Analysis of Aldehydes in Beer by Agilent PAL 3 Autosampler and 5977C GC/MSD

Technical Reference Summary

Covered Models & Document Scope

Application note covering the analysis of four aldehydes (hexanal, furfural, phenylacetaldehyde, trans-2-nonenal) in beer using an Agilent PAL3 RTC autosampler with SPME tool and 8890/5977C GC/MSD; no product model/size variants or ordering codes are given.

Technical Data

Parameter Value Condition
Flavor threshold, hexanal 350 μg/L p.2 ↗
Flavor threshold, furfural 15.157 mg/L p.2 ↗
Flavor threshold, phenylacetaldehyde 105 μg/L p.2 ↗
Flavor threshold, trans-2-nonenal 0.1 μg/L p.2 ↗
Calibration range, hexanal 0.05 to 10 µg/L p.6 ↗
Calibration range, furfural 5 to 1,000 µg/L p.6 ↗
Calibration range, phenylacetaldehyde 0.1 to 50 µg/L p.6 ↗
Calibration range, trans-2-nonenal 0.025 to 5 µg/L p.6 ↗
Coefficient of determination R², hexanal 0.999 p.6 ↗
Coefficient of determination R², furfural 0.998 p.6 ↗
Coefficient of determination R², phenylacetaldehyde 0.996 p.6 ↗
Coefficient of determination R², trans-2-nonenal 0.998 p.6 ↗
Repeatability %RSD (n = 3), hexanal in beer 1.3 to 4.9 % p.7 ↗ Brands 1–4
Repeatability %RSD (n = 3), furfural in beer 1.3 to 2.8 % p.7 ↗ Brands 1–4
Repeatability %RSD (n = 3), phenylacetaldehyde in beer 0.9 to 3.7 % p.7 ↗ Brands 1–4
Repeatability %RSD (n = 3), trans-2-nonenal in beer 1.7 to 2.8 % p.7 ↗ Brands 1–4
Method detection limit, hexanal 0.0009 µg/L p.10 ↗ 0.05 µg/L level
Method detection limit, furfural 0.52 µg/L p.10 ↗ 5 µg/L level
Method detection limit, phenylacetaldehyde 0.015 µg/L p.10 ↗ 0.1 µg/L level
Method detection limit, trans-2-nonenal 0.003 µg/L p.10 ↗ 0.025 µg/L level

Operating Limits

  • Calibration range, hexanal: 0.05 to 10 µg/L p.6 ↗
  • Calibration range, furfural: 5 to 1,000 µg/L p.6 ↗
  • Calibration range, phenylacetaldehyde: 0.1 to 50 µg/L p.6 ↗