Induztr
Engineering Digest Document Viewer Related Models

Characterization of Microplastics in Environmental Samples by Laser Direct Infrared Imaging

Technical Reference Summary

Covered Models & Document Scope

Application note for the Agilent 8700 LDIR chemical imaging system (QCL-based laser direct infrared imaging) used for characterization of microplastics in environmental samples; covers library generation from synthetic polymers (PTFE, PMMA, PET, PC, PS, PP, PA) and analysis of two beach samples.

Technical Data

Parameter Value Condition
Wavenumber range 1,800 to 975 cm–1 p.2 ↗ 8700 LDIR, QCL source
Infrared source semiconductor-based quantum cascade laser (QCL) p.2 ↗ 8700 LDIR
Detector single-point, thermoelectrically cooled mercury cadmium telluride (MCT) p.2 ↗ 8700 LDIR
Full spectrum acquisition time less than one second s p.2 ↗ objective parked at single point, QCL sweep
Studied particle size range 20 to 600 μm p.6 ↗ samples 1 and 2
Scanned area, sample 1 (red and white plastic) 3.04 × 3.06 mm2 p.7 ↗ sample 1
Scanned area, sample 2 (yellow film) 3.76 × 4.42 mm2 p.7 ↗ sample 2
Substrate 7.5 × 2.5 cm p.2 ↗ MirrIR, Kevley glass slides

Operating Limits

  • Studied particle size range: 20 to 600 μm p.6 ↗
  • Most particles ranged between: 20 to 200 μm p.6 ↗