Size Exclusion Chromatography MALS Characterization of mRNA
Technical Reference Summary
Covered Models & Document Scope
Application note covering SEC-MALS analysis of IVT mRNA using the Agilent 1290 Infinity III Bio LC System (G7131A Bio Flexible Pump, G7137A Bio Multisampler with InfinityLab Sample Thermostat option 101), Agilent 1260 Infinity II Multi-Angle Light Scattering Detector, AdvanceBio SEC 1000 Å wide-pore column (part number PL1180-5302), and Agilent Advanced GPC/SEC Software for OpenLab CDS 2.8
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Injection volume | 15–25 µL p.3 ↗ | |
| Stop time | 30 min p.3 ↗ | |
| Detection wavelength (UV) | 260 nm p.3 ↗ | peak width > 0.05 min (10 Hz) |
| Column temperature | 30 °C p.3 ↗ | |
| MALS detection angle | 90 ° p.3 ↗ | blue signal |
| Number of simultaneous MALS angles | 20 angles p.3 ↗ | |
| Column pore size | 1000 Å p.2 ↗ | AdvanceBio SEC |
| Mobile phase pH | 7 p.2 ↗ | adjusted with sodium hydroxide |
| Sample filtration pore size | 0.22 μm p.2 ↗ | |
| Software version | 2.8 p.2 ↗ | Agilent OpenLab CDS |
| Refractive index increment (dn/dc) | 0.172 p.3 ↗ | used as RI for SEC-MALS |
Operating Limits
- Column temperature: 30 °C p.3 ↗
- Mobile phase pH: 7 p.2 ↗
- Sample filtration pore size: 0.22 μm p.2 ↗
- Injection volume range: 15–25 µL p.3 ↗
- UV peak width detection threshold: > 0.05 min (10 Hz) p.3 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| PL1180-5302 p.2 ↗ | PL1180-5302 | AdvanceBio SEC column, 1000 Å pore size |