Maintaining Your Gas Chromatograph (GC)
Technical Reference Summary
Covered Models & Document Scope
Agilent 8890 Gas Chromatograph (GC). Covers maintenance of capillary columns, Split/Splitless Inlet (SSL), Purged Packed Inlet (PPI), Cool On-Column Inlet (COC), Multimode Inlet (MMI), Programmable Temperature Vaporization Inlet (PTV), Volatiles Interface (VI), Flame Ionization Detector (FID), Thermal Conductivity Detector (TCD), Electron Capture Detector (ECD), Nitrogen-Phosphorus Detector (NPD), Flame Photometric Detector Plus (FPD+), Auxiliary Electronic Pressure Control Module (Aux EPC), Pneumatics Control Module (PCM), and valves. Includes procedures for covers removal, consumables/parts identification, and Swagelok connections.
Section-by-Section Technical Breakdown
This section provides an overview of the maintenance procedures included in the manual and lists the tools needed for routine maintenance. It also describes the information needed to safely perform the various maintenance tasks.
- Manual details routine tasks needed to maintain the Agilent 8890 GC.
- Procedures assume basic knowledge of tool use and GC operation.
- Chapters cover capillary columns, inlets, detectors, EPC modules, and valves.
- Each chapter includes consumables/parts list, exploded parts view, and detailed procedures.
Table 1 lists the tools needed for most GC maintenance procedures. Specific tools required for a procedure are listed in step 1 of each procedure.
- Common tools include wrenches, screwdrivers, Torx keys, hex keys, flow meters, leak detector, magnifying loupe, metric ruler, bench vise, razor, tweezers, pliers, ESD wrist strap, heat-resistant gloves, and cotton swabs.
- Cleaning tools include cleaning brushes, jet cleaning wire, lint-free cloth, ultrasonic bath, nylon gloves, and steel wool.
- Part numbers are provided for many tools.
Before routine maintenance, the GC must be made ready by setting low temperatures, reduced flows, turning off and unplugging the GC, venting an MSD, and making other settings to prevent damage. Use the touchscreen to enter Maintenance Mode.
- Use Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Start Maintenance to put GC in standby.
- Follow prompts; select Finished to exit maintenance mode and restore previous settings.
- Maintenance Mode ends when GC is turned off.
- If not using built-in features, cool all heated zones, turn off GC, unplug power cord, and turn off cryogen supply and bleed residual backpressure.
The GC can display signal and other data useful during maintenance, such as instrument status, status details, and module status details.
- Instrument status: readiness state, running state.
- Instrument status details: not ready modules, error modules, warnings.
- Module status details: actual/setpoint pairs, quick on/off, firmware version, instrument IP address.
- Add items to the GC status listing as needed.
Agilent provides the Parts Finder tool to help locate replacement and consumable part numbers. Part numbers are also included in this manual.
- Parts Finder is installed with Agilent data systems or available on the Agilent GC and GC/MS User Manuals & Tools DVD.
- Navigate to the part graphically based on its location in the GC.
- Part numbers are also listed in the manual.
This section describes how to remove the detector top cover, pneumatics cover, and electronics cover as needed for routine maintenance. Only these covers should be removed.
- Detector top cover protects detectors, valve box, and valve assembly.
- Pneumatics cover removal procedure.
- Electronics cover removal procedure depends on whether an FPD+ is installed.
- Always turn off GC and unplug power cord before removing covers.
Table 2 lists consumable parts for the hydrogen sensor module accessory and for general GC maintenance, including fittings, tubing, gas regulators, and Gas Clean filters.
- Hydrogen sensor parts: calibration gas cylinder (2% H2 in 17 L compressed air), cylinder stand, pressure regulator.
- Fittings and hardware: 1/8-inch brass fittings, PTFE tape, copper tubing.
- Gas regulators: 2-stage, brass body, stainless steel diaphragms, 125 psi max, various CGA fittings.
- Gas Clean filters and connecting units for various gas lines.
Procedure to replace the calibration gas cylinder when it cannot provide correct flow rate or if expired. Connect only Agilent calibration gas (2% hydrogen in air).
- Cool oven and detector to < 40 °C before proceeding.
- Turn pressure regulator completely off (full CCW).
- Connect flowmeter tube to sensor tube in oven.
- Start calibration cycle via Settings > Calibration > Hydrogen Sensor > Start Calibration Cycle? > On/Yes.
- Adjust pressure regulator until flow rate is approximately 30 mL/min.
- Allow calibration cycle to complete (approximately 5 minutes).
- Check for leaks.
Tables 3 and 4 list nuts, ferrules, hardware, and hangers for capillary columns, organized by column inner diameter.
- Ferrules and column nuts for 0.530 mm, 0.320 mm, 0.250 mm, 0.100/0.200 mm column IDs.
- Universal column nut, collared self-tightening nuts, column cutter, diamond pencil, ferrule tool kit.
- Column hanger (1460-1914) and capillary column clip kit (G1530-61580).
Procedure to install a capillary column hanger in the oven. Place GC in maintenance mode, select front or back hanger position, and insert hanger ends into slots.
- Place GC in maintenance mode.
- Select front or back hanger position.
- Insert hanger ends into slots in selected position.
Procedure to install capillary column clips using the clip kit and a T-20 Torx screwdriver. Loosen four corner mounting screws, slip screws through large holes on clips, slide clips into slots, and tighten.
- Gather capillary column clip kit and T-20 Torx screwdriver.
- Place GC in maintenance mode.
- Loosen four corner mounting screws but do not remove.
- Slip each screw through large hole on clip, slide clip so screw is in slot.
- Tighten screws to hold clips; fully tighten after column installation.
Generic procedure to condition a capillary column. Always follow column manufacturer's recommendations. Do not use hydrogen as carrier for conditioning.
- Set inlet and oven temperature to < 40 °C.
- Install column into inlet using new ferrules.
- Cap any open column fittings.
- Turn off all detectors.
- Set minimum velocity of 30 cm/s or as recommended.
- Set oven to 120 °C and let gas flow for 15 to 30 minutes.
- Program oven from 120 °C to maximum temperature limit at 10 to 15 °C/min, hold 30 minutes.
- Cool oven to < 40 °C before attaching column to detector.
Procedure to cut a loop from a column when needed. Loosen inlet column nut, remove column, uncoil one loop, cut unwanted loop, and reinstall with new ferrules.
- Gather new ferrule(s) and column cutter.
- Place GC in maintenance mode.
- Loosen inlet column nut and remove column from inlet.
- Uncoil one loop of column from hanger.
- Cut unwanted loop from column.
- Install column into inlet using new ferrules.
- Exit maintenance mode.
Procedure to reverse column position and bake out contaminants. Disconnect column, reverse ends, reattach to inlet and detector, then bake out at elevated temperature.
- Disconnect column from inlet and detector.
- Reverse column position on hanger.
- Attach column to inlet and detector.
- Set column flow to normal operating value or velocity to 30 cm/s.
- For Multimode, Split/Splitless, PTV, and VI inlets, select split mode and set split vent flow to 200 mL/min.
- Purge column with carrier flow for at least 10 minutes before heating.
- Set inlet temperature to 300 °C or 25 °C above normal operating temperature.
- Set oven 25 °C above method final oven temperature, do not exceed column maximum.
- Bakeout for 30 minutes.
Procedure to attach a capillary column using UltiMetal Plus Flexible Metal ferrules. Ferrules are swaged to the column by tightening the internal nut or column nut a specified number of degrees.
- Available ferrule packages: G3188-27501 (0.1–0.25 mm id), G3188-27502 (0.32 mm id), G3188-27503 (0.45–0.53 mm id), G3188-27504 (plug), G3188-27505 (0.25–0.32 mm UltiMetal id), G3188-27506 (0.53 mm UltiMetal id).
- Swaging tool for CFT fitting: G3440-80227; for inlet: G3440-80218.
- Tighten internal nut or column nut to lower end of range, check grip, continue in 5–15 degree increments until gripping occurs.
- Then turn an additional 15–20 degrees clockwise.
- Degrees to tighten: G3188-27501: 50–100°, G3188-27502: 30–70°, G3188-27503: 20–50°, G3188-27504: 60°, G3188-27505: 40–90°, G3188-27506: 20–50°.
Tables 8, 9, and 10 list liners, nuts, ferrules, septa, O-rings, and other consumables for the Split/Splitless Inlet.
- Liners for split, splitless, direct inject, and direct column connect modes.
- Nuts and ferrules for capillary columns by column ID.
- Septum retainer nuts, 11-mm septa (high-temperature low-bleed, prepierced long life), O-rings, split vent trap, gold-plated seals.
Procedure to replace the NPD bead assembly, including closing covers, exiting maintenance mode, restoring gas flows, heating detector, checking leakage current, and adjusting offset.
- Close NPD cover, install GC detector top cover and electronics top cover; all covers must be closed for stable baseline.
- Exit maintenance mode via Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Finished.
- Restore normal NPD operating gas flows.
- Heat detector to 150 °C for 15 minutes, then 250 °C for 15 minutes.
- Increase to operating value (310 to 320 °C recommended) and allow 15 minutes for equilibration.
- Check NPD leakage current; if > 2.0 pA, verify bead installation or see Troubleshooting manual.
- Start Adjust offset process; default offset is 20 pA for Blos beads; 25 to 30 pA sufficient for most applications; higher offset may shorten bead life.
- Reset bead counter.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Hydrogen sensor calibration gas is 2% hydrogen in air; flow rate should be adjusted to approximately 30 mL/min during calibration. p.27 ↗ | Use only Agilent calibration gas; higher hydrogen concentrations can create a dangerous condition or damage the sensor. | 8890 GC with hydrogen sensor module |
| Cool oven and detector to < 40 °C before performing hydrogen sensor calibration. p.27 ↗ | Prevents burns; calibration can be performed at any temperature but cooling is required for safe handling. | 8890 GC |
| For column conditioning, set minimum velocity of 30 cm/s, oven to 120 °C, flow gas for 15 to 30 minutes, then program from 120 °C to column maximum at 10 to 15 °C/min and hold 30 minutes. p.34 ↗ | Do not use hydrogen as carrier for conditioning; follow column manufacturer's recommendations. | Capillary columns |
| When reversing a column for bakeout, set split vent flow to 200 mL/min for Multimode, Split/Splitless, PTV, and VI inlets, purge for at least 10 minutes, set inlet to 300 °C or 25 °C above normal, oven 25 °C above method final temperature, and bakeout for 30 minutes. p.38 ↗ | Do not exceed column manufacturer's maximum temperature limit. | Capillary columns with specified inlets |
| UltiMetal Plus Flexible Metal ferrules are swaged by tightening the internal nut or column nut to a specified degree range, then an additional 15–20 degrees. p.40 ↗ | Applying excessive force will not provide a better seal; compressing the ferrule less minimizes damage to fittings. | UltiMetal Plus Flexible Metal fittings |
| Ferrule swaging degrees: G3188-27501: 50–100°, G3188-27502: 30–70°, G3188-27503: 20–50°, G3188-27504: 60°, G3188-27505: 40–90°, G3188-27506: 20–50°. p.42 ↗ | Use the correct degree range for the ferrule part number to ensure a leak-free seal without over-tightening. | UltiMetal Plus Flexible Metal ferrules |
| NPD operating temperature recommended is 310 to 320 °C. p.223 ↗ | After bead replacement, heat detector to 150 °C for 15 min, then 250 °C for 15 min, then to operating value and allow 15 min for equilibration. | NPD detector |
| NPD leakage current should be ≤ 2.0 pA; if > 2.0 pA, verify bead installation or see Troubleshooting manual. p.223 ↗ | High leakage current indicates improper bead installation or other issues. | NPD detector |
| Default NPD offset is 20 pA for Blos beads; an offset of 25 to 30 pA is sufficient for most applications; higher offset may shorten bead life. p.223 ↗ | Adjust offset according to application needs while considering bead life. | NPD detector |
| All covers must be closed to get a stable NPD baseline. p.223 ↗ | Ensure NPD cover, detector top cover, and electronics top cover are installed before checking baseline. | NPD detector |
| Use Maintenance Mode via touchscreen: Maintenance > Instrument > Perform Maintenance > Maintenance Mode > Start Maintenance to make GC safe for maintenance. p.13 ↗ | Maintenance Mode ends when GC is turned off; use Finished to exit and restore previous settings. | 8890 GC |
| Only the detector top cover, pneumatics cover, and electronics cover should be removed; removing other covers can compromise safety features. p.17 ↗ | Removing other covers may lead to personal injury or damage to the instrument. | 8890 GC |
| Gas regulators for GC supply are 2-stage, brass body, stainless steel diaphragms, 125 psi max, with various CGA fittings. p.24 ↗ | Select regulator with appropriate CGA fitting for the gas type; for 1/4-inch tubing purchase a 1/4-inch adapter. | 8890 GC gas supply |
| Hydrogen sensor calibration cycle takes approximately 5 minutes total. p.27 ↗ | Allow sufficient time for calibration to complete. | 8890 GC with hydrogen sensor module |
| When installing capillary column clips, use a T-20 Torx screwdriver and loosen four corner mounting screws but do not remove them. p.33 ↗ | Slip each screw through the large hole on the clip, slide clip so screw is in slot, then tighten. | 8890 GC oven |
| For column conditioning, do not use hydrogen as carrier gas; it could vent into the oven and present an explosion hazard. p.34 ↗ | Use an inert carrier gas such as helium or nitrogen for conditioning. | Capillary columns |
| When reversing a column, set the inlet temperature to 300 °C or 25 °C above the normal operating temperature. p.39 ↗ | This helps bake contaminants out of the inlet through the split vent. | Capillary columns with specified inlets |
| The GC can display instrument status, status details, and module status details including actual/setpoint pairs, firmware version, and IP address. p.14 ↗ | Use status data during maintenance to verify module readiness and settings. | 8890 GC |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| NPD leakage current > 2.0 pA p.223 ↗ | Improper bead installation or other issue. | Verify bead installation or see the Troubleshooting manual. | After replacing NPD bead assembly |