Characterization of Microplastics in Environmental Samples by Laser Direct Infrared Imaging
Technical Reference Summary
Covered Models & Document Scope
Application note for the Agilent 8700 LDIR chemical imaging system (QCL-based laser direct infrared imaging) used for characterization of microplastics in environmental samples; covers library generation from synthetic polymers (PTFE, PMMA, PET, PC, PS, PP, PA) and analysis of two beach samples.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Wavenumber range | 1,800 to 975 cm–1 p.2 ↗ | 8700 LDIR, QCL source |
| Infrared source | semiconductor-based quantum cascade laser (QCL) p.2 ↗ | 8700 LDIR |
| Detector | single-point, thermoelectrically cooled mercury cadmium telluride (MCT) p.2 ↗ | 8700 LDIR |
| Full spectrum acquisition time | less than one second s p.2 ↗ | objective parked at single point, QCL sweep |
| Studied particle size range | 20 to 600 μm p.6 ↗ | samples 1 and 2 |
| Scanned area, sample 1 (red and white plastic) | 3.04 × 3.06 mm2 p.7 ↗ | sample 1 |
| Scanned area, sample 2 (yellow film) | 3.76 × 4.42 mm2 p.7 ↗ | sample 2 |
| Substrate | 7.5 × 2.5 cm p.2 ↗ | MirrIR, Kevley glass slides |
Operating Limits
- Studied particle size range: 20 to 600 μm p.6 ↗
- Most particles ranged between: 20 to 200 μm p.6 ↗