.
  • Default offset is 30 pA for white or black beads; 25–30 pA sufficient for most applications; higher offset may shorten bead life.
  • Reset bead counter after maintenance.
pp. 146-178
Maintaining the FID

Covers maintenance of the Flame Ionization Detector (FID), including jet selection, capillary column adapter installation, collector assembly replacement, jet replacement, leakage current check, baseline check, and bakeout.

  • FID cleaning kit 9301-0985 contains appropriate brushes for cleaning detectors and inlets.
  • Selecting an FID jet is covered.
  • Check FID leakage current and baseline.
  • Bakeout the FID.
pp. 182-190
Maintaining the TCD

Covers maintenance of the Thermal Conductivity Detector (TCD), including capillary column installation, optional capillary column adapter, and bakeout. Warns that the TCD filament will be damaged if exposed to air while hot.

  • Turn off TCD filament before maintenance to prevent damage from air exposure while hot.
  • Optional TCD capillary column adapter available.
  • Bakeout contaminants from the TCD.
pp. 192-205
Maintaining the µECD

Covers maintenance of the micro Electron Capture Detector (µECD), including important safety information, replacement of the fused silica indented mixing liner, installation of the makeup gas adapter, capillary column installation, insulating cup installation, and bakeout.

  • Important safety information about the µECD is provided.
  • Replace fused silica indented mixing liner and install makeup gas adapter.
  • Install capillary column in the µECD.
  • Install insulating cup for the µECD.
  • Bakeout the µECD.
pp. 234-249
Maintaining the FPD+

Covers maintenance of the Flame Photometric Detector Plus (FPD+), including packed column adapter installation, capillary column attachment, wavelength filter change, cover removal and installation, ignitor replacement, and brazement cleaning.

  • Install packed column adapter in the FPD+.
  • Attach capillary column to the FPD+.
  • Change the FPD+ wavelength filter.
  • Replace the FPD+ ignitor.
  • Clean the FPD+ brazement.
pp. 252-266
Maintaining the G3435A/G3436A FPD

Covers maintenance of the G3435A/G3436A Flame Photometric Detector (FPD), including capillary column adapter installation, capillary column attachment, wavelength filter change, vent tube removal and installation, and ignitor replacement.

  • Install capillary column adapter in the FPD.
  • Attach capillary column to the FPD.
  • Change the FPD wavelength filter.
  • Remove and install the FPD vent tube and cover.
  • Replace the FPD ignitor.
pp. 268-270
Maintaining the PCM

Covers maintenance of the Pressure Control Module (PCM), including calibration of the PCM interface and installation or replacement of frits.

  • Calibrating the PCM interface.
  • Installing or replacing frits in the PCM.
pp. 274-284
Maintaining a Valve

Covers maintenance of GC valves, including replacement of gas sampling valve loops, alignment of rotary valve rotors, replacement of rotary valves in the valve box, and removal and installation of the upper valve box.

  • Replace a gas sampling valve loop.
  • Align a rotary valve rotor.
  • Replace a rotary valve in the valve box.
  • Remove and install the upper valve box.
pp. 288-292
Swagelok Connections

Provides procedures for making Swagelok connections and using a Swagelok tee.

  • Making Swagelok connections.
  • Using a Swagelok tee.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
General GC Maintenance Method sets oven to 35 °C, all inlet temperatures to 35 °C, inlet gas pressures to 0.0, all detector temperatures to 35 °C, and detector flows to Off. p.13 ↗ Use this method before column, detector, or general GC maintenance to protect the instrument and avoid burns or shocks. 7820A GC
Inlets and detectors at operating temperature may require 12 hours or longer to reach maintenance method setpoints. p.13 ↗ Plan maintenance with sufficient cool-down time; do not assume rapid cooling. 7820A GC
Once zones reach < 70 °C, general GC maintenance can be performed. p.13 ↗ Verify zone temperatures are below 70 °C before touching components to avoid burns. 7820A GC
For FID and NPD maintenance, turn the electrometer Off to disable high voltage. p.13 ↗ Prevents electric shock during detector maintenance. FID and NPD detectors
The TCD filament will be damaged if exposed to air while hot; turn it Off to protect it. p.14 ↗ Always turn off TCD filament before introducing air or performing maintenance. TCD
For FPD+ maintenance, turn off the GC and unplug the power cord. p.13 ↗ Required safety step before FPD+ maintenance. FPD+
Conditioning a capillary column: set minimum velocity 30 cm/s, purge at room temperature 15–30 minutes, program oven at 10–15 °C/min to column maximum temperature, hold 30 minutes. p.29 ↗ Standard conditioning procedure to remove air and contaminants before use. Capillary columns
Do not use hydrogen as carrier for conditioning; it could vent into the oven and present an explosion hazard. p.28 ↗ Use inert carrier gas such as helium or nitrogen for conditioning. Capillary columns
Bakeout procedure for Split/Splitless inlet: split mode, column flow normal or 30 cm/s, split vent 200 mL/min, purge at least 10 minutes, inlet 300 °C or 25 °C above normal, oven 25 °C above method final temperature, bake 30 minutes. p.57 ↗ Removes contaminants from inlet and column; do not exceed column maximum temperature. Split/Splitless inlet
After column installation, wait 10 minutes after GC becomes ready before igniting detector flame or bead. p.43 ↗ Allows flows and temperatures to stabilize before ignition. Split/Splitless inlet and Purged Packed inlet procedures
For FPD+, immediately turn off the flame after column installation; for NPD, immediately set bead voltage to 0.0. p.67 ↗ Prevents detector damage during column installation. FPD+ and NPD
NPD bead conditioning: heat to 150 °C for 15 minutes, then 250 °C for 15 minutes, then operating temperature 310–320 °C for 15 minutes. p.223 ↗ Proper bead conditioning extends bead life and ensures stable response. NPD
Check NPD leakage current; if > 2.0 pA, verify bead installation or see Troubleshooting manual. p.223 ↗ Leakage current above 2.0 pA indicates a problem requiring investigation. NPD
NPD default offset is 30 pA for white or black beads; 25–30 pA is sufficient for most applications; higher offset may shorten bead life. p.223 ↗ Use the lowest offset that gives acceptable performance to maximize bead life. NPD
Split vent trap PM kit single cartridge part number 5188-6495. p.38 ↗ Replacement filter cartridge for split vent line maintenance. Split/Splitless inlet
Nonstick fluorocarbon liner O-ring for temperatures up to 350 °C (5188-5365); graphite O-ring for temperatures above 350 °C (5180-4168 for split, 5180-4173 for splitless). p.38 ↗ Select O-ring material based on operating temperature. Split/Splitless inlet
Gold-plated seal standard application 5188-5367; gold-plated seal with cross for high split flows 5182-9652. p.38 ↗ Use cross seal for high split flow applications. Split/Splitless inlet
Finger-tight column nut for 0.53-mm columns is 5020-8293; for 0.100- to 0.320-mm columns is 5020-8292. p.24 ↗ Select correct column nut by column inner diameter. Capillary columns
Column hanger part number 1460-1914; capillary column clip kit for 7-inch column basket G1530-61580. p.25 ↗ Hardware for installing capillary columns in the oven. Capillary columns
FID cleaning kit 9301-0985 contains appropriate brushes for cleaning detectors and inlets. p.11 ↗ Use correct cleaning tools to avoid damage. FID and inlets